Magnetic resonance only workflow and validation of dose calculations for radiotherapy of prostate cancer

Abstract Background: Current state of the art radiotherapy planning of prostate cancer utilises magnetic resonance (MR) for soft tissue delineation and computed tomography (CT) to provide an electron density map for dose calculation. This dual scan workflow is prone to setup and registration error. This study evaluates the feasibility of an MR-only workflow and the validity of dose calculation from an MR derived pseudo CT. Material and methods: Thirty prostate cancer patients were CT and MR scanned. Clinical treatment plans were generated on CT using a single 18 MV arc volumetric modulated arc therapy (VMAT) with a prescription of 78 Gy/39 fractions. Dose was recalculated on pseudo CT and assuming uniform water density. Pseudo CT and uniform density based dose calculations were compared to CT dose calculations by gamma analysis. One patient was treated with a plan based solely on MR and pseudo CT including daily image guided radiotherapy (IGRT) performed by manual match of implanted gold markers. Results: A pseudo CT was generated for 29 of the 30 patients. Median gamma pass rates for 1%/1 mm passing criteria for dose calculated on pseudo CT when compared to CT were 100% for most evaluated structures. Dose calculated on uniform density also yielded high median pass rates, but with a higher occurrence of pass rates below 95%. Cases of pass rate below 95% on pseudo CT proved to originate from the presence of rectal air on CT, not represented by the pseudo CT. Treatment based on MR alone was successfully delivered to one patient, including manually performed daily IGRT. Conclusions: Median gamma pass rates were high for pseudo CT and proved superior to uniform density. Local differences in dose calculations were concluded not to have clinical relevance. Feasibility of the MR-only workflow was demonstrated through successful delivery of a treatment course planned based on MR alone.

[1]  David Followill,et al.  Comparison of 2D and 3D gamma analyses. , 2014, Medical physics.

[2]  Adam Johansson,et al.  CT substitute derived from MRI sequences with ultrashort echo time. , 2011, Medical physics.

[3]  Tiina Seppälä,et al.  A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer. , 2013, Medical physics.

[4]  R. Hoogeveen,et al.  MR-only simulation for radiotherapy planning , 2015 .

[5]  Jan J W Lagendijk,et al.  The magnetic resonance imaging-linac system. , 2014, Seminars in radiation oncology.

[6]  M van Herk,et al.  Definition of the prostate in CT and MRI: a multi-observer study. , 1999, International journal of radiation oncology, biology, physics.

[7]  Alan Pollack,et al.  MRI-based treatment planning for radiotherapy: dosimetric verification for prostate IMRT. , 2004, International journal of radiation oncology, biology, physics.

[8]  V S Khoo,et al.  New developments in MRI for target volume delineation in radiotherapy. , 2006, The British journal of radiology.

[9]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[10]  Sasa Mutic,et al.  The ViewRay system: magnetic resonance-guided and controlled radiotherapy. , 2014, Seminars in radiation oncology.

[11]  Max A. Viergever,et al.  elastix: A Toolbox for Intensity-Based Medical Image Registration , 2010, IEEE Transactions on Medical Imaging.

[12]  Olivier Salvado,et al.  An atlas-based electron density mapping method for magnetic resonance imaging (MRI)-alone treatment planning and adaptive MRI-based prostate radiation therapy. , 2012, International journal of radiation oncology, biology, physics.

[13]  Karin Haustermans,et al.  The value of magnetic resonance imaging for radiotherapy planning. , 2014, Seminars in radiation oncology.

[14]  Jurgen Fripp,et al.  Automatic Substitute Computed Tomography Generation and Contouring for Magnetic Resonance Imaging (MRI)-Alone External Beam Radiation Therapy From Standard MRI Sequences. , 2015, International journal of radiation oncology, biology, physics.