Estimation of current leakage in left and right ventricular conductance volumetry using a dynamic finite element model

Leakage of electric current through cardiac structures surrounding the ventricle is a primary source of error during ventricular volume measurements using a conductance catheter. This error can be represented as a leakage volume, V/sub L/. V/sub L/ is generally estimated by a saline-bolus method, and is assumed constant throughout the cardiac cycle. However, dynamic changes in ventricular volume and cardiac wall thickness could change V/sub L/. To estimate V/sub L/, a dynamic finite element model of the heart was developed based on MR images. Conductance measurements were simulated using a modeled conductance catheter, and true V/sub L/ was calculated. V/sub L/ varied from 22.7 ml (end-systole) to 26.4 ml (end-diastole) in the left ventricle and from 19.9 ml (end-systole) to 26.9 ml (end-diastole) in the right ventricle. The saline-bolus method underestimated V/sub L/ in both the left (V/sub L/=19.4 ml) and the right (V/sub L/=4.1 ml) ventricular volume measurements. V/sub L/ increased linearly with the ratio of blood to tissue resistivity, and changed minimally with catheter position. These results indicate that V/sub L/ has to be estimated dynamically throughout the cardiac cycle to obtain accurate cardiac volume measurements. The results also show that the saline bolus method does not estimate current leakage accurately, especially in the right ventricular volume measurement.

[1]  J. Petre,et al.  A scheme to correct for electric field non-uniformity in right ventricular volume measurement using a conductance catheter , 1998, Computers in Cardiology 1998. Vol. 25 (Cat. No.98CH36292).

[2]  B Gopakumaran,et al.  Real-time continuous measurement of right ventricular volume using a conductance catheter. , 1996, Biomedical instrumentation & technology.

[3]  L. Benson,et al.  Measurement of right ventricular volume by conductance catheter in closed-chest pigs. , 1995, The American journal of physiology.

[4]  M. Valentinuzzi,et al.  Volume profiles obtained by a conductimetric method. , 1993, Journal of biomedical engineering.

[5]  S. Kun,et al.  Conductance volumetric model of an eccentrically positioned catheter within a three-compartment ellipsoidal ventricle , 1993, IEEE Transactions on Biomedical Engineering.

[6]  H. Spotnitz,et al.  Factors confounding impedance catheter volume measurements in vitro. , 1993, The Annals of thoracic surgery.

[7]  C. Bertram,et al.  Detecting Right Ventricular Volume Changes Using the Conductance Catheter , 1992, Pacing and clinical electrophysiology : PACE.

[8]  M. Lipton,et al.  Left ventricular segmental volume by conductance catheter and Cine-CT. , 1992, European heart journal.

[9]  A A Shoukas,et al.  Does volume catheter parallel conductance vary during a cardiac cycle? , 1990, The American journal of physiology.

[10]  W. Little,et al.  Simultaneous conductance catheter and dimension assessment of left ventricle volume in the intact animal. , 1990, Circulation.

[11]  S A Glantz,et al.  Left ventricular volume measurement by conductance catheter in intact dogs. Parallel conductance volume depends on left ventricular size. , 1989, Circulation.

[12]  C. Bertram,et al.  Right Ventricular Volumetry by Catheter Measurement of Conductance , 1987, Pacing and clinical electrophysiology : PACE.

[13]  D. Kass,et al.  Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique. , 1986, Circulation.

[14]  Rodney W. Salo,et al.  Measurement of Ventricular Volume by Intracardiac Impedance: Theoretical and Empinrcal Approaches , 1986, IEEE Transactions on Biomedical Engineering.

[15]  D. Kass,et al.  Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts. , 1985, Circulation.

[16]  B Buis,et al.  Continuous measurement of left ventricular volume in animals and humans by conductance catheter. , 1984, Circulation.

[17]  Jan Baan,et al.  Computation of the Input Impedances of a Catheter for Cardiac Volumetry , 1984, IEEE Transactions on Biomedical Engineering.

[18]  G V Heller,et al.  Instantaneous measurement of left and right ventricular stroke volume and pressure-volume relationships with an impedance catheter. , 1984, Circulation.

[19]  E. T. van der Velde,et al.  Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter. , 1981, Cardiovascular research.

[20]  H. C. Burger,et al.  Specific electric resistance of body tissues. , 1961, Physics in medicine and biology.

[21]  B. Gopakumaran Finite element based techniques for improved measurement of right ventricular volume using a conductance catheter , 1997 .

[22]  Francis A. Duck,et al.  Electrical Properties of Tissue , 1990 .