Improved DQE by means of X-ray spectra and scintillator optimization for FFDM

The focus of this work was to improve the DQE performance of a full-field digital mammography (FFDM) system by means of selecting an optimal X-ray tube anode-filter combination in conjunction with an optimal scintillator configuration. The flat panel detector in this work is a Varian PaxScan 3024M. The detector technology is comprised of a 2816 row × 3584 column amorphous silicon (a-Si) photodiode array with a pixel pitch of 83μm. The scintillator is cesium iodide and is deposited directly onto the photodiode array and available with configurable optical and x-ray properties. Two X-ray beam spectra were generated with the anode/filter combinations, Molybdenum/Molybdenum (Mo/Mo) and Tungsten/Aluminum (W/Al), to evaluate the imaging performance of two types of scintillators, high resolution (HR) type and high light output (HL) type. The results for the HR scintillator with W/Al anode-filter (HRW/ Al) yielded a DQE(0) of 67%, while HR-Mo/Mo was lower with a DQE(0) of 50%. In addition, the DQE(0) of the HR-W/Al configuration was comparable to the DQE(0) of the HL-Mo/Mo configuration. The significance of this result is the HR type scintillator yields about twice the light output with the W/Al spectrum, at about half the dose, as compared to the Mo/Mo spectrum. The light output or sensitivity was measured in analog-to-digital convertor units (ADU) per dose. The sensitivities (ADU/uGy) were 8.6, 16.8 and 25.4 for HR-Mo/Mo, HR-W/Al, HL-Mo/Mo, respectively. The Nyquist frequency for the 83 μm pixel is 6 lp/mm. The MTF at 5 lp/mm for HR-Mo/Mo and HR-W/Al were equivalent at 37%, while the HL-Mo/Mo MTF was 24%. According to the DQE metric, the more favorable anodefilter combination was W/Al with the HR scintillator. Future testing will evaluate the HL-W/Al configuration, as well as other x-ray filters materials and other scintillator optimizations. While higher DQE values were achieved, the more general conclusion is that the imaging performance can be tuned as required by the application by modifying optical and x-ray properties of the scintillator to match the spectral output of the chosen anode-filter combination.

[1]  R. Shaw,et al.  The Equivalent Quantum Efficiency of the Photographic Process , 1963 .

[2]  D. Dreossi,et al.  Measurement of the linear attenuation coefficients of breast tissues by synchrotron radiation computed tomography , 2010, Physics in medicine and biology.

[3]  Hilde Bosmans,et al.  Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system , 2006, European Radiology.

[4]  Harold L. Kundel,et al.  Handbook of Medical Imaging, Volume 1. Physics and Psychophysics , 2000 .

[5]  Kirby G. Vosburgh,et al.  X-Ray Image Intensifiers , 1977 .

[6]  R. A. Baker,et al.  Xeroradiography , 1975, Lancet.

[7]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[8]  Mari Varjonen,et al.  Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography , 2008, SPIE Medical Imaging.

[9]  P. C. Johns,et al.  X-ray characterisation of normal and neoplastic breast tissues. , 1987, Physics in medicine and biology.

[10]  G. Entine,et al.  Structured CsI(Tl) scintillators for X-ray imaging applications , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[11]  Michael P. Andre,et al.  Optimization of tungsten x-ray spectra for digital mammography: a comparison of model to experiment , 1997, Medical Imaging.

[12]  D R Dance,et al.  Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study. , 2000, The British journal of radiology.

[13]  Ehsan Samei,et al.  Measurement of the detective quantum efficiency in digital detectors consistent with the IEC 62220-1 standard: Practical considerations regarding the choice of filter material. , 2005, Medical physics.

[14]  Biao Chen,et al.  Imaging technique optimization of tungsten anode FFDM system , 2009, Medical Imaging.

[15]  T. Mertelmeier,et al.  Comparison of Anode/Filter Combinations in Digital Mammography with Respect to the Average Glandular Dose , 2008, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.