Validation of a Monte Carlo model used for simulating tube current modulation in computed tomography over a wide range of phantom conditions/challenges.

PURPOSE Monte Carlo (MC) simulation methods have been widely used in patient dosimetry in computed tomography (CT), including estimating patient organ doses. However, most simulation methods have undergone a limited set of validations, often using homogeneous phantoms with simple geometries. As clinical scanning has become more complex and the use of tube current modulation (TCM) has become pervasive in the clinic, MC simulations should include these techniques in their methodologies and therefore should also be validated using a variety of phantoms with different shapes and material compositions to result in a variety of differently modulated tube current profiles. The purpose of this work is to perform the measurements and simulations to validate a Monte Carlo model under a variety of test conditions where fixed tube current (FTC) and TCM were used. METHODS A previously developed MC model for estimating dose from CT scans that models TCM, built using the platform of mcnpx, was used for CT dose quantification. In order to validate the suitability of this model to accurately simulate patient dose from FTC and TCM CT scan, measurements and simulations were compared over a wide range of conditions. Phantoms used for testing range from simple geometries with homogeneous composition (16 and 32 cm computed tomography dose index phantoms) to more complex phantoms including a rectangular homogeneous water equivalent phantom, an elliptical shaped phantom with three sections (where each section was a homogeneous, but different material), and a heterogeneous, complex geometry anthropomorphic phantom. Each phantom requires varying levels of x-, y- and z-modulation. Each phantom was scanned on a multidetector row CT (Sensation 64) scanner under the conditions of both FTC and TCM. Dose measurements were made at various surface and depth positions within each phantom. Simulations using each phantom were performed for FTC, detailed x-y-z TCM, and z-axis-only TCM to obtain dose estimates. This allowed direct comparisons between measured and simulated dose values under each condition of phantom, location, and scan to be made. RESULTS For FTC scans, the percent root mean square (RMS) difference between measurements and simulations was within 5% across all phantoms. For TCM scans, the percent RMS of the difference between measured and simulated values when using detailed TCM and z-axis-only TCM simulations was 4.5% and 13.2%, respectively. For the anthropomorphic phantom, the difference between TCM measurements and detailed TCM and z-axis-only TCM simulations was 1.2% and 8.9%, respectively. For FTC measurements and simulations, the percent RMS of the difference was 5.0%. CONCLUSIONS This work demonstrated that the Monte Carlo model developed provided good agreement between measured and simulated values under both simple and complex geometries including an anthropomorphic phantom. This work also showed the increased dose differences for z-axis-only TCM simulations, where considerable modulation in the x-y plane was present due to the shape of the rectangular water phantom. Results from this investigation highlight details that need to be included in Monte Carlo simulations of TCM CT scans in order to yield accurate, clinically viable assessments of patient dosimetry.

[1]  J J DeMarco,et al.  A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models , 2003, Physics in medicine and biology.

[2]  Steven L Simon,et al.  Organ doses for reference adult male and female undergoing computed tomography estimated by Monte Carlo simulations. , 2011, Medical physics.

[3]  Ehsan Samei,et al.  Patient-specific radiation dose and cancer risk estimation in CT: part I. development and validation of a Monte Carlo program. , 2010, Medical physics.

[4]  J J DeMarco,et al.  A CT-based Monte Carlo simulation tool for dosimetry planning and analysis. , 1998, Medical physics.

[5]  A. Fanaroff,et al.  Radiation Dose to the Fetus for Pregnant Patients Undergoing Multidetector CT Imaging: Monte Carlo Simulations Estimating Fetal Dose for a Range of Gestational Age and Patient Size , 2009 .

[6]  Daniel Kolditz,et al.  Fast on-site Monte Carlo tool for dose calculations in CT applications. , 2012, Medical physics.

[7]  Martin Caon,et al.  Monte Carlo Calculated Effective Dose to Teenage Girls from Computed Tomography Examinations , 2000 .

[8]  Willi A. Kalender,et al.  Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomography , 2008, European Radiology.

[9]  M Caon,et al.  An EGS4-ready tomographic computational model of a 14-year-old female torso for calculating organ doses from CT examinations. , 1999, Physics in medicine and biology.

[10]  Walter Huda,et al.  X-ray tube current modulation and patient doses in chest CT. , 2011, Radiation protection dosimetry.

[11]  Cynthia H McCollough,et al.  Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. , 2009, Medical physics.

[12]  Nina Petoussi-Henss,et al.  The GSF family of voxel phantoms , 2002, Physics in medicine and biology.

[13]  W. Huda,et al.  CT doses in cylindrical phantoms. , 1995, Physics in medicine and biology.

[14]  Wesley E Bolch,et al.  Organ doses for reference pediatric and adolescent patients undergoing computed tomography estimated by Monte Carlo simulation. , 2012, Medical physics.

[15]  Maryam Khatonabadi,et al.  A comparison of methods to estimate organ doses in CT when utilizing approximations to the tube current modulation function. , 2012, Medical physics.

[16]  J J DeMarco,et al.  A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms. , 2005, Physics in medicine and biology.

[17]  Wesley E Bolch,et al.  Monte Carlo simulations of adult and pediatric computed tomography exams: validation studies of organ doses with physical phantoms. , 2012, Medical physics.

[18]  Ehsan Samei,et al.  Patient-specific radiation dose and cancer risk for pediatric chest CT. , 2011, Radiology.

[19]  W. L. Dunn,et al.  A Monte Carlo model for absorbed dose calculations in computed tomography. , 1983, Medical physics.

[20]  Cynthia H McCollough,et al.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations. , 2009, Medical physics.

[21]  B Bednarz,et al.  The development, validation and application of a multi-detector CT (MDCT) scanner model for assessing organ doses to the pregnant patient and the fetus using Monte Carlo simulations , 2009, Physics in medicine and biology.

[22]  D. Rogers Fifty years of Monte Carlo simulations for medical physics , 2006, Physics in medicine and biology.

[23]  Cynthia H McCollough,et al.  Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT , 2009, Physics in medicine and biology.

[24]  M. McNitt-Gray,et al.  Precision of dosimetry-related measurements obtained on current multidetector computed tomography scanners. , 2010, Medical physics.

[25]  J J DeMarco,et al.  Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose , 2007, Physics in medicine and biology.

[26]  M Caon,et al.  A comparison of radiation dose measured in CT dosimetry phantoms with calculations using EGS4 and voxel-based computational models. , 1997, Physics in medicine and biology.

[27]  B. Schmidt,et al.  A PC program for estimating organ dose and effective dose values in computed tomography , 1999, European Radiology.