Quantitative magnetic resonance imaging phantoms: A review and the need for a system phantom

The MRI community is using quantitative mapping techniques to complement qualitative imaging. For quantitative imaging to reach its full potential, it is necessary to analyze measurements across systems and longitudinally. Clinical use of quantitative imaging can be facilitated through adoption and use of a standard system phantom, a calibration/standard reference object, to assess the performance of an MRI machine. The International Society of Magnetic Resonance in Medicine AdHoc Committee on Standards for Quantitative Magnetic Resonance was established in February 2007 to facilitate the expansion of MRI as a mainstream modality for multi‐institutional measurements, including, among other things, multicenter trials. The goal of the Standards for Quantitative Magnetic Resonance committee was to provide a framework to ensure that quantitative measures derived from MR data are comparable over time, between subjects, between sites, and between vendors. This paper, written by members of the Standards for Quantitative Magnetic Resonance committee, reviews standardization attempts and then details the need, requirements, and implementation plan for a standard system phantom for quantitative MRI. In addition, application‐specific phantoms and implementation of quantitative MRI are reviewed. Magn Reson Med 79:48–61, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

[1]  S. Reeder,et al.  An acetone‐based phantom for quantitative diffusion MRI , 2017, Journal of magnetic resonance imaging : JMRI.

[2]  S. Reeder,et al.  Multisite, multivendor validation of the accuracy and reproducibility of proton‐density fat‐fraction quantification at 1.5T and 3T using a fat–water phantom , 2017, Magnetic resonance in medicine.

[3]  Debiao Li,et al.  Noninvasive measurement of pressure gradient across a coronary stenosis using phase contrast (PC)‐MRI: A feasibility study , 2017, Magnetic resonance in medicine.

[4]  Özcan Gündoğdu,et al.  Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities , 2017 .

[5]  Robert J. Nordstrom,et al.  The Quantitative Imaging Network in Precision Medicine , 2016, Tomography.

[6]  J. M. Nóbrega,et al.  Hollow Polypropylene Yarns as a Biomimetic Brain Phantom for the Validation of High-Definition Fiber Tractography Imaging. , 2016, ACS applied materials & interfaces.

[7]  N. Hylton,et al.  Variability and bias assessment in breast ADC measurement across multiple systems , 2016, Journal of magnetic resonance imaging : JMRI.

[8]  Fang-Cheng Yeh,et al.  Diffusion Capillary Phantom vs. Human Data: Outcomes for Reconstruction Methods Depend on Evaluation Medium , 2016, Front. Neurosci..

[9]  N. Hylton,et al.  Design of a breast phantom for quantitative MRI , 2016, Journal of magnetic resonance imaging : JMRI.

[10]  Peng Hu,et al.  Comparison of T1 measurement using ISMRM/NIST system phantom | NIST , 2016 .

[11]  T. Ebbers,et al.  4D Flow MRI‐based pressure loss estimation in stenotic flows: Evaluation using numerical simulations , 2016, Magnetic resonance in medicine.

[12]  Yunhong Shu,et al.  Integrated image reconstruction and gradient nonlinearity correction , 2015, Magnetic resonance in medicine.

[13]  John Kornak,et al.  Gradient nonlinearity correction to improve apparent diffusion coefficient accuracy and standardization in the american college of radiology imaging network 6698 breast cancer trial , 2015, Journal of magnetic resonance imaging : JMRI.

[14]  Theodoros N. Arvanitis,et al.  MRS thermometry calibration at 3 T: effects of protein, ionic concentration and magnetic field strength , 2015, NMR in biomedicine.

[15]  Christine L. Tardif,et al.  On the accuracy of T1 mapping: Searching for common ground , 2015, Magnetic resonance in medicine.

[16]  H. Barnhart,et al.  The emerging science of quantitative imaging biomarkers terminology and definitions for scientific studies and regulatory submissions , 2015, Statistical methods in medical research.

[17]  Mithat Gönen,et al.  Quantitative imaging biomarkers: A review of statistical methods for technical performance assessment , 2015, Statistical methods in medical research.

