Gated Silicon Drift Detector Fabricated from a Low-Cost Silicon Wafer

Inexpensive high-resolution silicon (Si) X-ray detectors are required for on-site surveys of traces of hazardous elements in food and soil by measuring the energies and counts of X-ray fluorescence photons radially emitted from these elements. Gated silicon drift detectors (GSDDs) are much cheaper to fabricate than commercial silicon drift detectors (SDDs). However, previous GSDDs were fabricated from 10-kΩ·cm Si wafers, which are more expensive than 2-kΩ·cm Si wafers used in commercial SDDs. To fabricate cheaper portable X-ray fluorescence instruments, we investigate GSDDs formed from 2-kΩ·cm Si wafers. The thicknesses of commercial SDDs are up to 0.5 mm, which can detect photons with energies up to 27 keV, whereas we describe GSDDs that can detect photons with energies of up to 35 keV. We simulate the electric potential distributions in GSDDs with Si thicknesses of 0.5 and 1 mm at a single high reverse bias. GSDDs with one gate pattern using any resistivity Si wafer can work well for changing the reverse bias that is inversely proportional to the resistivity of the Si wafer.

[1]  A New N-Channel Junction Field-Effect Transistor Embedded in the i Layer of a Pin Diode , 1999 .

[2]  B. L. Anderson,et al.  Fundamentals of Semiconductor Devices , 2004 .

[3]  A. Vacchi,et al.  Laboratory and test beam results from a large-area silicon drift detector , 2000 .

[4]  H. Matsuura Simulation of Thick Gated Silicon Drift X-ray Detector Operated by a Single High-Voltage Source , 2013 .

[5]  Pavel Rehak,et al.  Semiconductor drift chamber: an application of a novel charge transport scheme , 1984 .

[8]  H. Grubin The physics of semiconductor devices , 1979, IEEE Journal of Quantum Electronics.

[9]  Peter Lechner,et al.  Ultrathin entrance windows for silicon drift detectors , 1997 .

[10]  H. Matsuura,et al.  Simulation of 1.5-mm-thick and 15-cm-diameter gated silicon drift X-ray detector operated with a single high-voltage source , 2015 .

[11]  U. Prechtel,et al.  Spiral silicon drift detectors , 1989 .

[12]  Heike Soltau,et al.  Silicon drift detectors for high resolution room temperature X-ray spectroscopy , 1996 .

[13]  J. Kemmer,et al.  Silicon drift detectors with enlarged sensitive areas , 2004 .

[14]  Peter Holl,et al.  A new silicon drift detector with reduced lateral diffusion , 1996 .

[15]  Hideharu Matsuura,et al.  Simulation and Fabrication of Gated Silicon Drift X-Ray Detector Operated by Peltier Cooling , 2013 .

[16]  C. Fiorini,et al.  A monolithic array of silicon drift detectors for high-resolution gamma-ray imaging , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[17]  Carlo Fiorini,et al.  A new detection system for x-ray microanalysis based on a silicon drift detector with Peltier cooling , 1997 .

[18]  K. Hodgson,et al.  Simulation and modelling of a new silicon X-ray drift detector design for synchrotron radiation applications , 1996 .

[19]  J. Kemmer,et al.  Large area silicon drift detectors for new applications in nuclear medicine imaging , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[20]  H. Soltau,et al.  Novel high-resolution silicon drift detectors , 2004 .

[21]  M. Sampietro,et al.  Large area cylindrical silicon drift detector , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[22]  F. Lazzarotto,et al.  Room-temperature spectroscopic performance of a very-large area silicon drift detector , 2011 .

[23]  I. Eisele,et al.  Epitaxy — a new technology for fabrication of advanced silicon radiation detectors , 2005 .

[24]  C.R. Tull,et al.  High efficiency silicon X-ray detectors , 2004, IEEE Transactions on Nuclear Science.

[25]  K. Hansen,et al.  A novel multicell silicon drift detector module for X-ray spectroscopy and imaging applications , 2000 .

[26]  C. Kenney,et al.  3D — A proposed new architecture for solid-state radiation detectors , 1997 .

[27]  A. Vacchi,et al.  The X-Ray Spectroscopic Performance of a Very Large Area Silicon Drift Detector , 2009, IEEE Transactions on Nuclear Science.

[28]  G. Bertuccio,et al.  Room temperature X-ray spectroscopy with a silicon diode detector and an ultra low noise preamplifier , 1994 .

[29]  Keith O. Hodgson,et al.  Large area silicon drift detectors for X-rays-new results , 1999 .

[30]  Rohde,et al.  High-Resolution X-ray Spectroscopy Close to Room Temperature , 1998, Microscopy and Microanalysis.

[31]  S. Kavadias,et al.  A high resolution, 6 channels, silicon drift detector array with integrated JFET's designed for XAFS spectroscopy: first X-ray fluorescence excitation spectra recorded at the ESRF , 1996 .

[32]  A. Vacchi,et al.  Silicon drift detector with a continuous implanted resistor as divider-drift electrode , 1998 .

[33]  D. Siddons,et al.  Performance of a Thin-Window Silicon Drift Detector X-Ray Fluorescence Spectrometer , 2009, IEEE Transactions on Nuclear Science.