Differential Attenuation Imaging for the Characterization of High Intensity Focused Ultrasound Lesions

High intensity focused ultrasound (HIFU) is an effective technique for creating coagulative necrotic lesions in biological tissue, with a view to treating localized tumors. Although good results have already been obtained, notably in urology, current systems lack a real time monitoring system to check the efficacy of the treatment procedures. This study describes the development and assessment of a noninvasive system for making local measurements of attenuation variations during HIFU treatment procedures. An apparatus (Ablatherm, Edap-Technomed, France), combining a 2.5 MHz therapeutic transducer and a 5.5 MHz twin plane imaging probe (connected to an ultrasound scanner), was used to produce lesions. The rf signals needed to calculate the attenuation were recorded as outputs from the ultrasound scanner, before and after the high intensity firing sequences, which were performed on ten pieces of porcine liver. Each firing sequence involved producing a lesion volume comprising 42 individual lesions. A number of recordings were also made without producing lesions, in order to test the reproducibility of the measurements. The attenuation function was evaluated locally using the centroid and the multinarrowband methods. Initially, changes in the integrated attenuation αbar; (mean attenuation in the 4–7 MHz range) and the attenuation slope β were examined for the lesion volume. β values did not vary significantly within this range, whereas α values varied significantly (in the region of 86% of the initial level) in comparison to measurements performed without forming lesions. The differential attenuation Δα (representing local variations in α) was subsequently used to generate images revealing the lesion areas. There was a strong similarity between these ‘Δα images’ and the lesion volumes defined by the operator. ‘Δα images’ offer several advantages over existing attenuation imaging techniques. Any problems related to the heterogeneity of the medium are eliminated, since only the change in attenuation is taken into account. Furthermore, there is no need to compensate for diffraction when estimating Δα, as the rf signals are captured in exactly the same positions before and after treatment. This technique can be used during in vivo treatment procedures. It can be implemented in real time, since the computational algorithms (based primarily on FFT calculations) are very fast. The technique should provide clinical practitioners with valuable qualitative and quantitative information for use in HIFU ultrasound surgery.

[1]  J M Thijssen,et al.  Precision and accuracy of acoustospectrographic parameters. , 1996, Ultrasound in Medicine and Biology.

[2]  M. Marberger,et al.  Tissue ablation in benign prostatic hyperplasia with high intensity focused ultrasound. , 1994, The Journal of urology.

[3]  G Kossoff,et al.  Average velocity of ultrasound in the human female breast. , 1973, The Journal of the Acoustical Society of America.

[4]  M. Fink,et al.  Optimal Precision in Ultrasound Attenuation Estimation and Application to the Detection of Duchenne Muscular Dystrophy Carriers , 1987 .

[5]  C. Damianou,et al.  Dependence of ultrasonic attenuation and absorption in dog soft tissues on temperature and thermal dose. , 1997, The Journal of the Acoustical Society of America.

[6]  J. G. Miller,et al.  Attenuation Estimation in Reflection: Progress and Prospects , 1984, Ultrasonic imaging.

[7]  J. Havlice,et al.  Estimation of the Slope of the Acoustic Attenuation Coefficient , 1984 .

[8]  J. Driller,et al.  A Therapeutic Ultrasound System Incorporating Real-Time Ultrasonic Scanning , 1986, IEEE 1986 Ultrasonics Symposium.

[9]  W J FRY,et al.  Ultrasonic lesions in the mammalian central nervous system. , 1955, Science.

[10]  J. Ophir,et al.  Ultrasonic attenuation measurements of in vivo human muscle. , 1982, Ultrasonic imaging.

[11]  N J Pelc,et al.  Spectral characterization and attenuation measurements in ultrasound. , 1983, Ultrasonic imaging.

[12]  K. Hynynen,et al.  MRI-guided noninvasive ultrasound surgery. , 1993, Medical physics.

[13]  A. Herment,et al.  Application of autoregressive spectral analysis for ultrasound attenuation estimation: interest in highly attenuating medium , 1995, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[14]  J. Cardoso,et al.  Diffraction Effects in Pulse-Echo Measurement , 1984, IEEE Transactions on Sonics and Ultrasonics.

[15]  R. Kuc,et al.  Estimating the Acoustic Attenuation Coefficient Slope for Liver from Reflected Ultrasound Signals , 1979, IEEE Transactions on Sonics and Ultrasonics.

[16]  G. Berger,et al.  Absolute backscatter coefficient over a wide range of frequencies in a tissue-mimicking phantom containing two populations of scatterers , 1996, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[17]  K. Hynynen The threshold for thermally significant cavitation in dog's thigh muscle in vivo. , 1991, Ultrasound in medicine & biology.

[18]  R. Seip,et al.  Noninvasive estimation of tissue temperature response to heating fields using diagnostic ultrasound , 1995, IEEE Transactions on Biomedical Engineering.

[19]  L. Wilson,et al.  Frequency domain processing for ultrasonic attenuation measurement in liver. , 1984, Ultrasonic imaging.

[20]  J C Bamber,et al.  Acoustic properties of lesions generated with an ultrasound therapy system. , 1993, Ultrasound in medicine & biology.

[21]  T. Ponchon,et al.  Characterization of extracorporeal ablation of normal and tumor-bearing liver tissue by high intensity focused ultrasound. , 1993, Ultrasound in medicine & biology.

[22]  J M Dubernard,et al.  Treatment of prostate cancer with transrectal focused ultrasound: early clinical experience. , 1996, European urology.

[23]  R. Kuc,et al.  Variation of acoustic attenuation coefficient slope estimates for in vivo liver. , 1982, Ultrasound in medicine & biology.

[24]  D E Robinson,et al.  Beam Pattern (Diffraction) Correction for Ultrasonic Attenuation Measurement , 1984, Ultrasonic imaging.

[25]  J C Bamber,et al.  Effect of gaseous inclusions on the frequency dependence of ultrasonic attenuation in liver. , 1985, Ultrasound in medicine & biology.

[26]  H. Saunders,et al.  Fundamentals of Acoustics (3rd Ed.) , 1983 .