Implant-bone interface stress distribution in immediately loaded implants of different diameters: a three-dimensional finite element analysis.

PURPOSE To establish a 3D finite element model of a mandible with dental implants for immediate loading and to analyze stress distribution in bone around implants of different diameters. MATERIALS AND METHODS Three mandible models, embedded with thread implants (ITI, Straumann, Switzerland) with diameters of 3.3, 4.1, and 4.8 mm, respectively, were developed using CT scanning and self-developed Universal Surgical Integration System software. The von Mises stress and strain of the implant-bone interface were calculated with the ANSYS software when implants were loaded with 150 N vertical or buccolingual forces. RESULTS When the implants were loaded with vertical force, the von Mises stress concentrated on the mesial and distal surfaces of cortical bone around the neck of implants, with peak values of 25.0, 17.6 and 11.6 MPa for 3.3, 4.1, and 4.8 mm diameters, respectively, while the maximum strains (5854, 4903, 4344 muepsilon) were located on the buccal cancellous bone around the implant bottom and threads of implants. The stress and strain were significantly lower (p < 0.05) with the increased diameter of implant. When the implants were loaded with buccolingual force, the peak von Mises stress values occurred on the buccal surface of cortical bone around the implant neck, with values of 131.1, 78.7, and 68.1 MPa for 3.3, 4.1, and 4.8 mm diameters, respectively, while the maximum strains occurred on the buccal surface of cancellous bone adjacent to the implant neck, with peak values of 14,218, 12,706, and 11,504 microm, respectively. The stress of the 4.1-mm diameter implants was significantly lower (p < 0.05) than those of 3.3-mm diameter implants, but not statistically different from that of the 4.8 mm implant. CONCLUSIONS With an increase of implant diameter, stress and strain on the implant-bone interfaces significantly decreased, especially when the diameter increased from 3.3 to 4.1 mm. It appears that dental implants of 10 mm in length for immediate loading should be at least 4.1 mm in diameter, and uniaxial loading to dental implants should be avoided or minimized.

[1]  Y. Çiftçi,et al.  The effect of veneering materials on stress distribution in implant-supported fixed prosthetic restorations. , 2000, The International journal of oral & maxillofacial implants.

[2]  H. Iplikçioğlu,et al.  Comparative evaluation of the effect of diameter, length and number of implants supporting three-unit fixed partial prostheses on stress distribution in the bone. , 2002, Journal of dentistry.

[3]  K. Akca,et al.  Evaluation of the Effect of the Residual Bone Angulation on Implant-Supported Fixed Prosthesis in Mandibular Posterior Edentulism Part II: 3-D Finite Element Stress Analysis , 2001, Implant dentistry.

[4]  G. Anneroth,et al.  Endosseus titanium implants in extraction sockets , 1985 .

[5]  J. Rubenstein,et al.  Ten-year results for Brånemark implants immediately loaded with fixed prostheses at implant placement. , 1997, The International journal of oral & maxillofacial implants.

[6]  F. Mante,et al.  Evaluating parameters of osseointegrated dental implants using finite element analysis--a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. , 1998, The Journal of oral implantology.

[7]  U C Belser,et al.  Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data. , 2004, Clinical oral implants research.

[8]  Aslihan Usumez,et al.  Three-dimensional finite-element analysis of functional stresses in different bone locations produced by implants placed in the maxillary posterior region of the sinus floor. , 2005, The Journal of prosthetic dentistry.

[9]  B Rangert,et al.  Influence of implant diameters on the integration of screw implants. An experimental study in rabbits. , 1997, International journal of oral and maxillofacial surgery.

[10]  P A Schnitman,et al.  Immediate fixed interim prostheses supported by two-stage threaded implants: methodology and results. , 1990, The Journal of oral implantology.

[11]  S. Hansson,et al.  The implant thread as a retention element in cortical bone: the effect of thread size and thread profile: a finite element study. , 2003, Journal of biomechanics.

[12]  V. Olgac,et al.  Histological and clinical evaluation of 3 different grafting materials for sinus lifting procedure based on 8 cases. , 2001, Journal of periodontology.

[13]  P Missika,et al.  Optimal implant stabilization in low density bone. , 2001, Clinical oral implants research.

[14]  F Bosman,et al.  Loading conditions of endosseous implants in an edentulous human mandible: a three-dimensional, finite-element study. , 1996, Journal of oral rehabilitation.

[15]  Y C Loo,et al.  Application of finite element method in dental implant research , 2009 .

[16]  J B Brunski,et al.  Biomechanics of oral implants: future research directions. , 1988, Journal of dental education.

[17]  H. Hansson,et al.  Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. , 1981, Acta orthopaedica Scandinavica.

[18]  D P Tarnow,et al.  Immediate loading of threaded implants at stage 1 surgery in edentulous arches: ten consecutive case reports with 1- to 5-year data. , 1997, The International journal of oral & maxillofacial implants.

