Continuous flow phantom for the calibration of an ultrasonic transit-time flowmeter

INTRODUCTION: Ultrasound Transit-Time flowmeters are based on the fact that the time required for an ultrasound pulse to propagate through a given distance in a moving medium is a function of the vectorial sum of pulse propagation velocity and medium velocity. The most common application of this flowmeter in medicine is in the evaluation of blood flow in arteries and veins during heart vascular surgery. The present article describes the design, construction and evaluation of a flow phantom for transit-time flowmeters calibration. METHODS: Basically, it is a hydraulic circuit containing degassed and distilled water. In such a circuit, a constant differential water level is established between two columns that are interconnected by tubes with defined resistance, which determines a known flow rate. A basic theoretical model to estimate the system Reynolds Number and resistance was developed. RESULTS: A flow range between 4.43 ± 0.18 ml.min-1 and 106.88 ± 0.27 ml.min-1 was found to be compatible with physiological values in small vessels. The pressure range was between 0.20 ± 0.03 cmH2O and 12.53 ± 0.07 cmH2O, and the larger Reynolds Number was 1134.07. Experimental and theoretical resistance values were similar. CONCLUSION: A reproducible phantom was designed and built to be assembled with standard low-cost materials and is capable of generating adjustable and continuous flows that can be used to calibrate TTFM systems.

[1]  R. Groom,et al.  Intra-operative quality assessment of coronary artery bypass grafts , 2001, Perfusion.

[2]  D. A. Christopher,et al.  A wall-less vessel phantom for Doppler ultrasound studies. , 1995, Ultrasound in medicine & biology.

[3]  W. O’Brien,et al.  A flexible blood flow phantom capable of independently producing constant and pulsatile flow with a predictable spatial flow profile for ultrasound flow measurement validations , 1992, IEEE Transactions on Biomedical Engineering.

[4]  Colin Berry,et al.  Comparison of Intravascular Ultrasound and Quantitative Coronary Angiography for the Assessment of Coronary Artery Disease Progression , 2007, Circulation.

[5]  O. Hess,et al.  Transit time flow measurement: experimental validation and comparison of three different systems. , 2000, The Annals of thoracic surgery.

[6]  E. Mattsson,et al.  Volume blood flow measurements with a transit time flowmeter: an in vivo and in vitro variability and validation study. , 1993, Clinical physiology.

[7]  J. Bednarik,et al.  Evaluation of a transit-time system for the chronic measurement of blood flow in conscious sheep. , 1995, Journal of applied physiology.

[8]  M. L. Sanderson,et al.  Guidelines for the use of ultrasonic non-invasive metering techniques , 2002 .

[9]  Hristo N Nikolov,et al.  A thin-walled carotid vessel phantom for Doppler ultrasound flow studies. , 2004, Ultrasound in medicine & biology.

[10]  O. Hess,et al.  Transit-time flow measurement for detection of early graft failure during myocardial revascularization. , 1998, The Annals of thoracic surgery.

[11]  D. Heisey,et al.  Clinical importance of measuring coronary graft flows in the revascularized heart. Ultrasonic or electromagnetic? , 1997, The Journal of cardiovascular surgery.

[12]  H. Spotnitz,et al.  Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements. , 1996, ASAIO journal.

[13]  M K Pugsley,et al.  Methods of blood flow measurement in the arterial circulatory system. , 2000, Journal of pharmacological and toxicological methods.

[14]  K V Ramnarine,et al.  Validation of a new blood-mimicking fluid for use in Doppler flow test objects. , 1998, Ultrasound in medicine & biology.

[15]  Y. Nakao,et al.  An evaluation of the intraoperative transit time measurements of coronary bypass flow. , 2001, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[16]  I. Gordon Effects of stenosis on transit-time ultrasound measurements of blood flow. , 1995, Ultrasound in medicine & biology.

[17]  L. Michelini,et al.  Validation of transit-time flowmetry for chronic measurements of regional blood flow in resting and exercising rats. , 1997, Brazilian journal of medical and biological research = Revista brasileira de pesquisas medicas e biologicas.