A retrospective study on perfusion incidents and safety devices

Despite the acceptance of extracorporeal circulation as an effective modality to facilitate cardiac surgery, patient outcomes can be negatively influenced by the occurrence of perfusion incidents. A perfusion survey was conducted to identify safety techniques and incidents related to cardiopulmonary bypass (CPB). An 80-question survey was mailed to chief perfusionists of all 1030 USA cardiac surgical centers using CPB. The survey was designed to examine practices and incidents that occurred during a 2-year period (July 1996 to July 1998). Five-hundred-and-fifty-two (54% response rate) surveys were returned, which accounted for 797 hospitals (79% of all cardiac centers) and 653 621 surgical procedures. Of the 27 identified CPB safety devices, the highest utilization was arterial line filters (98.5%) and the lowest arterial line bubble traps (3.4%). Of the reported cases, a CPB incident occurred once every 138 cases. The most common occurring incidents were protamine reactions (1:783), coagulation problems (1:771), and heater/cooler failures (1:1809). The rate of occurrence of an incident resulting in a serious injury or death was one for every 1453 procedures. Although techniques and safety devices create a relatively secure environment for CPB, lower incident rates may be achieved with further improvements in coagulation monitoring and incident reporting.

[1]  D. Hett,et al.  A survey of priming solutions used for cardiopulmonary bypass , 1994, Perfusion.

[2]  C. S. Thomas,et al.  Air embolism and other accidents using pump oxygenators. , 1980, The Annals of thoracic surgery.

[3]  J. Simpson,et al.  Australasian perfusion incident survey , 1997, Perfusion.

[4]  M. Mack,et al.  A survey on minimally invasive coronary artery bypass grafting. , 1997, The Annals of thoracic surgery.

[5]  G. Schwarz,et al.  Myocardial protection for acquired heart disease surgery: results of a national survey. , 1995, The Annals of thoracic surgery.

[6]  M. Kurusz Lessons from perfusion surveys , 1997, Perfusion.

[7]  D. W. Miller,et al.  Results of a survey of the professional activities of 811 cardiopulmonary perfusionists. , 1982, The Journal of thoracic and cardiovascular surgery.

[8]  A. Stammers,et al.  Evaluation of a new generation cardioplegia administration system. , 1997, The journal of extra-corporeal technology.

[9]  B. Butler,et al.  Air embolism during cardiopulmonary bypass , 1995, Perfusion.

[10]  D A Palanzo,et al.  Perfusion safety: past, present, and future. , 1997, Journal of cardiothoracic and vascular anesthesia.

[11]  D. Birnbaum,et al.  Heparin-coated equipment reduces the risk of oxygenator failure. , 1998, The Annals of thoracic surgery.

[12]  D. Reich,et al.  Cardiopulmonary bypass for adult patients: a survey of equipment and techniques. , 1995, Journal of Cardiothoracic and Vascular Anesthesia.

[13]  D. Wheeldon Can cardiopulmonary bypass be a safe procedure , 1981 .

[14]  R. Groom,et al.  Paediatric perfusion practice in North America: an update , 1995, Perfusion.

[15]  A. Stammers,et al.  Evaluation of a variable ratio cardioplegia system. , 1996, The journal of extra-corporeal technology.