Calibration and Use of the Aerodynamic Particle Sizer (APS 3300)

The Aerodynamic Particle Sizer (APS) has been in use at the National Institute for Occupational Safety and Health (NIOSH) for over two years, beginning with a prototype model and more recently with a commercial version (Model 3300). The APS has been tested and used in a variety of laboratory and field situations. It has been a very useful instrument for testing aerodynamic sizing devices and provided a much needed means of rapid aerodynamic sizing of particles. Limits to the accuracy of the APS in determining aerodynamic diameter of particles were investigated. The calibration of the APS was originally carried out by using monodisperse di-octyl phthalate (DOP) oil aerosol in the 3–15 μm range. Using a laser imaging system, the flattening of droplets into oblate spheroids was observed for larger particles (20–100 μm). The APS was recalibrated with solid latex particles and the DOP particles were measured to determine the effect of the droplet flattening. A 15 μm droplet is measured as being 20% smaller by ...