Optimization of 3-D MP-RAGE Sequences for Structural Brain Imaging

Abstract An optimized MR sequence for structural three-dimensional brain scans is presented, giving good T1 contrast and excellent white matter/gray matter segmentation. Modification of the usual linear phase encoding order to centric phase encoding restores the contrast loss, which usually occurs after magnetization preparation during the acquisition process when large volumes are imaged. The deleterious effects on the point-spread function are compensated by means of an appropriate k-space filter. RF coil inhomogeneities are corrected by means of shaped excitation pulses. High contrast-to-noise images of the entire brain with 1 mm isotropic resolution can be obtained in 12 min. The contrast-to-noise-ratio is about 100% higher than for sequences based on linear phase encoding.

[1]  S J Riederer,et al.  Centric phase‐encoding order in three‐dimensional MP‐RAGE sequences: Application to abdominal imaging , 1992, Journal of magnetic resonance imaging : JMRI.

[2]  L R Schad,et al.  A comparison of magnetization prepared 3D gradient-echo (MP-RAGE) sequences for imaging of intracranial lesions. , 1996, Magnetic resonance imaging.

[3]  J. Mugler,et al.  Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE) , 1990, Magnetic resonance in medicine.

[4]  J. Mugler,et al.  Optimization of parameter values for complex pulse sequences by simulated annealing: Application to 3D MP‐RAGE imaging of the brain , 1994, Magnetic resonance in medicine.

[5]  S J Riederer,et al.  The importance of phase‐encoding order in ultra‐short TR snapshot MR imaging , 1990, Magnetic resonance in medicine.

[6]  M. L. Wood,et al.  Spoiling of transverse magnetization in steady‐state sequences , 1991, Magnetic resonance in medicine.

[7]  A. Haase,et al.  Snapshot flash mri. applications to t1, t2, and chemical‐shift imaging , 1990, Magnetic resonance in medicine.

[8]  R Deichmann,et al.  Quantification of T1 values by SNAPSHOT-FLASH NMR imaging , 1992 .

[9]  J. Ashburner,et al.  Multimodal Image Coregistration and Partitioning—A Unified Framework , 1997, NeuroImage.

[10]  J. Mugler,et al.  Csf‐suppressed t2‐weighted threel‐dimensional mp‐rage MR imaging , 1995, Journal of magnetic resonance imaging : JMRI.

[11]  J. Mugler,et al.  Shaping the Signal Response during the Approach to Steady State in Three‐Dimensional Magnetization‐Prepared Rapid Gradient‐Echo Imaging Using Variable Flip Angles , 1992, Magnetic resonance in medicine.

[12]  E. McVeigh,et al.  Signal‐to‐noise measurements in magnitude images from NMR phased arrays , 1997 .

[13]  V M Runge,et al.  Clinical comparison of three‐dimensional MP‐RAGE and FLASH techniques for MR imaging of the head , 1991, Journal of magnetic resonance imaging : JMRI.

[14]  T. Parrish,et al.  3-D FLASH imaging using a single surface coil and a new adiabatic pulse, BIR-4. , 1990, Investigative radiology.

[15]  D. Norris,et al.  A simple method of generating variable T1 contrast images using temporally reordered phase encoding , 1990, Magnetic resonance in medicine.

[16]  M. S. Silver,et al.  Highly selective {π}/{2} and π pulse generation , 1984 .

[17]  R. Deichmann,et al.  Fast T1 mapping on a whole‐body scanner , 1999, Magnetic resonance in medicine.

[18]  D Matthaei,et al.  Rapid three-dimensional MR imaging using the FLASH technique. , 1986, Journal of computer assisted tomography.

[19]  J. Mugler,et al.  Rapid three‐dimensional T1‐weighted MR imaging with the MP‐RAGE sequence , 1991, Journal of magnetic resonance imaging : JMRI.