Design of a Class-E Transcutaneous Energy Transmitter for an Implantable System

Purpose – The purpose of this paper is to develop a model of the inductive link for implantable systems. The model is suitable for a cochlear implant in which a lateral misalignment and distance coil can be up to 16 mm.Design/methodology/approach – The description of the generation of implantable systems' high‐power, such as a cochlear implant, are powered by transcutaneous inductive power links formed by two coils: the first is a printed spiral coil used in the receiver device and the second is a solenoid coil used in the emitter device. Optimizing the power efficiency of the wireless link is imperative to minimize the size of the external energy source, heating dissipation in the tissue, and interference with other devices. The authors have outlined the theoretical foundation of optimal power transmission efficiency in an inductive link, and combined it with semi‐empirical models to predict parasitic components. The power amplifier itself is a class‐E amplifier optimized in both output voltage and effic...

[1]  P.R. Troyk,et al.  Inductive links and drivers for remotely-powered telemetry systems , 2000, IEEE Antennas and Propagation Society International Symposium. Transmitting Waves of Progress to the Next Millennium. 2000 Digest. Held in conjunction with: USNC/URSI National Radio Science Meeting (C.

[2]  R. White,et al.  A Wide-Band Efficient Inductive Transdennal Power and Data Link with Coupling Insensitive Gain , 1987, IEEE Transactions on Biomedical Engineering.

[3]  Maysam Ghovanloo,et al.  Design and Optimization of Printed Spiral Coils for Efficient Transcutaneous Inductive Power Transmission , 2007, IEEE Transactions on Biomedical Circuits and Systems.

[4]  Maysam Ghovanloo,et al.  A wideband frequency-shift keying wireless link for inductively powered biomedical implants , 2004, IEEE Transactions on Circuits and Systems I: Regular Papers.

[5]  Paul Jespers The gm/ID Methodology, a sizing tool for low-voltage analog CMOS Circuits: The semi-empirical and compact model approaches , 2009 .

[6]  No Sokal,et al.  CLASS-E - NEW CLASS OF HIGH-EFFICIENCY TUNED SINGLE-ENDED SWITCHING POWER AMPLIFIERS , 1975 .

[7]  Mondher Chaoui An Optimal Geometry for Power Energy and Data Transmission in the Inductive Link , 2001 .

[8]  Maysam Ghovanloo,et al.  A modular 32-site wireless neural stimulation microsystem , 2004 .

[9]  Maysam Ghovanloo,et al.  Fully integrated power supply design for wireless biomedical implants , 2002, 2nd Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine and Biology. Proceedings (Cat. No.02EX578).

[10]  Marian K. Kazimierczuk,et al.  Exact analysis of class E tuned power amplifier at any Q and switch duty cycle , 1987 .

[11]  P.R. Troyk,et al.  Closed-loop class E transcutaneous power and data link for MicroImplants , 1992, IEEE Transactions on Biomedical Engineering.

[12]  Richard Perdriau,et al.  Reducing Load Effects in High-energy, High-efficiency Inductive Links , 2006 .

[13]  C. Zierhofer,et al.  Electronic design of a cochlear implant for multichannel high-rate pulsatile stimulation strategies , 1995 .

[14]  Frederick H. Raab,et al.  Idealized operation of the class E tuned power amplifier , 1977 .

[15]  C.M. Zierhofer,et al.  Geometric approach for coupling enhancement of magnetically coupled coils , 1996, IEEE Transactions on Biomedical Engineering.

[16]  M. Ghovanloo,et al.  Fully integrated wideband high-current rectifiers for inductively powered devices , 2004, IEEE Journal of Solid-State Circuits.

[17]  Frederick H. Raab,et al.  Effects of circuit variations on the class E tuned power amplifier , 1978 .

[18]  N. de N. Donaldson,et al.  Analysis of resonant coupled coils in the design of radio frequency transcutaneous links , 1983, Medical and Biological Engineering and Computing.

[19]  F. C. Flack,et al.  Mutual inductance of air-cored coils: Effect on design of radio-frequency coupled implants , 1971, Medical and biological engineering.

[20]  Maysam Ghovanloo,et al.  An Integrated Full-Wave CMOS Rectifier With Built-In Back Telemetry for RFID and Implantable Biomedical Applications , 2008, IEEE Transactions on Circuits and Systems I: Regular Papers.

[21]  C.M. Zierhofer,et al.  High-efficiency coupling-insensitive transcutaneous power and data transmission via an inductive link , 1990, IEEE Transactions on Biomedical Engineering.