[18]  P. Hubbard,et al.  Biomimetic phantom for the validation of diffusion magnetic resonance imaging , 2015, Magnetic resonance in medicine.

[19]  Anthony Gamst,et al.  Magnetic Resonance Elastography Predicts Advanced Fibrosis in Patients With Nonalcoholic Fatty Liver Disease: A Prospective Study , 2014, Hepatology.

[20]  Theodoros N. Arvanitis,et al.  MRS water resonance frequency in childhood brain tumours: a novel potential biomarker of temperature and tumour environment , 2014, NMR in biomedicine.

[21]  M. Suntharalingam,et al.  TU-C-12A-09: Modeling Pathologic Response of Locally Advanced Esophageal Cancer to Chemo-Radiotherapy Using Quantitative PET/CT Features, Clinical Parameters and Demographics. , 2014, Medical physics.

[22]  A. Jackson,et al.  TU-C-12A-08: Thermally-Stabilized Isotropic Diffusion Phantom for Multisite Assessment of Apparent Diffusion Coefficient Reproducibilty. , 2014, Medical physics.

[23]  Xia Li,et al.  Errors in Quantitative Image Analysis due to Platform-Dependent Image Scaling. , 2014, Translational oncology.

[24]  Ryan J Bosca,et al.  Methodological Development of a Multi-Parametric Quantitative Imaging Biomarker Framework for Assessing Treatment Response with MRI , 2014 .

[25]  V. Rieke,et al.  Comparison of temperature processing methods for monitoring focused ultrasound ablation in the brain , 2013, Journal of magnetic resonance imaging : JMRI.

[26]  Gaohong Wu,et al.  Prospective and retrospective high order eddy current mitigation for diffusion weighted echo planar imaging , 2013, Magnetic resonance in medicine.

[27]  B. Tuong,et al.  Development of a novel breast MRI phantom for quality control. , 2013, AJR. American journal of roentgenology.

[28]  F. Ståhlberg,et al.  Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom study , 2013, Acta radiologica.

[29]  Sabrina Fox‐Bosetti,et al.  Reproducibility of hepatic fat fraction measurement by magnetic resonance imaging , 2013, Journal of magnetic resonance imaging : JMRI.

[30]  R. Idilman,et al.  Hepatic steatosis: quantification by proton density fat fraction with MR imaging versus liver biopsy. , 2013, Radiology.

[31]  Timothy D Johnson,et al.  Multi‐system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice‐water phantom , 2013, Journal of magnetic resonance imaging : JMRI.

[32]  J. Duerk,et al.  Magnetic Resonance Fingerprinting , 2013, Nature.

[33]  S. Reeder,et al.  Comparison of R2* correction methods for accurate fat quantification in fatty liver , 2013, Journal of magnetic resonance imaging : JMRI.

[34]  Lynne E. Bilston,et al.  Combining MR elastography and diffusion tensor imaging for the assessment of anisotropic mechanical properties: A phantom study , 2013, Journal of magnetic resonance imaging : JMRI.

[35]  Inline directionally independent peak velocity evaluation reduces error in peak antegrade velocity estimation in patients referred for cardiac valvular assessment. , 2012, AJR. American journal of roentgenology.

[36]  S. Reeder,et al.  Proton density fat‐fraction: A standardized mr‐based biomarker of tissue fat concentration , 2012, Journal of magnetic resonance imaging : JMRI.

[37]  David N Firmin,et al.  A multi-center inter-manufacturer study of the temporal stability of phase-contrast velocity mapping background offset errors , 2012, Journal of Cardiovascular Magnetic Resonance.

[38]  Armando Manduca,et al.  Review of MR elastography applications and recent developments , 2012, Journal of magnetic resonance imaging : JMRI.

[39]  Werner Vach,et al.  Analysis of pulse wave velocity in the thoracic aorta by flow‐sensitive four‐dimensional MRI: Reproducibility and correlation with characteristics in patients with aortic atherosclerosis , 2012, Journal of magnetic resonance imaging : JMRI.

[40]  A. Fleisher,et al.  Altered default mode network connectivity in alzheimer's disease—A resting functional MRI and bayesian network study , 2011, Human brain mapping.

