Evaluation of a personal and environmental dosemeter based on CR-39 track detectors in quasi-monoenergetic neutron fields.
暂无分享,去创建一个
In this paper, the evaluation of the dosimetric capability of a detector based on a CR-39 solid-state nuclear track detector coupled to a 1 cm thickness of PMMA radiator was made with the aim of understanding the applicability of this technique to personal and environmental neutron dosimetry. The dosemeter has been exposed to monoenergetic and quasi-monoenergetic neutron beams at PTB in Braunschweig, Germany and at Ithemba Laboratories, in Faure, South Africa, with peak energies ranging from 0.565 to 100 MeV. The results showed a response that is almost independent of the neutron energy in the whole energy range.