[The effect of high-pass filter circuit on accurate measurement of ST-segment].

The high-pass filter which is designed to overcome the polarization voltage in the ECG measurements can affect the measurement of the ST-segment; but the accuracy of the ST-segment will have an impact on the diagnosis and treatment of heart diseases. This paper used the ECG detection platform to test the effect of the high-pass filter at different frequency and studied the effect of the high-pass filter on the ST segment to ensure proper use of different application modes.