[41]  Kevin M Johnson,et al.  Noninvasive assessment of transstenotic pressure gradients in porcine renal artery stenoses by using vastly undersampled phase-contrast MR angiography. , 2011, Radiology.

[42]  C. Meyer,et al.  Diffusion coefficient measurement using a temperature‐controlled fluid for quality control in multicenter studies , 2011, Journal of magnetic resonance imaging : JMRI.

[43]  Y. Ohno,et al.  T2* measurements of 3-T MRI with ultrashort TEs: capabilities of pulmonary function assessment and clinical stage classification in smokers. , 2011, AJR. American journal of roentgenology.

[44]  Tino Ebbers,et al.  Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor study , 2011, Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance.

[45]  Scott B Reeder,et al.  Quantification of hepatic steatosis with T1-independent, T2-corrected MR imaging with spectral modeling of fat: blinded comparison with MR spectroscopy. , 2011, Radiology.

[46]  Kyle J Myers,et al.  An anthropomorphic phantom for quantitative evaluation of breast MRI. , 2011, Medical physics.

[47]  Peter J Basser,et al.  Pore diameter mapping using double pulsed-field gradient MRI and its validation using a novel glass capillary array phantom. , 2011, Journal of magnetic resonance.

[48]  Mariya Doneva,et al.  Compressed sensing reconstruction for magnetic resonance parameter mapping , 2010, Magnetic resonance in medicine.

[49]  R. Goebel,et al.  Ground truth hardware phantoms for validation of diffusion‐weighted MRI applications , 2010, Journal of magnetic resonance imaging : JMRI.

[50]  Yasmin,et al.  Determinants of pulse wave velocity in healthy people and in the presence of cardiovascular risk factors: ‘establishing normal and reference values’ , 2010, European heart journal.

[51]  M. Alley,et al.  Bicuspid Aortic Valve : Four-dimensional MR Evaluation of Ascending Aortic Systolic Flow Patterns 1 , 2010 .

[52]  A. Mayer,et al.  A prospective diffusion tensor imaging study in mild traumatic brain injury , 2010, Neurology.

[53]  A. van Rossum,et al.  Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements , 2010, Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance.

[54]  A. Badano,et al.  A dynamic lesion phantom for quantitative evaluation of dynamic contrast enhanced MRI , 2010 .

[55]  A. Barker,et al.  Quantification of Hemodynamic Wall Shear Stress in Patients with Bicuspid Aortic Valve Using Phase-Contrast MRI , 2010, Annals of Biomedical Engineering.

[56]  L. Clarke,et al.  Quantitative imaging for evaluation of response to cancer therapy. , 2009, Translational oncology.

[57]  S. Reeder,et al.  T1 independent, T2* corrected MRI with accurate spectral modeling for quantification of fat: Validation in a fat‐water‐SPIO phantom , 2009, Journal of magnetic resonance imaging : JMRI.

[58]  D. Richman,et al.  Resting cerebral blood flow , 2009, Neurology.

[59]  Norbert Schuff,et al.  Measurement of MRI scanner performance with the ADNI phantom. , 2009, Medical physics.

[60]  S. Reeder,et al.  Quantification of hepatic steatosis with MRI: The effects of accurate fat spectral modeling , 2009, Journal of magnetic resonance imaging : JMRI.

[61]  R. Kauppinen,et al.  Monitoring of gliomas in vivo by diffusion MRI and 1H MRS during gene therapy‐induced apoptosis: interrelationships between water diffusion and mobile lipids , 2009, NMR in biomedicine.

[62]  P. Choyke,et al.  Diffusion-weighted magnetic resonance imaging as a cancer biomarker: consensus and recommendations. , 2009, Neoplasia.

[63]  Ram D. Sriram,et al.  Imaging as a Biomarker: Standards for Change Measurements in Therapy workshop summary. , 2008, Academic radiology.

[64]  J. Hennig,et al.  3D MR flow analysis in realistic rapid‐prototyping model systems of the thoracic aorta: Comparison with in vivo data and computational fluid dynamics in identical vessel geometries , 2008, Magnetic resonance in medicine.

