Design, development and experimental validation of a non-invasive device for recording respiratory events during bottle feeding

In newborns, a poor coordination between sucking, swallowing and breathing may undermine the effectiveness of oral feeding and signal immaturity of Central Nervous System. The aim of this work is to develop and validate a non-invasive device for recording respiratory events of newborns during bottle feeding. The proposed device working principle is based on the convective heat exchanged between two hot bodies and the infants' breathing. The sensing elements are inserted into a duct and the gas exchanged by infants is conveyed into this duct thanks to an ad hoc designed system to be mounted on a commercial feeding bottle. Two sets of experiments have been carried out in order to investigate the discrimination threshold of the device and characterize the sensor response at oscillating flows. The effect of distance and tilt between nostrils and device, and the breathing frequency, have been investigated simulating nostrils and neonatal respiratory pattern. The device has a discrimination threshold lower than 0.5 L/min at both 10° and 20° of tilt. Distance for these two settings does not affect the threshold in the investigated range (10-20 mm). Moreover, the device is able to detect breathing events, and to discriminate the onset of expiratory phase, during a neonatal respiratory task delivered by a lung simulator. The results foster the successful application of this device to the assessment of the temporal breathing pattern of newborns during bottle feeding with a non-invasive approach.

[1]  R. Schanler,et al.  Maturation of oral feeding skills in preterm infants , 2007, Acta paediatrica.

[2]  Sergio Silvestri,et al.  Ecological Sucking Monitoring of Newborns , 2013, IEEE Sensors Journal.

[3]  A. Bos,et al.  Uncoordinated sucking patterns in preterm infants are associated with abnormal general movements. , 2012, The Journal of pediatrics.

[4]  O P Mathew,et al.  Breathing patterns of preterm infants during bottle feeding: role of milk flow. , 1991, The Journal of pediatrics.

[5]  Tomohiro Yamada,et al.  Three-dimensional analysis of facial morphology in normal Japanese children as control data for cleft surgery. , 2002, The Cleft palate-craniofacial journal : official publication of the American Cleft Palate-Craniofacial Association.

[6]  Sergio Silvestri,et al.  A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: theory, working principle, and static calibration. , 2013, The Review of scientific instruments.

[7]  Gaurav Agnihotri,et al.  Craniofacial Anthropometry in Newborns and Infants , 2007 .

[8]  Sergio Silvestri,et al.  Linearity dependence on oxygen fraction and gas temperature of a novel Fleisch pneumotachograph for neonatal ventilation at low flow rates , 2012 .

[9]  Sergio Silvestri,et al.  A novel target-type low pressure drop bidirectional optoelectronic air flow sensor for infant artificial ventilation: measurement principle and static calibration. , 2011, The Review of scientific instruments.

[10]  Sergio Silvestri,et al.  A transistor based air flow transducer for thermohygrometric control of neonatal ventilatory applications. , 2008, The Review of scientific instruments.

[11]  Emiliano Schena,et al.  A new ecological method for the estimation of Nutritive Sucking Efficiency in newborns: Measurement principle and experimental assessment , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[12]  K. Carlsen,et al.  Lung function by tidal breathing in awake healthy newborn infants. , 1994, The European respiratory journal.

[13]  Sergio Silvestri,et al.  An orifice meter for bidirectional air flow measurements: Influence of gas thermo-hygrometric content on static response and bidirectionality , 2013 .

[14]  S. Cecchini,et al.  Design and experimental characterization of a gas flow generator to calibrate flow meters for neonatal ventilation , 2012, 2012 4th IEEE RAS & EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob).

[15]  F. Flack,et al.  Comparative Review of Techniques for Recording Respiratory Events at Rest and during Deglutition , 1997, Dysphagia.

[16]  Emiliano Schena,et al.  Technological Solutions and Main Indices for the Assessment of Newborns' Nutritive Sucking: A Review , 2014, Sensors.

[17]  Sergio Silvestri,et al.  An open-loop controlled active lung simulator for preterm infants. , 2011, Medical engineering & physics.

[18]  Lucena Sabrina Lopes,et al.  A non-invasive technique for evaluation of respiratory efforts in preterm infants during feeding , 2014 .