Electrical impedance imaging system using FPGAs for flexibility and interoperability

BackgroundModern EIT systems require simultaneously operating multiple functions for flexibility, interoperability, and clinical applicability. To implement versatile functions, expandable design and implementation tools are needed. On the other hand, it is necessary to develop an ASIC-based EIT system to maximize its performance. Since the ASIC design is expensive and unchangeable, we can use FPGAs as a prior step to the digital ASIC design and carefully classify which functions should be included in the ASIC. In this paper, we describe the details of the FPGA design adopted in the KHU Mark2.5 EIT system.MethodsWe classified all functions of the KHU Mark2.5 EIT system into two categories. One is the control and processing of current injection and voltage measurement. The other includes the collection and management of the multi-channel data with timing controls for internal and external interconnections. We describe the implementation of these functions in two kinds of FPGAs called the impedance measurement module (IMM) FPGA and the intra-network controller FPGA.ResultsWe present functional and timing simulations of the key functions in the FPGAs. From phantom and animal imaging experiments, we show that multiple functions of the system are successfully implemented in the FPGAs. As examples, we demonstrate fast multi-frequency imaging and ECG-gated imaging.ConclusionGiven an analog design of a parallel EIT system, it is important to optimize its digital design to minimize systematic artifacts and maximize performance. This paper described technical details of the FPGA-based fully parallel EIT system called the KHU Mark2.5 with numerous functions needed for clinical applications. Two kinds of FPGAs described in this paper can be used as a basis for future EIT digital ASIC designs for better application-specific human interface as well as hardware performance.

[1]  D S Holder,et al.  Identification of a suitable current waveform for acute stroke imaging , 2006, Physiological measurement.

[2]  G.J. Saulnier,et al.  ACT3: a high-speed, high-precision electrical impedance tomograph , 1991, IEEE Transactions on Biomedical Engineering.

[3]  Eung Je Woo,et al.  A fully parallel multi-frequency EIT system with flexible electrode configuration: KHU Mark2 , 2011, Physiological measurement.

[4]  P Bertemes-Filho,et al.  A comparison of modified Howland circuits as current generators with current mirror type circuits. , 2000, Physiological measurement.

[5]  Eung Je Woo,et al.  Validation of a multi-frequency electrical impedance tomography (mfEIT) system KHU Mark1: impedance spectroscopy and time-difference imaging , 2008, Physiological measurement.

[6]  Eung Je Woo,et al.  Feasibility of magnetic resonance electrical impedance tomography (MREIT) conductivity imaging to evaluate brain abscess lesion: In vivo canine model , 2013, Journal of magnetic resonance imaging : JMRI.

[7]  Andreas Demosthenous,et al.  Comparison of a new integrated current source with the modified Howland circuit for EIT applications , 2009, Physiological measurement.

[8]  I Frerichs,et al.  Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activities. , 2000, Physiological measurement.

[9]  Ryan Halter,et al.  FPGA Based High Speed Data Acquisition System for Electrical Impedance Tomography. , 2013, Journal of physics. Conference series.

[10]  Eung Je Woo,et al.  Multi-frequency EIT system with radially symmetric architecture: KHU Mark1 , 2007, Physiological measurement.

[11]  John G. Webster,et al.  An Impedance Camera for Spatially Specific Measurements of the Thorax , 1978, IEEE Transactions on Biomedical Engineering.

[12]  Andy Adler,et al.  Electrical impedance tomography system based on active electrodes , 2012, Physiological measurement.

[13]  Harki Tanaka,et al.  Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. , 2004, American journal of respiratory and critical care medicine.

[14]  Xicai Yue,et al.  FPGA design and implementation for EIT data acquisition. , 2008, Physiological measurement.

[15]  G.J. Saulnier,et al.  An Electrical Impedance Spectroscopy System for Breast Cancer Detection , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[16]  B H Brown,et al.  Electrical impedance tomography (EIT): a review , 2003, Journal of medical engineering & technology.

[17]  William R B Lionheart,et al.  GREIT: a unified approach to 2D linear EIT reconstruction of lung images , 2009, Physiological measurement.

[18]  David S. Holder,et al.  Electrical Impedance Tomography : Methods, History and Applications , 2004 .

[19]  Alexander S Ross,et al.  Current source design for electrical impedance tomography. , 2003, Physiological measurement.

[20]  J. Marcus,et al.  Noninvasive assessment and monitoring of the pulmonary circulation , 2005, European Respiratory Journal.

[21]  R H Smallwood,et al.  Mk3.5: a modular, multi-frequency successor to the Mk3a EIS/EIT system. , 2001, Physiological measurement.

[22]  Keith D. Paulsen,et al.  A Broadband High-Frequency Electrical Impedance Tomography System for Breast Imaging , 2008, IEEE Transactions on Biomedical Engineering.

[23]  F J Lidgey,et al.  Development of a real-time adaptive current tomograph. , 1994, Physiological measurement.

[24]  Eung Je Woo,et al.  Multi-Frequency Electrical Impedance Tomography System With Automatic Self-Calibration for Long-Term Monitoring , 2014, IEEE Transactions on Biomedical Circuits and Systems.