Experimental and numerical analyses of the steady flow field around an aortic Björk-Shiley standard valve prosthesis.

To develop a numerical method for the description of the flow field around a Björk-Shiley (BS) standard valve prosthesis in aortic position, detailed experimental measurements and numerical calculations are performed under steady flow conditions. The experiment was conducted at Reynolds numbers up to 800. In order to perform LDA measurement of velocity in the vicinity of the valve with a curved sinus boundary, a mixture of oil and kerosine was used as the fluid which exactly matches the refractive index of the perspex aortic model. The velocity profiles at six positions in the vicinity and downstream of the valve were measured, including both axial and radial velocity components. The results show very clearly the existence of two nearly symmetric spiral vortex streams downstream of the valve. There is no recirculation area in the aorta downstream and also no obvious stagnation area in the minor orifice region near the valve when Re less than or equal to 800. Theoretically, the flow field of a BS valve is simulated by the flow pattern around a circular plate with an angle of incidence to the approaching stream. The numerical calculations were carried out by means of a 2-D model using the FEM together with the penalty function method. The maximum Reynolds number is 700. The results agree with the experimental results in the plane of symmetry when the Reynolds number is small. However, as the Reynolds number increases, the difference becomes evident. Our conclusion is that the steady flow field of a BS valve is completely 3-dimensional, featured by two spiral vortices. It cannot be simulated exactly by 2-D numerical calculations. To get more detailed and complete information about the flow field of this valve, 3-D numerical calculations are needed.

[1]  W G Tiederman,et al.  Two-component laser velocimeter measurements downstream of heart valve prostheses in pulsatile flow. , 1986, Journal of biomechanical engineering.

[2]  钱勤建,et al.  Study of the closing mechanism of natural heart valves , 1986 .

[3]  K. Affeld,et al.  A Ten Times Enlarged Model of Artificial Heart Valve Flow , 1990 .

[4]  K. Chandran,et al.  Steady flow development past valve prostheses in a model human aorta. II. Tilting disc valves. , 1983, Journal of biomechanics.

[5]  Richard Figliola,et al.  Fluid Stresses in the Vicinity of Disk, Ball, and Tilting Disk Prosthetic Heart Valves From In-Vitro Measurements , 1977 .

[6]  S R Idelsohn,et al.  A comparative computational study of blood flow through prosthetic heart valves using the finite element method. , 1985, Journal of biomechanics.

[7]  M. Giersiepen,et al.  Three-dimensional visualization of axial velocity profiles downstream of six different mechanical aortic valve prostheses, measured with a hot-film anemometer in a steady state flow model. , 1987, Journal of biomechanics.

[8]  K. Chandran,et al.  Steady flow development past valve prostheses in a model human aorta. I. Centrally occluding valves. , 1983, Journal of biomechanics.

[9]  A P Yoganathan,et al.  In vitro velocity measurements in the vicinity of aortic prostheses. , 1979, Journal of biomechanics.

[10]  B Khalighi,et al.  Experimental study of physiological pulsatile flow past valve prosthesis in a model of human aorta--II. Tilting disc valves and the effect of orientation. , 1985, Journal of biomechanics.

[11]  B Khalighi,et al.  Experimental study of physiological pulsatile flow past valve prostheses in a model of human aorta--I. Caged ball valves. , 1985, Journal of biomechanics.

[12]  J. D. Janssen,et al.  A numerical analysis of steady flow in a three-dimensional model of the carotid artery bifurcation. , 1990, Journal of biomechanics.

[13]  R Pelissier,et al.  Velocity profiles in the wake of two prosthetic heart valves using a new cardiovascular simulator. , 1985, Journal of biomechanics.

[14]  W. M. Swanson,et al.  Dimensions and Geometric Relationships of the Human Aortic Value as a Function of Pressure , 1974, Circulation research.

[15]  A P Yoganathan,et al.  In vitro flow dynamics of four prosthetic aortic valves: a comparative analysis. , 1989, Journal of biomechanics.

[16]  H. Reul,et al.  Three-dimensional visualization of velocity fields downstream of six mechanical aortic valves in a pulsatile flow model. , 1988, Journal of biomechanics.

[17]  J. R. Carl,et al.  The Bjork-Shiley Aortic Prosthesis Flow Characteristics, Thrombus Formation and Tissue Overgrowth , 1978, Circulation.

[18]  Charles S. Peskin,et al.  Modeling prosthetic heart valves for numerical analysis of blood flow in the heart , 1980 .

[19]  H Reul,et al.  Model studies at mechanical aortic heart valve prostheses--Part I: Steady-state flow fields and pressure loss coefficients. , 1988, Journal of biomechanical engineering.

[20]  A. Au,et al.  Computer visualization of flow patterns for prosthetic heart valves. , 1976, Scandinavian journal of thoracic and cardiovascular surgery.

[21]  T. Schmitz-Rode,et al.  The geometry of the aortic root in health, at valve disease and after valve replacement. , 1990, Journal of biomechanics.

[22]  T. Mueller,et al.  On the hemolytic and thrombogenic potential of occluder prosthetic heart valves from in-vitro measurements. , 1981, Journal of biomechanical engineering.

[23]  J. J. Horsten On the analysis of moving heart valves : a numerical fluid-structure interaction model , 1990 .