Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Optimal acetabular cup orientation is of substantial importance to good long-term function and low complication rates after total hip arthroplasty (THA). The radiographic anteversion (RA) and inclination (RI) angles of the cup are typically studied due to the practicability, simplicity, and ease of interpretation of their measurements. A great number of methods have been developed to date, most of which have been performed on pelvic or hip anteroposterior radiographs. However, there are primarily two influencing factors for these methods: X-ray offset and pelvic rotation. In addition, there are three types of pelvic rotations about the transverse, longitudinal, and anteroposterior axes of the body. Their effects on the RA and RI angles of the cup are interactively correlated with the position and true orientation of the cup. To date, various fitted or analytical models have been established to disclose the correlations between the X-ray offset and pelvic rotation and the RA and RI angles of the cup. Most of these models do not incorporate all the potential influencing parameters. Advanced methods for performing X-ray offset and pelvic rotation corrections are mainly performed on a single pelvic AP radiograph, two synchronized radiographs, or a two-dimensional/three-dimensional (2D-3D) registration system. Some measurement systems, originally developed for evaluating implant migration or wear, could also be used for correcting the X-ray offset and pelvic rotation simultaneously, but some drawbacks still exist with these systems. Above all, the 2D-3D registration technique might be an alternative and powerful tool for accurately measuring cup orientation. In addition to the current methods used for postoperative assessment, navigation systems and augmented reality are also used for the preoperative planning and intraoperative guidance of cup placement. With the continuing development of artificial intelligence and machine learning, these techniques could be incorporated into robot-assisted orthopaedic surgery in the future.

[1]  M. Müller Total hip prostheses. , 1970, Clinical orthopaedics and related research.

[2]  R. McLaren Prosthetic hip angulation. , 1973, Radiology.

[3]  D Resnick,et al.  Evaluation of acetabular anteversion following total hip arthroplasty: necessity of proper centring. , 1975, The British journal of radiology.

[4]  J. Lewis,et al.  Dislocations after total hip-replacement arthroplasties. , 1978, The Journal of bone and joint surgery. American volume.

[5]  J. Visser,et al.  A new method for measuring angles after total hip arthroplasty. A study of the acetabular cup and femoral component. , 1981, The Journal of bone and joint surgery. British volume.

[6]  H. Lindberg,et al.  Radiologic Evaluation of the Position of the Acetabular Component of the Total Hip Prosthesis , 1982, Acta radiologica: diagnosis.

[7]  B F Morrey,et al.  Dislocations after total hip arthroplasty. , 1982, The Journal of bone and joint surgery. American volume.

[8]  R J Miller,et al.  Analysis of version in the acetabular cup. , 1982, Clinical orthopaedics and related research.

[9]  J P Conaty,et al.  Classification and treatment of dislocations of total hip arthroplasty. , 1983, Clinical orthopaedics and related research.

[10]  H. Uhthoff,et al.  Anteversion of the acetabular cup. Measurement of angle after total hip replacement. , 1986, The Journal of bone and joint surgery. British volume.

[11]  G. Selvik,et al.  Space Orientation of Total Hip Prosthesis , 1986, Acta radiologica: diagnosis.

[12]  G. Selvik,et al.  Comparison of Two- and Three-Dimensional Methods for Assessment of Orientation of the Total Hip Prosthesis , 1988, Acta radiologica.

[13]  G. Selvik Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. , 1989, Acta orthopaedica Scandinavica. Supplementum.

[14]  P Benum,et al.  Pelvic Inclination and Spatial Orientation of the Acetabulum , 1990, Acta radiologica.

[15]  B. Sahlstedt,et al.  Influence of Pelvic Position on Radiographic Measurements of the Prosthetic Acetabular Component; An Experimental Study on a Pelvic Model , 1990, Acta radiologica.

[16]  David A. Forsyth,et al.  Invariant Descriptors for 3D Object Recognition and Pose , 1991, IEEE Trans. Pattern Anal. Mach. Intell..

[17]  M. Ferri,et al.  Projective pose estimation of linear and quadratic primitives in monocular computer vision , 1993 .

[18]  D. Murray The definition and measurement of acetabular orientation. , 1993, The Journal of bone and joint surgery. British volume.

[19]  R Bauer,et al.  EBRA: a method to measure migration of acetabular components. , 1995, Journal of biomechanics.

