PURPOSE
The purpose of this investigation was to determine whether whole body composition could be estimated in tall individuals using summed measures from two partial DXA scans.
METHODS
Using a pencil-beam DXA, a convenience sample of young individuals (N = 19) were scanned three times. Two partial scans of the upper and lower body were combined to predict bone area, bone mineral content (BMC), bone mineral density (BMD), fat mass (FM), fat-free soft tissue mass (FFST), and percent fat (%Fat), from a complete whole body scan. Two different methods, dividing the body at the hip (HIP) or at the top of the shoulders (NECK), were used to determine optimal method of summing.
RESULTS
There were no significant differences in BMC, BMD, FM, FFST, and %Fat comparing the complete scan and summed scans regardless of summing method. There was excellent agreement between complete and summed values of BMC, BMD, FM, FFST, and %Fat values as indicated by regression analysis (r value range: 0.992-1.00) and nonsignificant intercepts. Bland-Altman plot analysis indicated that a small systemic bias occurred in estimates of FM and %Fat using the HIP method and FFST using the NECK method; however, all biases were negligible.
CONCLUSIONS
The findings suggest that accurate estimates of whole body composition from a complete scan can be obtained by summing two partial scans. Although both summing methods provide good estimates of body composition, dividing the body at the neck provides more accurate estimates of bone and soft tissue composition than dividing the body at the hip.
[1]
A. Beckett,et al.
AKUFO AND IBARAPA.
,
1965,
Lancet.
[2]
Douglas G. Altman,et al.
Measurement in Medicine: The Analysis of Method Comparison Studies
,
1983
.
[3]
P. Tothill,et al.
Precision and accuracy of measurements of whole-body bone mineral: comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers.
,
1994,
The British journal of radiology.
[4]
P. Tothill,et al.
Comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers and other techniques used for whole-body soft tissue measurements.
,
1994,
European journal of clinical nutrition.
[5]
E. Evans,et al.
In vivo validation of whole body composition estimates from dual-energy X-ray absorptiometry.
,
1997,
Journal of applied physiology.
[6]
W. Kohrt.
Preliminary evidence that DEXA provides an accurate assessment of body composition.
,
1998,
Journal of applied physiology.
[7]
T. Lohman,et al.
Reproducibility of DXA in obese women.
,
2002,
Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry.
[8]
H. Genant,et al.
Clinical applications of body composition measurements using DXA.
,
2003,
Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry.