Reduced anatomical clutter in digital breast tomosynthesis with statistical iterative reconstruction.

PURPOSE Digital breast tomosynthesis (DBT) has been shown to somewhat alleviate the breast tissue overlapping issues of two-dimensional (2D) mammography. However, the improvement in current DBT systems over mammography is still limited. Statistical image reconstruction (SIR) methods have the potential to reduce through-plane artifacts in DBT, and thus may be used to further reduce anatomical clutter. The purpose of this work was to study the impact of SIR on anatomical clutter in the reconstructed DBT image volumes. METHODS An SIR with a slice-wise total variation (TV) regularizer was implemented to reconstruct DBT images which were compared with the clinical reconstruction method (filtered backprojection). The artifact spread function (ASF) was measured to quantify the reduction of the through-plane artifacts level in phantom studies and microcalcifications in clinical cases. The anatomical clutter was quantified by the anatomical noise power spectrum with a power law fitting model: NPSa ( f) = α f-β . The β values were measured from the reconstructed image slices when the two reconstruction methods were applied to a cohort of clinical breast exams (N = 101) acquired using Hologic Selenia Dimensions DBT systems. RESULTS The full width half maximum (FWHM) of the measured ASF was reduced from 8.7 ± 0.1 mm for clinical reconstruction to 6.5 ± 0.1 mm for SIR which yields a 25% reduction in FWHM in phantom studies and the same amount of ASF reduction was also found in clinical measurements from microcalcifications. The measured β values for the two reconstruction methods were 3.17 ± 0.36 and 2.14 ± 0.39 for the clinical reconstruction method and the SIR method, respectively. This difference was statistically significant (P << 0.001). The dependence of β on slice location using either method was negligible. CONCLUSIONS Statistical image reconstruction enabled a significant reduction of both the through-plane artifacts level and anatomical clutter in the DBT reconstructions. The β value was found to be β≈2.14 with the SIR method. This value stays in the middle between the β≈1.8 for cone beam CT and β≈3.2 for mammography. In contrast, the measured β value in the clinical reconstructions (β≈3.17) remains close to that of mammography.

[1]  Berkman Sahiner,et al.  Application of boundary detection information in breast tomosynthesis reconstruction. , 2007, Medical physics.

[2]  E. Sidky,et al.  Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization , 2008, Physics in medicine and biology.

[3]  Ehsan Samei,et al.  Evaluation of Low-Contrast Detectability of Iterative Reconstruction across Multiple Institutions, CT Scanner Manufacturers, and Radiation Exposure Levels. , 2015, Radiology.

[4]  John M Boone,et al.  Characterizing anatomical variability in breast CT images. , 2008, Medical physics.

[5]  J. Rakowski,et al.  A comparison of reconstruction algorithms for C-arm mammography tomosynthesis. , 2006, Medical physics.

[6]  Arthur E Burgess,et al.  Signal detection in power-law noise: effect of spectrum exponents. , 2007, Journal of the Optical Society of America. A, Optics, image science, and vision.

[7]  Lubomir M. Hadjiiski,et al.  Selective-diffusion regularization for enhancement of microcalcifications in digital breast tomosynthesis reconstruction. , 2010, Medical physics.

[8]  Nuno Matela,et al.  Total variation minimization filter for DBT imaging. , 2015, Medical physics.

[9]  J. H. van Hateren,et al.  Modelling the Power Spectra of Natural Images: Statistics and Information , 1996, Vision Research.

[10]  Jeffrey A. Fessler,et al.  Alternating Direction Method of Multiplier for Tomography With Nonlocal Regularizers , 2014, IEEE Transactions on Medical Imaging.

[11]  Kyle J. Myers,et al.  Model observers for assessment of image quality , 1993 .

[12]  James T Dobbins,et al.  Digital x-ray tomosynthesis: current state of the art and clinical potential. , 2003, Physics in medicine and biology.

[13]  J H Siewerdsen,et al.  Investigation of lung nodule detectability in low-dose 320-slice computed tomography. , 2009, Medical physics.

[14]  Jeffrey A. Fessler,et al.  Fast X-Ray CT Image Reconstruction Using a Linearized Augmented Lagrangian Method With Ordered Subsets , 2014, IEEE Transactions on Medical Imaging.

[15]  Jeffrey A. Fessler,et al.  Alternating Dual Updates Algorithm for X-ray CT Reconstruction on the GPU , 2015, IEEE Transactions on Computational Imaging.

[16]  C. Bouman,et al.  A recursive filter for noise reduction in statistical iterative tomographic imaging , 2006, Electronic Imaging.

[17]  Francis R. Verdun,et al.  Importance of anatomical noise in mammography , 1997, Medical Imaging.

[18]  Kwang Eun Jang,et al.  Statistical reconstruction using dual formulation of subband-wise total variation regularization (SDST) for limited angle tomography , 2011, 2011 IEEE International Symposium on Biomedical Imaging: From Nano to Macro.

[19]  Bo Zhao,et al.  Image artifacts in digital breast tomosynthesis: investigation of the effects of system geometry and reconstruction parameters using a linear system approach. , 2008, Medical physics.

[20]  Berkman Sahiner,et al.  Truncation artifact and boundary artifact reduction in breast tomosynthesis reconstruction , 2008, SPIE Medical Imaging.

[21]  Xiaochuan Pan,et al.  Enhanced imaging of microcalcifications in digital breast tomosynthesis through improved image-reconstruction algorithms. , 2009, Medical physics.

[22]  Chulkyu Park,et al.  Improvement of image performance in digital breast tomosynthesis (DBT) by incorporating a compressed-sensing (CS)-based deblurring scheme , 2016 .

[23]  Lubomir M. Hadjiiski,et al.  A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis. , 2006, Medical physics.

[24]  M. Yaffe,et al.  Characterisation of mammographic parenchymal pattern by fractal dimension. , 1990, Physics in medicine and biology.

[25]  S. A. Zolotarev,et al.  Experimental investigation of iterative reconstruction techniques for high resolution mammography , 2014 .

[26]  Jeffrey A. Fessler,et al.  Detector Blur and Correlated Noise Modeling for Digital Breast Tomosynthesis Reconstruction , 2018, IEEE Transactions on Medical Imaging.

[27]  Anita Nosratieh,et al.  Anatomical complexity in breast parenchyma and its implications for optimal breast imaging strategies. , 2012, Medical physics.

[28]  Wei Zhao,et al.  Breast Structural Noise in Digital Breast Tomosynthesis and Its Dependence on Reconstruction Methods , 2010, Digital Mammography / IWDM.

[29]  Ken D. Sauer,et al.  Model-Based Iterative Reconstruction for Dual-Energy X-Ray CT Using a Joint Quadratic Likelihood Model , 2014, IEEE Transactions on Medical Imaging.

[30]  F R Verdun,et al.  Estimation of the noisy component of anatomical backgrounds. , 1999, Medical physics.

[31]  E. Samei,et al.  Dose dependence of mass and microcalcification detection in digital mammography: free response human observer studies. , 2007, Medical physics.

[32]  Ke Li,et al.  Anatomical background noise power spectrum in differential phase contrast and dark field contrast mammograms. , 2014, Medical physics.

[33]  Fiona J Gilbert,et al.  Development and experience of quality control methods for digital breast tomosynthesis systems. , 2015, The British journal of radiology.

[34]  D. Conover,et al.  Background DEDICATED BREAST CT: ANATOMIC POWER SPECTRUM , 2012 .

[35]  Ian Shaw,et al.  Design and performance of the prototype full field breast tomosynthesis system with selenium based flat panel detector , 2005, SPIE Medical Imaging.

[36]  L T Niklason,et al.  Digital breast imaging: tomosynthesis and digital subtraction mammography. , 1998, Breast disease.

[37]  Emily F Conant,et al.  Breast cancer screening using tomosynthesis in combination with digital mammography. , 2014, JAMA.

[38]  J. Boone,et al.  Association between power law coefficients of the anatomical noise power spectrum and lesion detectability in breast imaging modalities. , 2013, Physics in medicine and biology.

[39]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[40]  Junfeng Yang,et al.  A New Alternating Minimization Algorithm for Total Variation Image Reconstruction , 2008, SIAM J. Imaging Sci..

[41]  E. Conant,et al.  Breast Cancer Screening Using Tomosynthesis and Digital Mammography in Dense and Nondense Breasts. , 2016, JAMA.

[42]  E. Loli Piccolomini,et al.  A fast Total Variation-based iterative algorithm for digital breast tomosynthesis image reconstruction , 2016 .

[43]  Aydin Akan,et al.  Digital breast tomosynthesis image reconstruction using 2D and 3D total variation minimization , 2013, Biomedical engineering online.

[44]  Otto Zhou,et al.  Statistical iterative reconstruction to improve image quality for digital breast tomosynthesis. , 2015, Medical physics.

[45]  Jeffrey A. Fessler,et al.  A Splitting-Based Iterative Algorithm for Accelerated Statistical X-Ray CT Reconstruction , 2012, IEEE Transactions on Medical Imaging.

[46]  Ellen Warner,et al.  Clinical practice. Breast-cancer screening. , 2011, The New England journal of medicine.

[47]  Nuno Matela,et al.  3D Total Variation Minimization Filter for Breast Tomosynthesis Imaging , 2016, Digital Mammography / IWDM.

[48]  Emil Y. Sidky,et al.  Optimizing algorithm parameters based on a model observer detection task for image reconstruction in digital breast tomosynthesis , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[49]  J. Fessler,et al.  Segmented separable footprint projector for digital breast tomosynthesis and its application for subpixel reconstruction , 2017, Medical physics.

[50]  Jeffrey A. Fessler,et al.  Relaxed Linearized Algorithms for Faster X-Ray CT Image Reconstruction , 2015, IEEE Transactions on Medical Imaging.

[51]  I. Sechopoulos A review of breast tomosynthesis. Part I. The image acquisition process. , 2013, Medical physics.

[52]  Srinivasan Vedantham,et al.  Scaling-law for the energy dependence of anatomic power spectrum in dedicated breast CT. , 2013, Medical physics.

[53]  A. Burgess,et al.  Human observer detection experiments with mammograms and power-law noise. , 2001, Medical physics.

[54]  D. Kopans,et al.  Digital tomosynthesis in breast imaging. , 1997, Radiology.

[55]  Patrick L. Combettes,et al.  Signal Recovery by Proximal Forward-Backward Splitting , 2005, Multiscale Model. Simul..

[56]  James G. Nagy,et al.  A scaled gradient method for digital tomographic image reconstruction , 2018 .

[57]  Andriy I. Bandos,et al.  Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. , 2013, Radiology.

[58]  A. Rodríguez-Ruiz,et al.  Evaluation of the technical performance of three different commercial digital breast tomosynthesis systems in the clinical environment. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[59]  Tao Wu,et al.  A comparison of reconstruction algorithms for breast tomosynthesis. , 2004, Medical physics.