The influence of shape and sliding distance of femoral head movement loci on the wear of acetabular cups in total hip arthroplasty

Abstract Wear of the polyethylene acetabular component is the most serious threat to the long-term success of total hip replacements (THRs). Greatly reduced wear rates have been reported for unidirectional, compared to multidirectional, articulation in vitro. This study considers the multidirectional motions experienced at the hip joint as described by movement loci of points on the femoral head for individual THR patients. A three-dimensional computer program determined the movement loci of selected points on the femoral head for THR patients and normal subjects using kinematic data obtained from gait analysis. The sizes and shapes of these loci were quantified by their sliding distances and aspect ratios with substantial differences exhibited between individual THR patients. The average sliding distances ranged from 10.0 to 18.1 mm and the average aspect ratios of the loci ranged from 2.5 to 9.2 for the THR patients. Positive correlations were found between wear rate and average sliding distance, the inverse of the average aspect ratio of the loci and the product of the average sliding distance and the inverse of the average aspect ratio of the loci. Patients with a normal hip joint range of motion produce multidirectional motion loci and tend to experience more wear than patients with more unidirectional motion loci. Differing patterns of multidirectional motion at the hip joint for individual THR patients may explain widely differing wear rates in vivo.

[1]  Duncan Dowson,et al.  Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints , 1985 .

[2]  J. Fisher,et al.  Influence of simulator kinematics on the wear of metal-on-metal hip prostheses , 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[3]  H B Skinner,et al.  Pathokinesiology and Total Joint Arthroplasty , 1993, Clinical orthopaedics and related research.

[4]  D. Tabor,et al.  Friction and molecular structure: the behaviour of some thermoplastics , 1972, Proceedings of the Royal Society of London. A. Mathematical and Physical Sciences.

[5]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[6]  Martin Bland,et al.  An Introduction to Medical Statistics , 1987 .

[7]  J H Dumbleton,et al.  Mechanistic and Morphological Origins of Ultra-High Molecular Weight Polyethylene Wear Debris in Total Joint Replacement Prostheses , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[8]  M P Kadaba,et al.  Measurement of lower extremity kinematics during level walking , 1990, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[9]  Brian J. Edwards,et al.  Orientation softening in the deformation and wear of ultra-high molecular weight polyethylene , 1997 .

[10]  S. Smith,et al.  Simplified motion and loading compared to physiological motion and loading in a hip joint simulator , 2000, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[11]  J. Archard Contact and Rubbing of Flat Surfaces , 1953 .

[12]  J. Fisher,et al.  Biological reactions to wear debris in total joint replacement , 2000, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[13]  D. Howie,et al.  A rat model of resorption of bone at the cement-bone interface in the presence of polyethylene wear particles. , 1988, The Journal of bone and joint surgery. American volume.

[14]  W H Harris,et al.  Loci of movement of selected points on the femoral head during normal gait. Three-dimensional computer simulation. , 1996, The Journal of arthroplasty.

[15]  W. Capello,et al.  Periprosthetic Bone Loss in Total Hip Arthroplasty (THA) , 1998 .

[16]  R Baker,et al.  Movement loci of selected points on the femoral head for individual total hip arthroplasty patients using three-dimensional computer simulation. , 2000, The Journal of arthroplasty.

[17]  M. Ritter,et al.  Measurement of polyethylene thickness in total hip arthroplasty. A technique analysis. , 1994, The Journal of arthroplasty.

[18]  M. Semlitsch,et al.  Reactions of the articular capsule to wear products of artificial joint prostheses. , 1977, Journal of biomedical materials research.

[19]  J Fisher,et al.  A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[20]  J. Fisher,et al.  Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys used in metal on metal total hip replacements , 1999, Journal of materials science. Materials in medicine.

[21]  J. K. Lancaster,et al.  MATERIAL-SPECIFIC WEAR MECHANISMS : RELEVANCE TO WEAR MODELLING , 1990 .

[22]  V. Saikko,et al.  Type of motion and lubricant in wear simulation of polyethylene acetabular cup , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[23]  C. Bragdon,et al.  The Importance of Multidirectional Motion on the Wear of Polyethylene , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[24]  R. Brooks,et al.  The effects of the concentration of high-density polyethylene particles on the bone-implant interface. , 2000, The Journal of bone and joint surgery. British volume.

[25]  A. Wang,et al.  The significance of nonlinear motion in the wear screening of orthopaedic implant materials , 1997 .