[20]  L. G. Watson,et al.  Radiographic calculation of anteversion in acetabular prostheses. , 1995, The Journal of arthroplasty.

[21]  R B Bourne,et al.  Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique. , 1995, Clinical orthopaedics and related research.

[22]  J. Bahlmann,et al.  Migration of the uncemented Harris-Galante acetabular cup: results of the einbildroentgenanalyse (EBRA) method. , 1997, The Journal of arthroplasty.

[23]  E R Valstar,et al.  Roentgen stereophotogrammetric analysis of metal‐backed hemispherical cups without attached markers , 1997, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[24]  J Kärrholm,et al.  Radiostereometry of hip prostheses. Review of methodology and clinical results. , 1997, Clinical orthopaedics and related research.

[25]  S Berdia,et al.  Determination of Polyethylene Wear in Total Hip Replacements with Use of Digital Radiographs* , 1997, The Journal of bone and joint surgery. American volume.

[26]  P. Devane,et al.  Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris. , 1997, The Journal of arthroplasty.

[27]  T. Ilchmann,et al.  EBRA improves the accuracy of radiographic analysis of acetabular cup migration. , 1998, Acta orthopaedica Scandinavica.

[28]  Lawrence D. Dorr,et al.  Causes of and Treatment Protocol for Instability of Total Hip Replacement , 1998, Clinical orthopaedics and related research.

[29]  B. Jaramaz,et al.  Computer Assisted Measurement of Cup Placement in Total Hip Replacement , 1998, Clinical orthopaedics and related research.

[30]  P. Devane,et al.  Assessment of polyethylene wear in total hip replacement. , 1999, Clinical orthopaedics and related research.

[31]  H. J. Warnecke,et al.  Systematic Geometric Image Measurement Errors of Circular Object Targets: Mathematical Formulation and Correction , 1999 .

[32]  R. Pradhan,et al.  Planar anteversion of the acetabular cup as determined from plain anteroposterior radiographs. , 1999, The Journal of bone and joint surgery. British volume.

[33]  L. Fabeck,et al.  A method to measure acetabular cup anteversion after total hip replacement. , 1999, Acta orthopaedica Belgica.

[34]  L. Dorr,et al.  Radiographic anatomic structure of the arthritic acetabulum and its influence on total hip arthroplasty. , 2000, The Journal of arthroplasty.

[35]  Christophe Tilmant,et al.  Monitoring of polyethylene wear in nonmetal-backed acetubular cups by digitized anteroposterior pelvic radiography , 2003, IEEE Transactions on Medical Imaging.

[36]  G. Saillant,et al.  Hip-spine relationship: a radio-anatomical study for optimization in acetabular cup positioning , 2004, Surgical and Radiologic Anatomy.

[37]  G. Selvik,et al.  Instability and wear of total hip prostheses determined with roentgen stereophotogrammetry , 2004, Archives of orthopaedic and traumatic surgery.

[38]  Karl-Heinz Widmer,et al.  A simplified method to determine acetabular cup anteversion from plain radiographs. , 2004, The Journal of arthroplasty.

[39]  Guoyan Zheng,et al.  C-arm based navigation in total hip arthroplasty-background and clinical experience. , 2004, Injury.

[40]  Alon Wolf,et al.  A kinematic model for calculating cup alignment error during total hip arthroplasty. , 2005, Journal of biomechanics.

[41]  B. Jaramaz,et al.  Measurements of acetabular cup position and pelvic spatial orientation after total hip arthroplasty using computed tomography/radiography matching , 2005, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[42]  F. Langlotz,et al.  Estimation of pelvic tilt on anteroposterior X-rays—a comparison of six parameters , 2006, Skeletal Radiology.

[43]  Frank Langlotz,et al.  Anatomic Referencing of Cup Orientation in Total Hip Arthroplasty , 2005, Clinical orthopaedics and related research.

[44]  Burkhard Lembeck,et al.  Pelvic tilt makes acetabular cup navigation inaccurate , 2005, Acta orthopaedica.

[45]  M Tannast,et al.  Tilt and Rotation Correction of Acetabular Version on Pelvic Radiographs , 2005, Clinical orthopaedics and related research.

[46]  Alon Wolf,et al.  Cup Alignment Error Model for Total Hip Arthroplasty , 2005, Clinical orthopaedics and related research.

[47]  Laurent Sarry,et al.  Radiographic stereometry for non-metal-backed acetubular cups: 3D wear estimation and related uncertainty , 2005, Medical Image Anal..

