Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension

Background Four oscillometric devices for self-measurement of blood pressure (SBPM) were evaluated according to the European Society of Hypertension (ESH) international protocol and its 2010 revision in four separate studies. The Omron® M2, Omron M3, and Omron M6 measure blood pressure (BP) at the brachial level, while the Omron R2 measures BP at the wrist level. Methods The international protocol requires a total number of 33 subjects in which the validation is performed. The Omron M2 and Omron R2 were validated in 2009 according to the ESH international protocol, while the Omron M3 and Omron M6 were validated in 2010–2011 according to the 2010 ESH international protocol revision. The protocol procedures were followed precisely. Results All four tested devices passed the validation process. The mean differences between the device and mercury readings were 2.7 ± 5.0 and −1.4 ± 3.2 mmHg for systolic and diastolic BP, respectively, using the Omron M2 device, and 1.7 ± 3.2 and −0.9 ± 2.6 mmHg using the Omron M3, 1.6 ± 2.9 and −0.9 ± 2.5 mmHg using the Omron M6, and −1.1 ± 4.8 and −0.9 ± 4.3 mmHg using the Omron R2. Conclusion Readings from the Omron M2, Omron M3, Omron M6, and Omron R2, differing by less than 5, 10, and 15 mmHg, fulfill the ESH international protocol and its 2010 revision requirements. Therefore, each of these four devices can be used by patients for SBPM.

[1]  Michael Böhm,et al.  2013 ESH/ESC Guidelines for the management of arterial hypertension: the Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). , 2007, Journal of hypertension.

[2]  P. Kirchhof,et al.  The task force for the management ofarterial hypertension of the european society ofhypertension (esh) and of the european society of cardiology (esc) , 2013 .

[3]  Yutaka Imai,et al.  European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults , 2010, Blood pressure monitoring.

[4]  Yutaka Imai,et al.  European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring , 2008, Journal of hypertension.

[5]  G. Stergiou,et al.  Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol , 2007, Blood pressure monitoring.

[6]  A. Dominiczak,et al.  2007 Guidelines for the Management of Arterial Hypertension: The Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC) , 2007, European heart journal.

[7]  G. Stergiou,et al.  Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol , 2006 .

[8]  Benjamin J. Epstein,et al.  Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure , 2007 .

[9]  R. Asmar,et al.  Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) , 2006, Blood pressure monitoring.

[10]  L. Geddes,et al.  Characterization of the oscillometric method for measuring indirect blood pressure , 2006, Annals of Biomedical Engineering.

[11]  Yutaka Imai,et al.  Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. , 2005, Journal of hypertension.

[12]  G. Reboldi,et al.  Validation of the A&D UA-774 (UA-767Plus) device for self-measurement of blood pressure , 2004, Blood pressure monitoring.

[13]  A. Nolly,et al.  Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol , 2004, Blood pressure monitoring.

[14]  B. Alpert Validation of the Pharma-Smart PS-2000 public use blood pressure monitor , 2004, Blood pressure monitoring.

[15]  Daniel W. Jones,et al.  Seventh report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure. , 2003, Hypertension.

[16]  R. Asmar,et al.  Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT , 2003, Blood pressure monitoring.

[17]  Daniel W. Jones,et al.  The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report. , 2003, JAMA.

[18]  E. O'Brien The Working Group on Blood Pressure Monitoring of the European Society of Hypertension. , 2003, Blood pressure monitoring.

[19]  A. Shennan,et al.  Inflationary oscillometric blood pressure monitoring: validation of the OMRON-MIT , 2002, Blood pressure monitoring.

[20]  J. Mallion,et al.  Clinical evaluation of a self blood pressure monitor according to the First International Consensus Conference on Self Blood Pressure Measurement , 2002, Blood pressure monitoring.

[21]  A. Shennan,et al.  Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol , 2002, Blood pressure monitoring.

[22]  R. Asmar,et al.  Validation of the Omron HEM-907 device for blood pressure measurement , 2002, Blood pressure monitoring.

[23]  Yutaka Imai,et al.  Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults , 2002, Blood pressure monitoring.

[24]  T. Perry,et al.  Comparison of the oscillometric blood pressure monitor (BPM‐100 Beta ) with the auscultatory mercury sphygmomanometer , 2001, Blood pressure monitoring.

[25]  W. White,et al.  Evaluation of the overall efficacy of the Omron office digital blood pressure HEM‐907 monitor in adults , 2001, Blood pressure monitoring.

[26]  D. Yeshurun,et al.  Accuracy of the OMRON M4 automatic blood pressure measuring device , 2000, Journal of Human Hypertension.

[27]  A. Zanchetti,et al.  Guidelines for the use of self-blood pressure monitoring: a summary report of the First International Consensus Conference. Groupe Evaluation & Measure of the French Society of Hypertension. , 2000, Journal of hypertension.

[28]  O. Hanon,et al.  Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. , 1999, Blood pressure monitoring.

[29]  J. Staessen,et al.  Reference values for self-recorded blood pressure: a meta-analysis of summary data. , 1998, Archives of internal medicine.

[30]  R. Córdoba,et al.  Validación de un monitor de automedida de la presión arterial: OMRON-HM 722C , 1997 .

[31]  Modesti,et al.  Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. , 1996, Blood pressure monitoring.

[32]  F. Mee,et al.  Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. , 1996, Blood pressure monitoring.

[33]  T. Pickering,et al.  Recommendations for the use of home (self) and ambulatory blood pressure monitoring. American Society of Hypertension Ad Hoc Panel. , 1996, American journal of hypertension.

[34]  A. Coats,et al.  Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. , 1994, Journal of human hypertension.

[35]  E. O’Brien,et al.  An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. , 1993, Journal of hypertension.

[36]  F. Mee,et al.  Comparative accuracy of six ambulatory devices according to blood pressure levels , 1993, Journal of hypertension.

[37]  F M Charbonnier,et al.  AAMI/ANSI standard for automatic or advisory external defibrillators. Association for the Advancement of Medical Instrumentation. American National Standards Institute. , 1993, Journal of electrocardiology.

[38]  F. Mee,et al.  Evaluation of blood pressure measuring devices. , 1993, Clinical and experimental hypertension.

[39]  E. O’Brien,et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. , 1990, Journal of hypertension.

[40]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[41]  D. Sackett,et al.  IMPROVEMENT OF MEDICATION COMPLIANCE IN UNCONTROLLED HYPERTENSION , 1976, The Lancet.

[42]  Roland Asmara,et al.  Guidelines for the use of self-blood pressure monitoring: a summary report of the first international consensus conference , 2022 .