[65]  E. Achten,et al.  Simulation and experimental verification of the diffusion in an anisotropic fiber phantom. , 2008, Journal of magnetic resonance.

[66]  Janet Woodcock,et al.  The FDA critical path initiative and its influence on new drug development. , 2008, Annual review of medicine.

[67]  P. Basser,et al.  Polyvinylpyrrolidone (PVP) water solutions as isotropic phantoms for diffusion MRI studies , 2008 .

[68]  D. Donoho,et al.  Sparse MRI: The application of compressed sensing for rapid MR imaging , 2007, Magnetic resonance in medicine.

[69]  A. Manduca,et al.  Assessment of hepatic fibrosis with magnetic resonance elastography. , 2007, Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association.

[70]  Andrea Laghi,et al.  MRI and ultrasound for hepatic fat quantification:relationships to clinical and metabolic characteristics of pediatric nonalcoholic fatty liver disease , 2007, Acta paediatrica.

[71]  Raimo Sepponen,et al.  Phantom‐based evaluation of geometric distortions in functional magnetic resonance and diffusion tensor imaging , 2007, Magnetic resonance in medicine.

[72]  H. Kauczor,et al.  Value of MR phase-contrast flow measurements for functional assessment of pulmonary arterial hypertension , 2007, European Radiology.

[73]  Jan Henrik Ardenkjaer-Larsen,et al.  Metabolic imaging by hyperpolarized 13C magnetic resonance imaging for in vivo tumor diagnosis. , 2006, Cancer research.

[74]  Lee Friedman,et al.  Report on a multicenter fMRI quality assurance protocol , 2006, Journal of magnetic resonance imaging : JMRI.

[75]  G. Wright,et al.  Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity , 2006, Magnetic resonance in medicine.

[76]  A. Fenster,et al.  Computer-controlled positive displacement pump for physiological flow simulation , 2006, Medical and Biological Engineering and Computing.

[77]  Benjamin M Yeh,et al.  Accuracy of liver fat quantification at MR imaging: comparison of out-of-phase gradient-echo and fat-saturated fast spin-echo techniques--initial experience. , 2005, Radiology.

[78]  Eduard E de Lange,et al.  Hyperpolarized HHe 3 MRI of the lung in cystic fibrosis: assessment at baseline and after bronchodilator and airway clearance treatment. , 2005, Academic radiology.

[79]  E. Boltshauser,et al.  Diffusion tensor imaging in progressive multifocal leukoencephalopathy: early predictor for demyelination? , 2005, AJNR. American journal of neuroradiology.

[80]  L. Esserman,et al.  MRI measurements of breast tumor volume predict response to neoadjuvant chemotherapy and recurrence-free survival. , 2005, AJR. American journal of roentgenology.

[81]  D. LeBihan,et al.  Validation of q-ball imaging with a diffusion fibre-crossing phantom on a clinical scanner , 2005, Philosophical Transactions of the Royal Society B: Biological Sciences.

[82]  Maria Drangova,et al.  Multisite trial of MR flow measurement: Phantom and protocol design , 2005, Journal of magnetic resonance imaging : JMRI.

[83]  Bradford A Moffat,et al.  Functional diffusion map: a noninvasive MRI biomarker for early stratification of clinical brain tumor response. , 2005, Proceedings of the National Academy of Sciences of the United States of America.

[84]  M. Leach,et al.  Developing a quality control protocol for diffusion imaging on a clinical MRI system. , 2004, Physics in medicine and biology.

[85]  M. Greicius,et al.  Default-mode network activity distinguishes Alzheimer's disease from healthy aging: Evidence from functional MRI , 2004, Proc. Natl. Acad. Sci. USA.

[86]  A. Sorensen,et al.  Diffusion tensor imaging as potential biomarker of white matter injury in diffuse axonal injury. , 2004, AJNR. American journal of neuroradiology.

[87]  C. Jack,et al.  Comparison of different MRI brain atrophy rate measures with clinical disease progression in AD , 2004, Neurology.

