Effects of the Metallic Port in Tissue Expanders on Dose Distribution in Postmastectomy Radiotherapy: A Tridimensional Experimental Model of Dosimetry in Breast Reconstruction

Abstract The purpose of this research was to develop an experimental model of dosimetry using a breast phantom and evaluate the effects of the metallic port in tissue expanders on dose distribution in postmastectomy radiotherapy. Dose distribution was assessed using an experimental acrylic dosimetry simulator in the absence and presence of a metallic disc (MD), which is similar to that used in tissue expanders containing a magnetic port, by collecting coronal and sagittal radiochromic films. Dosimetry film analysis did not show any changes in dose distribution, except for a MD shadow in the sagittal film where the dose distribution rate was on average 14% lower than in other areas. This model very closely resembled actual breast reconstruction with a tissue expander containing a magnetic port. Scattering or attenuation of the irradiation dose due to interference of the MD was not evident in areas that could jeopardize the effectiveness of radiation therapy. Therefore, the presence of the MD inside the tissue expander does not likely have an impact on radiotherapy effectiveness during immediate breast reconstruction.

[1]  David Odgers,et al.  In vivo dosimetric impact of breast tissue expanders on post‐mastectomy radiotherapy , 2016, Journal of medical imaging and radiation oncology.

[2]  Victor G. L. Alves,et al.  Evaluation of the Radiotherapy Treatment Planning in the Presence of a Magnetic Valve Tissue Expander , 2015, PloS one.

[3]  T. Fitzgerald,et al.  Dosimetric impact of the AeroForm tissue expander in postmastectomy radiation therapy: an ex vivo analysis. , 2015, Practical radiation oncology.

[4]  A. Dal Cin,et al.  Does the presence of an implant including expander with internal port alter radiation dose? An ex vivo model , 2013, The Canadian journal of plastic surgery = Journal canadien de chirurgie plastique.

[5]  A. Losken,et al.  Impact of internal metallic ports in temporary tissue expanders on postmastectomy radiation dose distribution. , 2013, International journal of radiation oncology, biology, physics.

[6]  S. Cardoso,et al.  Influence of the Presence of Tissue Expanders on Energy Deposition for Post-Mastectomy Radiotherapy , 2013, PloS one.

[7]  P. Karaiskos,et al.  Dose perturbation in the radiotherapy of breast cancer patients implanted with the Magna‐Site: a Monte Carlo study , 2011, Journal of applied clinical medical physics.

[8]  P. Cordeiro,et al.  Do metallic ports in tissue expanders affect postmastectomy radiation delivery? , 2006, International journal of radiation oncology, biology, physics.

[9]  J. Dewar Postmastectomy radiotherapy. , 2006, Clinical oncology (Royal College of Radiologists (Great Britain)).

[10]  A. Morgan,et al.  Investigation into dosimetric effect of a MAGNA-SITE™ tissue expander on post-mastectomy radiotherapy. , 2005, Medical physics.

[11]  K. Griffith,et al.  Dosimetry around metallic ports in tissue expanders in patients receiving postmastectomy radiation therapy: an ex vivo evaluation. , 2004, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[12]  Jong-Hyeon Jeong,et al.  Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer. , 2002, The New England journal of medicine.