Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator images.

A new fast method is presented for the quantification of patient set-up errors during radiotherapy with external photon beams. The set-up errors are described as deviations in relative position and orientation of specified anatomical structures relative to specified field shaping devices. These deviations are determined from parameters of the image transformations that make their features in a portal image align with the corresponding features in a simulator image. Knowledge of some set-up parameters during treatment simulation is required. The method does not require accurate knowledge about the position of the portal imaging device as long as the positions of some of the field shaping devices are verified independently during treatment. By applying this method, deviations in a pelvic phantom set-up can be measured with a precision of 2 mm within 1 minute. Theoretical considerations and experiments have shown that the method is not applicable when there are out-of-plane rotations larger than 2 degrees or translations larger than 1 cm. Inter-observer variability proved to be a source of large systematic errors, which could be reduced by offering a precise protocol for the feature alignment.

[1]  R L Siddon,et al.  Solution to treatment planning problems using coordinate transformations. , 1981, Medical physics.

[2]  M van Herk,et al.  Radiation field edge detection in portal images. , 1991, Physics in medicine and biology.

[3]  J Bijhold,et al.  A method for the measurement of field placement errors in digital portal images. , 1990, Physics in medicine and biology.

[4]  M van Herk,et al.  A matrix ionisation chamber imaging device for on-line patient setup verification during radiotherapy. , 1988, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  M. Goitein,et al.  Accuracy of radiation field alignment in clinical practice. , 1985, International journal of radiation oncology, biology, physics.

[6]  M van Herk,et al.  First clinical experience with a newly developed electronic portal imaging device. , 1990, International journal of radiation oncology, biology, physics.

[7]  A G Visser,et al.  Accuracy in radiation field alignment in head and neck cancer: a prospective study. , 1988, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  Robert J. Shalek,et al.  Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures , 1977 .