[88]  Vinod Menon,et al.  Functional connectivity in the resting brain: A network analysis of the default mode hypothesis , 2002, Proceedings of the National Academy of Sciences of the United States of America.

[89]  A. Padhani Dynamic contrast‐enhanced MRI in clinical oncology: Current status and future directions , 2002, Journal of magnetic resonance imaging : JMRI.

[90]  Richard I. Kitney,et al.  Standardization of performance evaluation in MRI: 13 Years' experience of intersystem comparison , 2002 .

[91]  John M Buatti,et al.  Method and timing of tumor volume measurement for outcome prediction in cervical cancer using magnetic resonance imaging. , 2002, International journal of radiation oncology, biology, physics.

[92]  O Jolivet,et al.  Precision of magnetic resonance velocity and acceleration measurements: Theoretical issues and phantom experiments , 2001, Journal of magnetic resonance imaging : JMRI.

[93]  D. DeMets,et al.  Biomarkers and surrogate endpoints: Preferred definitions and conceptual framework , 2001, Clinical pharmacology and therapeutics.

[94]  R I Kitney,et al.  Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3 T , 2001, Journal of magnetic resonance imaging : JMRI.

[95]  P. Barber,et al.  Combined 1H MR spectroscopy and diffusion-weighted MRI improves the prediction of stroke outcome , 2000, Neurology.

[96]  M E Bastin,et al.  On the use of water phantom images to calibrate and correct eddy current induced artefacts in MR diffusion tensor imaging. , 2000, Magnetic resonance imaging.

[97]  S. Resnick,et al.  One-year age changes in MRI brain volumes in older adults. , 2000, Cerebral cortex.

[98]  R M Harrison,et al.  Quality assurance for MRI: practical experience. , 2000, The British journal of radiology.

[99]  C Baldock,et al.  Test liquids for quantitative MRI measurements of self‐diffusion coefficient in vivo , 2000, Magnetic resonance in medicine.

[100]  M E Bastin,et al.  Correction of eddy current-induced artefacts in diffusion tensor imaging using iterative cross-correlation. , 1999, Magnetic resonance imaging.

[101]  G. Hagemann,et al.  Fast, accurate, and reproducible automatic segmentation of the brain in T1‐weighted volume MRI data , 1999, Magnetic resonance in medicine.

[102]  O. Pahlm,et al.  Left-to-right cardiac shunts: comparison of measurements obtained with MR velocity mapping and with radionuclide angiography. , 1999, Radiology.

[103]  Norbert Schuff,et al.  Age-related metabolite changes and volume loss in the hippocampus by magnetic resonance spectroscopy and imaging☆ , 1999, Neurobiology of Aging.

[104]  R. Edelman,et al.  A Phantom for diffusion‐weighted imaging of acute stroke , 1998, Journal of magnetic resonance imaging : JMRI.

[105]  L. Clarke,et al.  Monitoring brain tumor response to therapy using MRI segmentation. , 1997, Magnetic resonance imaging.

[106]  Richard Frayne,et al.  Accuracy of MR phase contrast velocity measurements for unsteady flow , 1995, Journal of magnetic resonance imaging : JMRI.

[107]  A. Fenster,et al.  A geometrically accurate vascular phantom for comparative studies of x-ray, ultrasound, and magnetic resonance vascular imaging: construction and geometrical verification. , 1993, Medical physics.

[108]  J. D. de Certaines,et al.  Performance assessment and quality control in MRI by Eurospin test objects and protocols. , 1993, Magnetic resonance imaging.

[109]  Society of magnetic resonance in medicine , 1990 .

[110]  L. Axel,et al.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. , 1990, Medical physics.

[111]  F. Ståhlberg,et al.  A rotating phantom for the study of flow effects in MR imaging. , 1988, Magnetic resonance imaging.

[112]  P. M. Walker,et al.  V. Multi-center trial with protocols and prototype test objects for the assessment of MRI equipment , 1988 .

[113]  D W McRobbie,et al.  Multi-center trial with protocols and prototype test objects for the assessment of MRI equipment. EEC Concerted Research Project. , 1988, Magnetic resonance imaging.