A discussion on the threshold limit for hemolysis related to Reynolds shear stress.

Turbulence-related damage to blood is a major problem with the use of prosthetic devices, such as mechanical heart valves. An often-cited paper by Sallam and Hwang (1984). Biorheology 21, 783-797) quantified the threshold for hemolysis to be about 400 N m(-2), a value that has hitherto contributed to the evaluation of the potential dangerousness of a medical implantable device. We propose a discussion of the mentioned experiment, based on the application of stress analysis concepts to the original measurements: this is necessary to assess the peak turbulence shear stress value that could have been found in Sallam and Hwangs experiment, with a suitable orientation of the measurement axes. The result of our theoretical discussion is that the threshold value of 400 N m(-2) could probably be considerably underestimated: following this point of view, a 3-D stress analysis shows that the peak turbulence shear stress at the inception of hemolysis should be at least 600 N m(-2). This result, obtained on the basis of the study of RBCs' response to a turbulent environment, indicates that blood particles are probably more resistant to short-time shear stresses than it was thought.

[1]  H Nygaard,et al.  Turbulent stress measurements downstream of six mechanical aortic valves in a pulsatile flow model. , 1988, Journal of biomechanics.

[2]  L. J. Wurzinger,et al.  Mechanical bloodtrauma. An overview , 1986 .

[3]  M Grigioni,et al.  19 mm Sized Bileaflet Valve Prostheses’ flow Field Investigated by Bidimensional Laser Doppler Anemometry (part II: Maximum Turbulent Shear Stresses) , 1997, The International journal of artificial organs.

[4]  L. Edmunds,et al.  Thromboembolic complications of current cardiac valvular prostheses. , 1982, The Annals of thoracic surgery.

[5]  J D Hellums,et al.  Red blood cell damage by shear stress. , 1972, Biophysical journal.

[6]  L. E. Malvern Introduction to the mechanics of a continuous medium , 1969 .

[7]  P K Paulsen,et al.  Estimation of turbulent shear stresses in pulsatile flow immediately downstream of two artificial aortic valves in vitro. , 1990, Journal of biomechanics.

[8]  K V Arom,et al.  Ten-year follow-up study of patients who had double valve replacement with the St. Jude Medical prosthesis. , 1989, The Journal of thoracic and cardiovascular surgery.

[9]  H Schmid-Schönbein,et al.  Platelet and Coagulation Parameters Following Millisecond Exposure to Laminar Shear Stress , 1985, Thrombosis and Haemostasis.

[10]  G. Batchelor,et al.  The theory of axisymmetric turbulence , 1946, Proceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences.

[11]  A P Yoganathan,et al.  In vitro flow dynamics of four prosthetic aortic valves: a comparative analysis. , 1989, Journal of biomechanics.

[12]  A. Yoganathan,et al.  A 6 year experience with the St. Jude medical valve: hemodynamic performance, surgical results, biocompatibility and follow-up. , 1985, Journal of the American College of Cardiology.

[13]  H Reul,et al.  In-vitro wall shear measurements at aortic valve prostheses. , 1984, Journal of biomechanics.

[14]  S P Sutera,et al.  Deformation and fragmentation of human red blood cells in turbulent shear flow. , 1975, Biophysical journal.

[15]  H N Sabbah,et al.  Measured Turbulence and Its Effect on Thrombus Formation , 1974, Circulation research.

[16]  N H Hwang,et al.  Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. , 1984, Biorheology.

[17]  J. Hinze Fundamentals of the hydrodynamic mechanism of splitting in dispersion processes , 1955 .