CPMG measurements and ultrafast imaging in human lungs with hyperpolarized helium‐3 at low field (0.1 T)

This work reports the use of single‐shot spin echo sequences to achieve in vivo diffusion gas measurements and ultrafast imaging of human lungs, in vivo, with hyperpolarized 3He at 0.1 T. The observed transverse relaxation time of 3He lasted up to 10 s, which made it possible to use long Carr‐Purcell‐Meiboom‐Gill echo trains. Preliminary NMR studies showed that the resolution of lung images acquired with hyperpolarized 3He and single‐shot sequences is limited to about 6 mm because of the diffusion of the gas in applied field gradients. Ultrafast images of human lungs in normal subjects, achieved in less than 0.4 s with the equivalent of only 130 μmol of fully polarized 3He, are presented. Comparison with other studies shows that there is no SNR penalty by using low fields in the hyperpolarized case. Advantage was taken of the self diffusion‐weighting of the rapid acquisition with relaxation enhancement (RARE) sequence to acquire apparent diffusion coefficient (ADC) images of the lungs. Time scales of seconds could be explored for the first time because there is no hindrance from T  *2 as with the usual approaches. At 0.1 T, 180° RF pulses can be repeated every 10 ms without exceeding specific absorption rate limits, which would not be the case for higher fields. Moreover, at low field, susceptibility‐induced phenomena are expected to be milder. This supports the idea that low‐field imagers can be used for hyperpolarized noble gas MRI of lungs and may be preferred for ADC measurements. Magn Reson Med 47:75–81, 2002. © 2002 Wiley‐Liss, Inc.

[1]  E. Courtade,et al.  Kinetics of helium 3 laser optical pumping , 2000 .

[2]  D. Yablonskiy,et al.  MR imaging of diffusion of 3He gas in healthy and diseased lungs , 2000, Magnetic resonance in medicine.

[3]  L W Hedlund,et al.  Spatially resolved measurements of hyperpolarized gas properties in the lung in vivo. Part I: Diffusion coefficient , 1999, Magnetic resonance in medicine.

[4]  J R MacFall,et al.  Spatially resolved measurements of hyperpolarized gas properties in the lung in vivo. Part II: T∗︁2 , 1999, Magnetic resonance in medicine.

[5]  D. Yablonskiy,et al.  Rapid imaging of hyperpolarized gas using EPI , 1999, Magnetic resonance in medicine.

[6]  R. Ladebeck,et al.  Echo-Planar Imaging Image Artifacts , 1998 .

[7]  G. Guillot,et al.  Low-field 3He nuclear magnetic resonance in human lungs , 1997 .

[8]  L. C. Bourne,et al.  Thin film high temperature superconducting RF coils for low field MRI , 1994, Magnetic resonance in medicine.

[9]  R. Semelka,et al.  T2‐weighted MR imaging of focal hepatic lesions: Comparison of various RARE and fat‐suppressed spin‐echo sequences , 1993, Journal of magnetic resonance imaging : JMRI.

[10]  Partha P. Mitra,et al.  Time-Dependent Diffusion Coefficient of Fluids in Porous Media as a Probe of Surface-to-Volume Ratio , 1993 .

[11]  H. Pfeifer Principles of Nuclear Magnetic Resonance Microscopy , 1992 .

[12]  N. Pelc,et al.  Correction of spatially dependent phase shifts for partial Fourier imaging. , 1988, Magnetic resonance imaging.

[13]  E M Haacke,et al.  The effects of finite sampling in spin‐echo or field‐echo magnetic resonance imaging , 1987, Magnetic resonance in medicine.

[14]  J Hennig,et al.  Rapid nontomographic approach to MR myelography without contrast agents. , 1986, Journal of computer assisted tomography.

[15]  P. Lauterbur,et al.  The sensitivity of the zeugmatographic experiment involving human samples , 1979 .

[16]  J. Liner,et al.  Determination of the Temperature Dependence of Gaseous Diffusion Coefficients Using Gas Chromatographic Apparatus , 1972 .