A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system

&NA; AcurosPT is a Monte Carlo algorithm in the Eclipse 13.7 treatment planning system, which is designed to provide rapid and accurate dose calculations for proton therapy. Computational run‐time in minimized by simplifying or eliminating less significant physics processes. In this article, the accuracy of AcurosPT was benchmarked against both measurement and an independent MC calculation, TOPAS. Such a method can be applied to any new MC calculation for the detection of potential inaccuracies. To validate multiple Coulomb scattering (MCS) which affects primary beam broadening, single spot profiles in a Solidwater® phantom were compared for beams of five selected proton energies between AcurosPT, measurement and TOPAS. The spot Gaussian sigma in AcurosPT was found to increase faster with depth than both measurement and TOPAS, suggesting that the MCS algorithm in AcurosPT overestimates the scattering effect. To validate AcurosPT modeling of the halo component beyond primary beam broadening, field size factors (FSF) were compared for multi‐spot profiles measured in a water phantom. The FSF for small field sizes were found to disagree with measurement, with the disagreement increasing with depth. Conversely, TOPAS simulations of the same FSF consistently agreed with measurement to within 1.5%. The disagreement in absolute dose between AcurosPT and measurement was smaller than 2% at the mid‐range depth of multi‐energy beams. While AcurosPT calculates acceptable dose distributions for typical clinical beams, users are cautioned of potentially larger errors at distal depths due to overestimated MCS and halo implementation.

[1]  Matthias Fippel,et al.  A pencil beam algorithm for intensity modulated proton therapy derived from Monte Carlo simulations , 2005, Physics in medicine and biology.

[2]  L Grevillot,et al.  A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4 , 2011, Physics in medicine and biology.

[3]  R. Mohan,et al.  Commissioning dose computation models for spot scanning proton beams in water for a commercially available treatment planning system. , 2013, Medical physics.

[4]  A Mairani,et al.  Dosimetric commissioning and quality assurance of scanned ion beams at the Italian National Center for Oncological Hadrontherapy. , 2015, Medical physics.

[5]  S Safai,et al.  Proton beam monitor chamber calibration , 2014, Physics in medicine and biology.

[6]  R. Mohan,et al.  An MCNPX Monte Carlo model of a discrete spot scanning proton beam therapy nozzle. , 2010, Medical physics.

[7]  C Nauraye,et al.  Monte Carlo modelling of the treatment line of the Proton Therapy Center in Orsay , 2009, Physics in medicine and biology.

[8]  Sjirk N. Boon,et al.  Fast 2D phantom dosimetry for scanning proton beams. , 1998, Medical physics.

[9]  H. Paganetti Range uncertainties in proton therapy and the role of Monte Carlo simulations , 2012, Physics in medicine and biology.

[10]  A. Dell'Acqua,et al.  Geant4 - A simulation toolkit , 2003 .

[11]  Liyong Lin,et al.  Experimentally validated pencil beam scanning source model in TOPAS , 2014, Physics in medicine and biology.

[12]  T. Solberg,et al.  Experimental characterization of two-dimensional spot profiles for two proton pencil beam scanning nozzles , 2014, Physics in medicine and biology.

[13]  H Paganetti,et al.  TOPAS: an innovative proton Monte Carlo platform for research and clinical applications. , 2012, Medical physics.

[14]  Dennis Mah,et al.  Technical Note: Spot characteristic stability for proton pencil beam scanning. , 2016, Medical physics.

[15]  A. Lomax,et al.  Characterizing a proton beam scanning system for Monte Carlo dose calculation in patients , 2015, Physics in medicine and biology.

[16]  B. Schaffner Proton dose calculation based on in-air fluence measurements , 2008, Physics in medicine and biology.

[17]  F. Mourtada,et al.  Feasibility of a multigroup deterministic solution method for three-dimensional radiotherapy dose calculations. , 2008, International journal of radiation oncology, biology, physics.

[18]  S Lorentini,et al.  Characterization and validation of a Monte Carlo code for independent dose calculation in proton therapy treatments with pencil beam scanning , 2015, Physics in medicine and biology.

[19]  E Pedroni,et al.  Experimental characterization and physical modelling of the dose distribution of scanned proton pencil beams , 2005, Physics in medicine and biology.

[20]  O Jäkel,et al.  A calibration procedure for beam monitors in a scanned beam of heavy charged particles. , 2004, Medical physics.

[21]  Liyong Lin,et al.  A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles , 2013, Physics in medicine and biology.

[22]  F Dessy,et al.  Fundamental radiological and geometric performance of two types of proton beam modulated discrete scanning systems. , 2013, Medical physics.

[23]  David Sarrut,et al.  Optimization of GEANT4 settings for Proton Pencil Beam Scanning simulations using GATE , 2010 .

[24]  K Parodi,et al.  A Monte Carlo-based treatment planning tool for proton therapy , 2013, Physics in medicine and biology.

[25]  Radhe Mohan,et al.  Experimental characterization of the low-dose envelope of spot scanning proton beams , 2010, Physics in medicine and biology.

[26]  E Pedroni,et al.  Monte Carlo dose calculations for spot scanned proton therapy , 2005, Physics in medicine and biology.

[27]  K Parodi,et al.  On the parametrization of lateral dose profiles in proton radiation therapy. , 2015, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[28]  Timothy D. Solberg,et al.  Use of a novel two‐dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance , 2015, Journal of applied clinical medical physics.

[29]  Firas Mourtada,et al.  Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beams , 2010, Physics in medicine and biology.

[30]  Martin Bues,et al.  Impact of range shifter material on proton pencil beam spot characteristics. , 2015, Medical physics.

[31]  Alessandra Bolsi,et al.  Treatment planning and verification of proton therapy using spot scanning: initial experiences. , 2004, Medical physics.