[48]  Chiou-Shann Fuh,et al.  A New Tool for Measuring Cup Orientation in Total Hip Arthroplasties from Plain Radiographs , 2006, Clinical orthopaedics and related research.

[49]  Lars Perlick,et al.  Position of the acetabular cup -- accuracy of radiographic calculation compared to CT-based measurement. , 2006, European journal of radiology.

[50]  Niclas Börlin,et al.  RSA wear measurements with or without markers in total hip arthroplasty. , 2006, Journal of biomechanics.

[51]  Matthias Wiese,et al.  Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography , 2006, Archives of Orthopaedic and Trauma Surgery.

[52]  David J. Hawkes,et al.  Postoperative Calculation of Acetabular Cup Position Using 2-D–3-D Registration , 2007, IEEE Transactions on Biomedical Engineering.

[53]  Guoyan Zheng,et al.  Precise Estimation of Postoperative Cup Alignment from Single Standard X-Ray Radiograph with Gonadal Shielding , 2007, MICCAI.

[54]  Jae Sung Myung,et al.  Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty , 2008, Skeletal Radiology.

[55]  D. Estok,et al.  Dislocation rate after conversion from hip hemiarthroplasty to total hip arthroplasty. , 2008, The Journal of bone and joint surgery. American volume.

[56]  C. Fuh,et al.  A Simple Mathematical Standardized Measurement of Acetabulum Anteversion after Total Hip Arthroplasty , 2008 .

[57]  C. Lüring,et al.  Cup positioning in THA: current status and pitfalls. A systematic evaluation of the literature , 2009, Archives of Orthopaedic and Trauma Surgery.

[58]  Stephen J. McKenna,et al.  Automated assessment of polyethylene wear in cemented acetabular components using anteroposterior radiographs of total hip replacements , 2008, Comput. Medical Imaging Graph..

[59]  Brian Derbyshire,et al.  Correction of acetabular cup orientation measurements for X-ray beam offset. , 2008, Medical engineering & physics.

[60]  Dohyung Lim,et al.  Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. , 2008, Clinical biomechanics.

[61]  Jens Kowal,et al.  Validation of a new method for determination of cup orientation in THA , 2009, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[62]  B. Jaramaz,et al.  Imaging and Navigation Measurement of Acetabular Component Position in THA , 2009, Clinical orthopaedics and related research.

[63]  Guoyan Zheng,et al.  HipMatch: An object-oriented cross-platform program for accurate determination of cup orientation using 2D-3D registration of single standard X-ray radiograph and a CT volume , 2009, Comput. Methods Programs Biomed..

[64]  Stephen Lyman,et al.  CT Outperforms Radiography for Determination of Acetabular Cup Version after THA , 2009, Clinical orthopaedics and related research.

[65]  A. Kosiyatrakul,et al.  Measurement of acetabular cup anteversion with the circle theorem. , 2009, Journal of the Medical Association of Thailand = Chotmaihet thangphaet.

[66]  Chiou-Shann Fuh,et al.  Measurement of the acetabular cup anteversion on simulated radiographs. , 2009, The Journal of arthroplasty.

[67]  R. Bader,et al.  Accuracy and reliability of different methods to evaluate the acetabular cup version from plain radiographs , 2010, Surgical and Radiologic Anatomy.

[68]  L. Dorr,et al.  Quantification of Pelvic Tilt in Total Hip Arthroplasty , 2010, Clinical orthopaedics and related research.

[69]  S. Bhatnagar,et al.  Cup anteversion in hip resurfacing: validation of EBRA and the presentation of a simple clinical grading system. , 2010, The Journal of arthroplasty.

[70]  B. Jaramaz,et al.  A validation model for measurement of acetabular component position. , 2010, The Journal of arthroplasty.

[71]  A. Paakkala,et al.  Assessment of inter- and intra-observer reliability in the determination of radiographic version and inclination of the cup in metal-on-metal hip resurfacing , 2012, International Orthopaedics.

[72]  T. Schmalzried,et al.  Measuring acetabular component position on lateral radiographs - ischio-lateral method. , 2011, Bulletin of the NYU hospital for joint diseases.

[73]  David Wright,et al.  Acetabular orientation: anatomical and functional measurement , 2012, International Journal of Computer Assisted Radiology and Surgery.

