Quantification of flow rates using harmonic grey-scale imaging and an ultrasound contrast agent: an in vitro and in vivo study.

It is unclear if the dye-dilution theory and its corresponding parameters are capable of measuring brain perfusion using harmonic grey-scale imaging. We performed a study on a flow phantom using a SONOS 5500 (1.8--3.6-MHz harmonic imaging) and Levovist as the ultrasound (US) contrast agent (UCA). We applied the UCA in six different doses (0.1 to 3.0 mL) and used eight different flow-rates (180 to 540 mL/min). Additionally, we performed a study on dog brain using Levovist boluses of 1.5 mL and 3 mL. We evaluated the influence of dose and flow-rate on the parameters of the time-intensity curve: peak signal intensity (PSI), area under the curve (AUC) and mean transit time (MTT). Along with an increase of the Levovist dose, the AUC and the PSI increased only in the dose range between 0.1 and 0.5 mL Levovist; further increase led to no change of parameters. Flow-rate showed no influence on AUC, MTT or PSI. The dye-dilution theory is not a useful theoretical model for the analysis of perfusion using harmonic grey-scale imaging. A possible explanation for this effect is the bubble saturation.

[1]  M. Kaps,et al.  Visualization of brain perfusion with harmonic gray scale and power doppler technology : an animal pilot study. , 2000, Stroke.

[2]  A J Hindle,et al.  A perfusion phantom for the evaluation of ultrasound contrast agents. , 1994, Ultrasound in medicine & biology.

[3]  B. Rosen,et al.  Perfusion imaging with NMR contrast agents , 1990, Magnetic resonance in medicine.

[4]  J. Fike,et al.  Measurement of regional cerebral blood flow using ultrafast computed tomography. Theoretical aspects. , 1991, Stroke.

[5]  K R Hoffmann,et al.  Determination of instantaneous and average blood flow rates from digital angiograms of vessel phantoms using distance-density curves. , 1991, Investigative radiology.

[6]  G. Seidel,et al.  Harmonic imaging of the human brain. Visualization of brain perfusion with ultrasound. , 2000, Stroke.

[7]  T. Porter,et al.  Transient myocardial contrast after initial exposure to diagnostic ultrasound pressures with minute doses of intravenously injected microbubbles. Demonstration and potential mechanisms. , 1995, Circulation.

[8]  H. Przuntek,et al.  Transient response harmonic imaging: an ultrasound technique related to brain perfusion. , 1998, Stroke.

[9]  N de Jong,et al.  Absorption and scatter of encapsulated gas filled microspheres: theoretical considerations and some measurements. , 1992, Ultrasonics.

[10]  P A Heidenreich,et al.  In vitro calculation of flow by use of contrast ultrasonography. , 1993, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[11]  M. Kaps,et al.  The influence of Doppler system settings on the clearance kinetics of different ultrasound contrast agents. , 1999, European journal of ultrasound : official journal of the European Federation of Societies for Ultrasound in Medicine and Biology.

[12]  A Herment,et al.  In vitro flow quantification with contrast power Doppler imaging. , 2000, Ultrasound in medicine & biology.