A prototype master-slave system consisting of two MR-compatible manipulators with interchangeable surgical tools: part of a unified support system for diagnosis and treatment

A prototype master-slave system consisting of two MR-compatible manipulators with interchangeable surgical tools was designed, implemented, and evaluated. The manipulators are designed so as to fit into vertical-magnetic-field open-configuration MR imagers. They are also designed so that the surgical tool mounted at the tip of their arm is interchangeable. The MR compatibility of the manipulator including a surgical tool was evaluated by moving it close to a phantom in the field of view of an open-configuration MR imager. No noticeable deformation, but deterioration in the signal-to-noise ratio of approximately 10 percent, was observed in the MR images. Consulted clinicians then evaluated the manipulability of the master-slave system by maneuvering a pair of slave manipulators through an operation input device consisting of a pair of master manipulators. After several minutes of trial and error, they could perform each primitive motion for suturing tasks on training material mimicking human tissue.

[1]  Nobuhiko Hata,et al.  MR Compatible Surgical Assist Robot: System Integration and Preliminary Feasibility Study , 2000, MICCAI.

[2]  Toshikazu Kawai,et al.  Micromanipulator system (NeuRobot): clinical application in neurosurgery , 2003, CARS.

[3]  Masakatsu G. Fujie,et al.  A magnetic resonance compatible surgical manipulator: part of a unified support system for the diagnosis and treatment of heart disease , 2003, Adv. Robotics.

[4]  Masakatsu G. Fujie,et al.  Development of MR Compatible Surgical Manipulator toward a Unified Support System for Diagnosis and Treatment of Heart Disease , 2002, MICCAI.

[5]  Nobuhiko Hata,et al.  Surgical assist robot for the active navigation in the intraoperative MRI: hardware design issues , 2000, Proceedings. 2000 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS 2000) (Cat. No.00CH37113).

[6]  Masakatsu G. Fujie,et al.  Microscopic-manipulator system for minimally invasive neurosurgery: preliminary study for clinical application , 2001, CARS.

[7]  Masakatsu G. Fujie,et al.  An MR-compatible master-slave manipulator with interchangeable surgical tools , 2003, CARS.

[8]  Tatsuo Arai,et al.  Robotic Assist for MR-Guided Surgery Using Leverage and Parallelepiped Mechanism , 2000, MICCAI.

[9]  Ron Kikinis,et al.  MR Compatibility of Mechatronic Devices: Design Criteria , 1999, MICCAI.

[10]  H Iseki,et al.  Development of an MRI-compatible needle insertion manipulator for stereotactic neurosurgery. , 1995, Journal of image guided surgery.