[74]  Guoyan Zheng,et al.  Validation of a statistical shape model-based 2D/3D reconstruction method for determination of cup orientation after THA , 2012, International Journal of Computer Assisted Radiology and Surgery.

[75]  R. Barrack,et al.  The reliability and variation of acetabular component anteversion measurements from cross-table lateral radiographs. , 2011, The Journal of arthroplasty.

[76]  Henrik Malchau,et al.  The John Charnley Award: Risk Factors for Cup Malpositioning: Quality Improvement Through a Joint Registry at a Tertiary Hospital , 2011, Clinical orthopaedics and related research.

[77]  J. Lazenneca,et al.  Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: Measurements reproductibility with EOS imaging system versus conventional radiographies , 2011 .

[78]  R. Barrack,et al.  Does Standing Affect Acetabular Component Inclination and Version After THA? , 2012, Clinical orthopaedics and related research.

[79]  Carolyn Anglin,et al.  A multiplanar radiography method for assessing cup orientation in total hip arthroplasty. , 2012, Journal of biomechanical engineering.

[80]  Y-K Lee,et al.  Reliability and validity of measuring version of the acetabular component. , 2012, The Journal of bone and joint surgery. British volume.

[81]  A. Sautet,et al.  New method for measuring acetabular component positioning with EOS imaging: feasibility study on dry bone , 2012, International Orthopaedics.

[82]  M. Cross,et al.  Measuring Acetabular Component Version After THA: CT or Plain Radiograph? , 2012, Clinical orthopaedics and related research.

[83]  J. Tiberi,et al.  A More Reliable Method to Assess Acetabular Component Position , 2012, Clinical orthopaedics and related research.

[84]  M. Dhillon,et al.  A new method of measuring acetabular cup anteversion on simulated radiographs , 2012, International Orthopaedics.

[85]  Ming-yu Lu,et al.  Reliability and Validity of Measuring Acetabular Component Orientation by Plain Anteroposterior Radiographs , 2013, Clinical orthopaedics and related research.

[86]  P. Foguet,et al.  Assessment of cup orientation in hip resurfacing: a comparison of TraumaCad and computed tomography , 2013, Journal of Orthopaedic Surgery and Research.

[87]  T. Matsushita,et al.  Accuracy and precision of version angle measurements of the acetabular component after total hip arthroplasty. , 2013, The Journal of arthroplasty.

[88]  R. Barrack,et al.  Accuracy of acetabular component position in hip arthroplasty. , 2013, The Journal of bone and joint surgery. American volume.

[89]  C. Fuh,et al.  Computerized ellipse method for measuring acetabular version after total hip replacement – a precision study using synthetic and real radiographs , 2013, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[90]  T. Kobayashi,et al.  An analysis of the best method for evaluating anteversion of the acetabular component after total hip replacement on plain radiographs. , 2014, The bone & joint journal.

[91]  Cup Version Can be Accurately Measured on the False Profile View Radiograph , 2014, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[92]  Y. Tanabe,et al.  Image registration method for assessing 3D hip alignment and implant position during standing posture , 2014 .

[93]  H. Gill,et al.  The relationship between operative and radiographic acetabular component orientation: which factors influence resultant cup orientation? , 2014, The bone & joint journal.

[94]  Validity, reliability and reproducibility of plain radiographic measurements after total hip arthroplasty , 2015, Skeletal Radiology.

[95]  A. Gonzalez Della Valle,et al.  A critical analysis of radiographic factors in patients who develop dislocation after elective primary total hip arthroplasty , 2016, International Orthopaedics.

[96]  Chris R Mellano,et al.  How does pelvic rotation or tilt affect radiographic measurement of acetabular component inclination angle during THA? , 2015, Journal of orthopaedics.

[97]  W. C. Shin,et al.  The reliability and accuracy of measuring anteversion of the acetabular component on plain anteroposterior and lateral radiographs after total hip arthroplasty. , 2015, The bone & joint journal.

[98]  I. Clarke,et al.  Total Hip Prostheses in Standing, Sitting and Squatting Positions: An Overview of Our 8 Years Practice Using the EOS Imaging Technology , 2015, The open orthopaedics journal.

[99]  T. Renkawitz,et al.  Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography , 2017, International Orthopaedics.

[100]  M. Teeter,et al.  Radiostereometric analysis using clinical radiographic views: Validation measuring total hip replacement wear , 2016, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[101]  O. Reikerås,et al.  Cross table lateral radiography for measurement of acetabular cup version. , 2016, Annals of translational medicine.

