Ultrafast single breath-hold cone-beam CT lung cancer imaging with faster linac gantry rotation.

PURPOSE Lung tumors treated with hypo-fractionated deep-inspiration breath-hold stereotactic body radiotherapy benefit from fast imaging and treatment. Single breath-hold cone-beam-CT (CBCT) could reduce motion artifacts and improve treatment precision. Thus, gantry speed was accelerated to 18°/s, limiting acquisition time to 10-20 s. Image quality, dosimetry and registration accuracy were compared with standard-CBCT (3°/s). METHODS AND MATERIALS For proof-of-concept, image quality was analyzed following customer acceptance tests, CT-dose index measured, and registration accuracy determined with an off-centered ball-bearing-phantom. A lung-tumor patient was simulated with differently shaped tumor-mimicking inlays in a thorax-phantom. Signal-to-noise-ratio, contrast-to-noise-ratio and geometry of the inlays quantified image quality. Dose was measured in representative positions. Registration accuracy was determined with inlays scanned in pre-defined positions. Manual, automatic (clinical software) and objective-automatic (in-house-developed) registration was performed on planning-CT, offsets between results and applied shifts were compared. RESULTS Image quality of ultrafast-CBCT was adequate for high-contrast areas, despite contrast-reduction of ∼80% due to undersampling. Dose-output was considerably reduced by 60-83% in presented setup; variations are due to gantry-braking characteristics. Registration accuracy was maintained better than 1 mm, mean displacement errors were 0.0 ± 0.2 mm with objective-automatic registration. Ultrafast-CBCT showed no significant registration differences to standard-CBCT. CONCLUSIONS This study of first tests with faster gantry rotation of 18°/s showed promising results for ultrafast high-contrast lung tumor CBCT imaging within single breath-hold of 10-20 s. Such fast imaging times, in combination with fast treatment delivery, could pave the way for intra-fractional combined imaging and treatment within one breath-hold phase, and thus mitigate residual motion and increase treatment accuracy and patient comfort. Even generally speaking, faster gantry rotation could set a benchmark with immense clinical impact where time matters most: palliative patient care, general reduction in uncertainty, and increase in patient throughput especially important for emerging markets with high patient numbers.

[1]  F. Lohr,et al.  Towards clinical implementation of ultrafast combined kV-MV CBCT for IGRT of lung cancer , 2016, Strahlentherapie und Onkologie.

[2]  B G Fallone,et al.  A grey-level image alignment algorithm for registration of portal images and digitally reconstructed radiographs. , 1996, Medical physics.

[3]  F. Wilcoxon,et al.  Individual comparisons of grouped data by ranking methods. , 1946, Journal of economic entomology.

[4]  Emiliano Spezi,et al.  Imaging dose from cone beam computed tomography in radiation therapy. , 2015, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[5]  R. Morin,et al.  Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180. , 2018, Medical physics.

[6]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[7]  Steve B. Jiang,et al.  Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation. , 2002, Physics in medicine and biology.

[8]  Frederik Wenz,et al.  Automated ultrafast kilovoltage-megavoltage cone-beam CT for image guided radiotherapy of lung cancer: System description and real-time results. , 2018, Zeitschrift fur medizinische Physik.

[9]  Steve B. Jiang,et al.  The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. , 2007, Medical physics.

[10]  F. Lohr,et al.  Phantom-based evaluation of dose exposure of ultrafast combined kV-MV-CBCT towards clinical implementation for IGRT of lung cancer , 2017, PloS one.

[11]  Frederik Wenz,et al.  Deep Inspiration Breath Hold-Based Radiation Therapy: A Clinical Review. , 2016, International journal of radiation oncology, biology, physics.

[12]  Max Dahele,et al.  Dosimetric impact of the interplay effect during stereotactic lung radiation therapy delivery using flattening filter-free beams and volumetric modulated arc therapy. , 2013, International journal of radiation oncology, biology, physics.

[13]  Volker Steil,et al.  Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  Jan-Jakob Sonke,et al.  Comparison of different strategies to use four-dimensional computed tomography in treatment planning for lung cancer patients. , 2008, International journal of radiation oncology, biology, physics.

[15]  R. Komaki,et al.  The Potential Role of Respiratory Motion Management and Image Guidance in the Reduction of Severe Toxicities Following Stereotactic Ablative Radiation Therapy for Patients with Centrally Located Early Stage Non-Small Cell Lung Cancer or Lung Metastases , 2014, Front. Oncol..

[16]  S M Seltzer,et al.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology. , 2001, Medical physics.

[17]  Rebecca Fahrig,et al.  Optimization of an on-board imaging system for extremely rapid radiation therapy. , 2015, Medical physics.

[18]  F. Lohr,et al.  An offline technique to evaluate residual motion of the diaphragm during deep inspiratory breath-hold from cone-beam CT datasets , 2018, Strahlentherapie und Onkologie.

[19]  F. Lohr,et al.  Multiple breath-hold CBCT for online image guided radiotherapy of lung tumors: simulation with a dynamic phantom and first patient data. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  Marcel van Herk,et al.  Short-term and long-term reproducibility of lung tumor position using active breathing control (ABC). , 2006, International journal of radiation oncology, biology, physics.

[21]  Pietro Mancosu,et al.  Volumetric modulated arc therapy with flattening filter free (FFF) beams for stereotactic body radiation therapy (SBRT) in patients with medically inoperable early stage non small cell lung cancer (NSCLC). , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[22]  Frederik Wenz,et al.  Intra-breath-hold residual motion of image-guided DIBH liver-SBRT: An estimation by ultrasound-based monitoring correlated with diaphragm position in CBCT. , 2018, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.