A novel method of calibrating a MEMS inertial reference unit on a turntable under limited working conditions

Micro-electro-mechanical systems (MEMS) inertial measurement devices tend to be widely used in inertial navigation systems and have quickly emerged on the market due to their characteristics of low cost, high reliability and small size. Calibration is the most effective way to remove the deterministic error of an inertial reference unit (IRU), which in this paper consists of three orthogonally mounted MEMS gyros. However, common testing methods in the lab cannot predict the corresponding errors precisely when the turntable's working condition is restricted. In this paper, the turntable can only provide a relatively small rotation angle. Moreover, the errors must be compensated exactly because of the great effect caused by the high angular velocity of the craft. To deal with this question, a new method is proposed to evaluate the MEMS IRU's performance. In the calibration procedure, a one-axis table that can rotate a limited angle in the form of a sine function is utilized to provide the MEMS IRU's angular velocity. A new algorithm based on Fourier series is designed to calculate the misalignment and scale factor errors. The proposed method is tested in a set of experiments, and the calibration results are compared to a traditional calibration method performed under normal working conditions to verify their correctness. In addition, a verification test in the given rotation speed is implemented for further demonstration.

[1]  Chris J. Bleakley,et al.  Fast calibration of a 9-DOF IMU using a 3 DOF position tracker and a semi-random motion sequence , 2016 .

[2]  Yuanxin Wu,et al.  Improved multi-position calibration for inertial measurement units , 2009 .

[3]  Quan Zhang,et al.  An in situ hand calibration method using a pseudo-observation scheme for low-end inertial measurement units , 2012 .

[4]  Jan Kenneth Bekkeng Calibration of a Novel MEMS Inertial Reference Unit , 2009, IEEE Transactions on Instrumentation and Measurement.

[5]  Chunhua Ren,et al.  A Novel Self-Calibration Method for MIMU , 2015, IEEE Sensors Journal.

[6]  Rui Zhang,et al.  Calibration of an IMU Using 3-D Rotation Platform , 2014, IEEE Sensors Journal.

[7]  Hai Yang,et al.  Development of an analytical method for IMU calibration , 2016, 2016 13th International Multi-Conference on Systems, Signals & Devices (SSD).

[8]  Osama Moselhi,et al.  Inertial measurement unit based indoor localization for construction applications , 2016 .

[9]  Naser El-Sheimy,et al.  A new multi-position calibration method for MEMS inertial navigation systems , 2007 .

[10]  Jari Kangas,et al.  An enhanced multi-position calibration method for consumer-grade inertial measurement units applied and tested , 2010 .

[11]  Jiazhen Lu,et al.  Applied system-level method in calibration validation for personal navigation system in field , 2017 .

[12]  Naser El-Sheimy,et al.  A Standard Testing and Calibration Procedure for Low Cost MEMS Inertial Sensors and Units , 2008, Journal of Navigation.

[13]  Bing Luo,et al.  Observability Analysis of a MEMS INS/GPS Integration System with Gyroscope G-Sensitivity Errors , 2014, Sensors.

[14]  Jiancheng Fang,et al.  Integrated Calibration Method for Dithered RLG POS Using a Hybrid Analytic/Kalman Filter Approach , 2013, IEEE Transactions on Instrumentation and Measurement.