[102]  T. Renkawitz,et al.  Measuring Acetabular Cup Orientation on Antero-Posterior Radiographs of the Hip after Total Hip Arthroplasty with a Vector Arithmetic Radiological Method. Is It Valid and Verified for Daily Clinical Practice? , 2016, Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren.

[103]  Carolyn Anglin,et al.  Good vs Poor Results After Total Hip Arthroplasty: An Analysis Method Using Implant and Anatomic Parameters With the EOS Imaging System. , 2016, The Journal of arthroplasty.

[104]  N. Boukhelifa,et al.  Comparison of Three-Dimensional Planning-Assisted and Conventional Acetabular Cup Positioning in Total Hip Arthroplasty: A Randomized Controlled Trial. , 2016, The Journal of bone and joint surgery. American volume.

[105]  T. Schwarz,et al.  Central X-ray beam correction of radiographic acetabular cup measurement after THA: an experimental study , 2017, International Journal of Computer Assisted Radiology and Surgery.

[106]  Mark H. Gonzalez,et al.  Validating a Modified Circle Theorem Method for the Measurement of Acetabular Cup Anteversion on Plain Radiography with Intra-Operative Data from Robotic Assisted Total Hip Arthroplasty. , 2016, The Journal of arthroplasty.

[107]  Kai-Nan An,et al.  Radiographic cup anteversion measurement corrected from pelvic tilt. , 2017, Medical engineering & physics.

[108]  M. Cross,et al.  Quantification of Imaging Error in the Measurement of Cup Position: A Cadaveric Comparison of Radiographic and Computed Tomography Imaging. , 2017, Orthopedics.

[109]  M. Teeter,et al.  Effect of Acetabular Position on Polyethylene Liner Wear Measured Using Simultaneous Biplanar Acquisition. , 2017, The Journal of arthroplasty.

[110]  Brian Derbyshire A mathematical method for precisely calculating the radiographic angles of the cup after total hip arthroplasty. , 2017, Medical engineering & physics.

[111]  Joachim Grifka,et al.  Correction of Pelvic Tilt and Pelvic Rotation in Cup Measurement after THA – An Experimental Study , 2017, RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren.

[112]  H. Weinans,et al.  The trigonometric algorithm defining the true three-dimensional acetabular cup orientation: Correlation between measured and calculated cup orientation angles , 2017 .

[113]  P. Devane,et al.  Measuring the migration of the components and polyethylene wear after total hip arthroplasty: BEADS AND SPECIALISED RADIOGRAPHS ARE NOT NECESSARY , 2017, The bone & joint journal.

[114]  J. Vigdorchik,et al.  Undetected Intraoperative Pelvic Movement Can Lead to Inaccurate Acetabular Cup Component Placement during Total Hip Arthroplasty: A Mathematical Simulation Estimating Change in Cup Position , 2017, The Journal of Hip Surgery.

[115]  Ahmed K. Emara,et al.  The Accuracy of Acetabular Component Position Using a Novel Method to Determine Anteversion. , 2017, The Journal of arthroplasty.

[116]  X. Montet,et al.  Standardized fluoroscopy‐based technique to measure intraoperative cup anteversion , 2017, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[117]  K. Tokunaga,et al.  Implant Orientation Measurement After THA Using the EOS X-Ray Image Acquisition System. , 2018, Advances in experimental medicine and biology.

[118]  Guoyan Zheng,et al.  Computer-Aided Orthopaedic Surgery: State-of-the-Art and Future Perspectives. , 2018, Advances in experimental medicine and biology.

[119]  Zhe Zhao,et al.  Three-dimensional orientation and location-dependent varying rules of radiographic angles of the acetabular cup , 2018, International Orthopaedics.

[120]  Guoyan Zheng,et al.  Effect of Pelvic Tilt and Rotation on Cup Orientation in Both Supine and Standing Positions. , 2017, The Journal of arthroplasty.

[121]  B. Penenberg,et al.  Digital Radiography in Total Hip Arthroplasty: Technique and Radiographic Results , 2018, The Journal of bone and joint surgery. American volume.

[122]  Elmar Kotter,et al.  Correct Assessment of Acetabular Component Orientation in Total Hip Arthroplasty From Plane Radiographs. , 2018, The Journal of arthroplasty.

