Minimizing fouling at hydrogenated conical-tip carbon electrodes during dopamine detection in vivo.

In this paper, physically small conical-tip carbon electrodes (∼2-5 μm diameter and ∼4 μm axial length) were hydrogenated to develop a probe capable of withstanding fouling during dopamine detection in vivo. Upon hydrogenation, the resultant hydrophobic sp(3) carbon surface deters adsorption of amphiphilic lipids, proteins, and peptides present in extracellular fluid and hence minimizes electrode fouling. These hydrogenated carbon electrodes showed a 35% decrease in sensitivity but little change in the limit of detection for dopamine over a 7-day incubation in a synthetic laboratory solution containing 1.0% (v/v) caproic acid (a lipid), 0.1% (w/v) bovine serum albumin and 0.01% (w/v) cytochrome C (both are proteins), and 0.002% (w/v) human fibrinopeptide B (a peptide). Subsequently, during dopamine detection in vivo, over 70% of the dopamine oxidation current remained after the first 30 min of a 60-min experiment, and at least 50% remained over the next half-period at the hydrogenated carbon electrodes. On the basis of these results, an initial average electrode surface fouling rate of 1.2% min(-1) was estimated, which gradually declined to 0.7% min(-1). These results support minimal fouling at hydrogenated carbon electrodes applied to dopamine detection in vivo.

[1]  R. Wightman,et al.  Electrochemical monitoring of biogenic amine neurotransmission in real time. , 1999, Journal of pharmaceutical and biomedical analysis.

[2]  B. Rezek,et al.  Electronic properties of plasma hydrogenated diamond surfaces: A microscopic study , 2006 .

[3]  Wen-Li Jia,et al.  Highly selective and sensitive determination of dopamine using a Nafion/carbon nanotubes coated poly(3-methylthiophene) modified electrode. , 2006, Biosensors & bioelectronics.

[4]  Ralph N. Adams,et al.  Nafion-coated electrodes with high selectivity for CNS electrochemistry , 1984, Brain Research.

[5]  Q. Xue,et al.  Characterization of hydrogenated diamond-like carbon films electrochemically deposited on a silicon substrate , 2004 .

[6]  S. Alwarappan,et al.  Evaluation of hydrogenated physically small carbon electrodes in resisting fouling during voltammetric detection of dopamine , 2007 .

[7]  Junho Choi,et al.  Structural and mechanical properties of DLC films prepared by bipolar PBII&D☆☆ , 2011 .

[8]  R. Wightman,et al.  In vivo comparison of norepinephrine and dopamine release in rat brain by simultaneous measurements with fast‐scan cyclic voltammetry , 2011, Journal of neurochemistry.

[9]  J. Xu,et al.  Anthraquinonedisulfonate electrochemistry:  a comparison of glassy carbon, hydrogenated glassy carbon, highly oriented pyrolytic graphite, and diamond electrodes. , 1998, Analytical chemistry.

[10]  Paul A. Garris,et al.  Wireless transmission of fast-scan cyclic voltammetry at a carbon-fiber microelectrode: proof of principle , 2004, Journal of Neuroscience Methods.

[11]  C. Casiraghi Effect of hydrogen on the UV Raman intensities of diamond-like carbon , 2011 .

[12]  J. Butler,et al.  The structural and electrochemical properties of boron-doped nanocrystalline diamond thin-film electrodes grown from Ar-rich and H2-rich source gases , 2009 .

[13]  A. Andrews,et al.  Boron-Doped Diamond Microelectrodes Reveal Reduced Serotonin Uptake Rates in Lymphocytes from Adult Rhesus Monkeys Carrying the Short Allele of the 5-HTTLPR. , 2010, ACS chemical neuroscience.

[14]  M. Rice,et al.  Chronoamperometry in brain slices: Quantitative evaluations of in vivo electrochemistry , 1983, Brain Research.

[15]  Peter Levine,et al.  Handbook of Biosensors and Biochips (Marks, R.S., et al., Eds.; 2007) [Book Review] , 2009, IEEE Signal Processing Magazine.

[16]  R. Wightman,et al.  Aversive stimulus differentially triggers subsecond dopamine release in reward regions , 2012, Neuroscience.

[17]  J. Musil Microwave plasma: its characteristics and applications in thin film technology , 1986 .

[18]  Danny K.Y. Wong,et al.  Evaluation of physically small p-phenylacetate-modified carbon electrodes against fouling during dopamine detection in vivo , 2013 .

[19]  D. Wong,et al.  An in vivo probe based on mechanically strong but structurally small carbon electrodes with an appreciable surface area. , 2001, Analytical chemistry.

