UNLABELLED
Defective detector blocks in PET may cause serious image artifacts. To estimate the influence of malfunctioning detectors on image quality, a method is described for transferring the actual detector defect onto previously acquired scans.
METHODS
Consequences of detector defects of varying types and extensions were simulated in phantom studies as well as in clinical 18F-fluorodeoxyglucose investigations. First, a condition frame was obtained by dividing the sinogram of a blank measurement, obtained with rod sources on the defective PET camera, by the sinogram of a reference blank acquired before the appearance of the defect. Second, the sinogram of a previously acquired typical patient study was multiplied by the condition frame and reconstructed. Thereafter, images from corrupted sinograms were compared visually with their originals. For repairing defective sinograms, linear interpolation and the constrained Fourier space method were tested.
RESULTS
The effects of detector defects can be simulated accurately in patient studies. The correction methods applied are especially helpful in cases of (a) several neighboring defective detectors and small study objects, (b) small hot artifacts and (c) several nonadjacent defective detectors. Linear interpolation is faster than the constrained Fourier space method; it is more widely applicable and provides similar results.
CONCLUSION
The proposed approach allows specific evaluation of clinical consequences of detector defects. This technique simplifies the decision as to whether a planned patient study can be performed or must be postponed. Even in cases of serious detector problems, sinogram repair may help eliminate image artifacts and minimize the loss of image quality.
[1]
F. Natterer.
The Mathematics of Computerized Tomography
,
1986
.
[2]
J. M. Ollinger,et al.
An evaluation of three algorithms for reconstructing images from data with missing projections
,
1988
.
[3]
S Grootoonk,et al.
Performance Evaluation of the Positron Scanner ECAT EXACT
,
1992,
Journal of computer assisted tomography.
[4]
Victor Perez-Mendez,et al.
Versatility of the CFR algorithm for limited angle reconstruction
,
1990
.
[5]
G Muehllehner,et al.
Constrained Fourier space method for compensation of missing data in emission computed tomography.
,
1988,
IEEE transactions on medical imaging.
[6]
Limited-angle 3D reconstruction of PET images for dose localization in light ion tumour therapy.
,
1992,
Physics in medicine and biology.
[7]
S.-C. Huang,et al.
Capability evaluation of a sinogram error detection and correction method in computed tomography
,
1991,
Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.
[8]
D. Mankoff,et al.
Continuous-slice PENN-PET: a positron tomograph with volume imaging capability.
,
1990,
Journal of nuclear medicine : official publication, Society of Nuclear Medicine.