[19]  Svatava Konvicková,et al.  Influence of implant length and diameter on stress distribution: a finite element analysis. , 2004, The Journal of prosthetic dentistry.

[20]  P I Brånemark,et al.  A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. , 1981, International journal of oral surgery.

[21]  H. Haanaes,et al.  A comparison of 3 different endosseous nonsubmerged implants in edentulous mandibles: a clinical report. , 1999, The International journal of oral & maxillofacial implants.

[22]  R A Jaffin,et al.  Immediate loading of implants in partially and fully edentulous jaws: a series of 27 case reports. , 2000, Journal of periodontology.

[23]  T W Korioth,et al.  Modeling the mechanical behavior of the jaws and their related structures by finite element (FE) analysis. , 1997, Critical reviews in oral biology and medicine : an official publication of the American Association of Oral Biologists.

[24]  A. Piattelli,et al.  Immediate loading of titanium plasma-sprayed implants: an histologic analysis in monkeys. , 1998, Journal of periodontology.

[25]  Makoto Shiota,et al.  The influence of implant location and length on stress distribution for three-unit implant-supported posterior cantilever fixed partial dentures. , 2004, The Journal of prosthetic dentistry.

[26]  G. Liu,et al.  Application of finite element analysis in implant dentistry: a review of the literature. , 2001, The Journal of prosthetic dentistry.

[27]  P. Branemark,et al.  Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. , 1990, The International journal of oral & maxillofacial implants.

[28]  James Ra,et al.  Comparative three-dimensional analysis of two finite-element endosseous implant designs. , 1994 .

[29]  D. Buser,et al.  Tissue integration of one-stage ITI implants: 3-year results of a longitudinal study with Hollow-Cylinder and Hollow-Screw implants. , 1991, The International journal of oral & maxillofacial implants.

[30]  O. Miyakawa,et al.  Influence of restoration type on stress distribution in bone around implants: a three-dimensional finite element analysis. , 1998, The International journal of oral & maxillofacial implants.

[31]  T Suetsugu,et al.  Two-dimensional FEM analysis of hydroxyapatite implants: diameter effects on stress distribution. , 1990, The Journal of oral implantology.

[32]  Lum Lb,et al.  A biomechanical rationale for the use of short implants. , 1991 .

[33]  M R Rieger,et al.  Finite element analysis of six endosseous implants. , 1990, The Journal of prosthetic dentistry.

[34]  D Pipko,et al.  Three-dimensional finite element stress analysis in and around the Screw-Vent implant. , 1991, The International journal of oral & maxillofacial implants.

[35]  C. Watson,et al.  Finite element analysis of fixed prostheses attached to osseointegrated implants. , 1990, Quintessence international.

[36]  U. Lekholm Clinical procedures for treatment with osseointegrated dental implants. , 1983, The Journal of prosthetic dentistry.

[37]  N Matsumoto,et al.  Influence of bone quality on the stress distribution. An in vitro experiment. , 1997, Clinical oral implants research.

[38]  J. Kuiper,et al.  Stress distribution around dental implants: influence of superstructure, length of implants, and height of mandible. , 1992, The Journal of prosthetic dentistry.

[39]  Thomas D Taylor,et al.  Twenty years of progress in implant prosthodontics. , 2002, The Journal of prosthetic dentistry.

[40]  A. Scarano,et al.  Immediate loading of titanium plasma-sprayed screw-shaped implants in man: a clinical and histological report of two cases. , 1997, Journal of periodontology.

[41]  A. Barone,et al.  Radiographic bone density around immediately loaded oral implants. , 2003, Clinical oral implants research.

[42]  O. Bahat,et al.  What are the clinical limitations of wide-diameter (4 mm or greater) root-form endosseous implants? , 2000, The International journal of oral & maxillofacial implants.

[43]  Dinçer Bozkaya,et al.  Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. , 2004, The Journal of prosthetic dentistry.

[44]  F. Bosman,et al.  A comparison of three finite element models of an edentulous mandible provided with implants. , 1993, Journal of oral rehabilitation.

[45]  T. Albrektsson,et al.  Tissue regeneration in bony defects adjacent to immediately loaded titanium implants placed into extraction sockets: a study in dogs. , 1997, The International journal of oral & maxillofacial implants.

[46]  Michel Barquins,et al.  Two dental implants designed for immediate loading: a finite element analysis. , 2002, The International journal of oral & maxillofacial implants.

[47]  G. van der Perre,et al.  Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog. , 1998, Journal of oral rehabilitation.

[48]  H Vaillancourt,et al.  Factors affecting crestal bone loss with dental implants partially covered with a porous coating: a finite element analysis. , 1996, The International journal of oral & maxillofacial implants.

[49]  Afonso C. C. Lemonge,et al.  Three-dimensional finite element stress analysis of a cuneiform-geometry implant. , 2003, The International journal of oral & maxillofacial implants.

[50]  G Van der Perre,et al.  The influence of bone mechanical properties and implant fixation upon bone loading around oral implants. , 1998, Clinical oral implants research.