[20]  R. Ramaraj,et al.  Electrochemically deposited nanostructured platinum on Nafion coated electrode for sensor applications , 2005 .

[21]  Chang Ming Li,et al.  Highly sensitive and selective method to detect dopamine in the presence of ascorbic acid by a new polymeric composite film. , 2007, Analytical biochemistry.

[22]  T. Kuwana,et al.  Vacuum heat-treatment for activation of glassy carbon electrodes , 1985 .

[23]  A. Barnard,et al.  Prediction and measurement of the size-dependent stability of fluorescence in diamond over the entire nanoscale. , 2009, Nano letters.

[24]  A. Swift Surface analysis of corrosion inhibitor films by XPS and ToFSIMS , 1995 .

[25]  F. Tajabadi,et al.  Simultaneous determination of dopamine, ascorbic acid, and uric acid using carbon ionic liquid electrode. , 2006, Analytical biochemistry.

[26]  Richard G Compton,et al.  Oxygenated edge plane sites slow the electron transfer of the ferro-/ferricyanide redox couple at graphite electrodes. , 2006, Chemphyschem : a European journal of chemical physics and physical chemistry.

[27]  Jonathan A. Stamford,et al.  Fast cyclic voltammetry: improved sensitivity to dopamine with extended oxidation scan limits , 1990, Journal of Neuroscience Methods.

[28]  E. Vogler,et al.  Structure of hydrogenated diamond like carbon by Micro-Raman spectroscopy , 2012 .

[29]  Philip J. Martin,et al.  The properties of fluorine containing diamond-like carbon films prepared by plasma-enhanced chemical vapour deposition , 2009 .

[30]  Akira Fujishima,et al.  Electroanalysis of dopamine and NADH at conductive diamond electrodes , 1999 .

[31]  Kelly L. Adams,et al.  High osmolarity and L-DOPA augment release via the fusion pore in PC12 cells. , 2007, Chemphyschem : a European journal of chemical physics and physical chemistry.

[32]  J. Galligan,et al.  Diamond microelectrodes for in vitro electroanalytical measurements: current status and remaining challenges. , 2008, The Analyst.

[33]  J. Luong,et al.  Selective nanomolar detection of dopamine using a boron-doped diamond electrode modified with an electropolymerized sulfobutylether-beta-cyclodextrin-doped poly(N-acetyltyramine) and polypyrrole composite film. , 2009, Analytical chemistry.

[34]  J. Lowry,et al.  Brain nitric oxide: Regional characterisation of a real-time microelectrochemical sensor , 2012, Journal of Neuroscience Methods.

[35]  P. Zambonin,et al.  XPS and SIMS surface chemical analysis of some important classes of polymeric biomaterials , 1996 .

[36]  Keith R Hallam,et al.  XPS and Laser Raman Analysis of Hydrogenated Amorphous Carbon Films , 2003 .

[37]  C. Blaha,et al.  Electrochemistry in vivo: Monitoring dopamine release in the brain of the conscious, freely moving rat , 1987, Brain Research Bulletin.

[38]  Ronen Polsky,et al.  Integrated carbon fiber electrodes within hollow polymer microneedles for transdermal electrochemical sensing. , 2011, Biomicrofluidics.

[39]  C. Haynes,et al.  Electroanalytical eavesdropping on single cell communication. , 2011, Analytical chemistry.

[40]  Romyani Goswami,et al.  Transparent polymer and diamond-like hydrogenated amorphous carbon thin films by PECVD technique , 2008 .

[41]  D. Britz,et al.  Diffusion-limited chronoamperometry at conical-tip microelectrodes , 2010 .

[42]  G. Swain,et al.  Structural characterization, electrochemical reactivity, and response stability of hydrogenated glassy carbon electrodes , 1998 .

[43]  Pedram Mohseni,et al.  Development of the Wireless Instantaneous Neurotransmitter Concentration System for intraoperative neurochemical monitoring using fast-scan cyclic voltammetry. , 2009, Journal of neurosurgery.

[44]  Malcolm L. H. Green,et al.  The influence of edge-plane defects and oxygen-containing surface groups on the voltammetry of acid-treated, annealed and “super-annealed” multiwalled carbon nanotubes , 2008 .

[45]  R. Wightman,et al.  Voltammetric detection of 5-hydroxytryptamine release in the rat brain. , 2009, Analytical chemistry.

[46]  Bernard Delmon,et al.  Studies in Surface Science and Catalysis , 1988 .

[47]  S. Kitazawa,et al.  Phasic reward responses in the monkey striatum as detected by voltammetry with diamond microelectrodes , 2011, Neuroscience Research.