[123]  B. Derbyshire Correction of radiographic measurements of acetabular cup wear for variations in pelvis orientation , 2018, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[124]  Masaaki Matsubara,et al.  A Pilot Study of Augmented Reality Technology Applied to the Acetabular Cup Placement During Total Hip Arthroplasty. , 2018, The Journal of arthroplasty.

[125]  H. Endo,et al.  Accuracy of Cup Positioning With the Computed Tomography-Based Two-dimensional to Three-Dimensional Matched Navigation System: A Prospective, Randomized Controlled Study. , 2018, The Journal of arthroplasty.

[126]  M. Stover,et al.  Quantitative Effect of Pelvic Position on Radiographic Assessment of Acetabular Component Position. , 2018, The Journal of arthroplasty.

[127]  M. Teeter,et al.  Change in Acetabular Cup Orientation From Supine to Standing Position and Its Effect on Wear of Highly Crosslinked Polyethylene. , 2018, The Journal of arthroplasty.

[128]  H. Gill,et al.  Pelvic positioning in the supine position leads to more consistent orientation of the acetabular component after total hip arthroplasty. , 2018, The bone & joint journal.

[129]  L. Dorr,et al.  Late Dislocation Following Total Hip Arthroplasty: Spinopelvic Imbalance as a Causative Factor , 2018, The Journal of bone and joint surgery. American volume.

[130]  M. Mahran,et al.  Accuracy of plain antero-posterior radiographic-based methods for measurement of acetabular cup version , 2018, International Orthopaedics.

[131]  John J Callaghan,et al.  Artificial Intelligence and Machine Learning in Lower Extremity Arthroplasty: A Review. , 2019, The Journal of arthroplasty.

[132]  G. Saka,et al.  A new radiographic acetabular cup anteversion measurement method in total hip arthroplasty: a clinical study , 2019, European Journal of Orthopaedic Surgery & Traumatology.

[133]  D. Beverland,et al.  Operative and radiographic acetabular component orientation in total hip replacement: Influence of pelvic orientation and surgical positioning technique. , 2019, Medical engineering & physics.

[134]  Guoyan Zheng,et al.  Effect of pelvic tilt and rotation on cup orientation in standing anteroposterior radiographs , 2020, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[135]  Changsheng Li,et al.  Evolution and Current Applications of Robot-Assisted Fracture Reduction: A Comprehensive Review , 2019, Annals of Biomedical Engineering.

[136]  H. Weinans,et al.  Lack of consensus on optimal acetabular cup orientation because of variation in assessment methods in total hip arthroplasty: a systematic review , 2019, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[137]  D. Shi,et al.  Femoral and tibial torsion measurements based on EOS imaging compared to 3D CT reconstruction measurements. , 2019, Annals of translational medicine.

[138]  M. Stover,et al.  Are single plane intraoperative and biplanar postoperative radiographic measurements of acetabular cup position the same? , 2020, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[139]  M. Stover,et al.  Intraoperative evaluation of acetabular cup position during anterior approach total hip arthroplasty: are we accurately interpreting? , 2020, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[140]  M. Bradley,et al.  Accuracy of Acetabular Component Positioning Using Computer-assisted Navigation in Direct Anterior Total Hip Arthroplasty , 2019, Cureus.

[141]  E. Auvinet,et al.  Can an Augmented Reality Headset Improve Accuracy of Acetabular Cup Orientation in Simulated THA? A Randomized Trial , 2018, Clinical orthopaedics and related research.

[142]  F. Boettner,et al.  Intraoperative measurement of cup inclination using fluoroscopy requires a correction factor , 2019, Archives of Orthopaedic and Trauma Surgery.

[143]  R. Castelein,et al.  Non‐equivalent Results from Different Anteversion Measurements Methods for the Evaluation of the Acetabular Cup Orientation in Total Hip Arthroplasty , 2019, Orthopaedic surgery.

[144]  Eric J Topol,et al.  High-performance medicine: the convergence of human and artificial intelligence , 2019, Nature Medicine.

[145]  Zhe Zhao,et al.  The synergetic effect of pelvic rotation and X-ray offset on radiographic angles of the acetabular cup , 2019, Medical & Biological Engineering & Computing.

[146]  Jing-xin Zhao,et al.  Erratum to radiographic assessment of the cup orientation after total hip arthroplasty: a literature review. , 2020, Annals of translational medicine.