Line fiducial material and thickness considerations for ultrasound calibration

Ultrasound calibration is a necessary procedure in many image-guided interventions, relating the position of tools and anatomical structures in the ultrasound image to a common coordinate system. This is a necessary component of augmented reality environments in image-guided interventions as it allows for a 3D visualization where other surgical tools outside the imaging plane can be found. Accuracy of ultrasound calibration fundamentally affects the total accuracy of this interventional guidance system. Many ultrasound calibration procedures have been proposed based on a variety of phantom materials and geometries. These differences lead to differences in representation of the phantom on the ultrasound image which subsequently affect the ability to accurately and automatically segment the phantom. For example, taut wires are commonly used as line fiducials in ultrasound calibration. However, at large depths or oblique angles, the fiducials appear blurred and smeared in ultrasound images making it hard to localize their cross-section with the ultrasound image plane. Intuitively, larger diameter phantoms with lower echogenicity are more accurately segmented in ultrasound images in comparison to highly reflective thin phantoms. In this work, an evaluation of a variety of calibration phantoms with different geometrical and material properties for the phantomless calibration procedure was performed. The phantoms used in this study include braided wire, plastic straws, and polyvinyl alcohol cryogel tubes with different diameters. Conventional B-mode and synthetic aperture images of the phantoms at different positions were obtained. The phantoms were automatically segmented from the ultrasound images using an ellipse fitting algorithm, the centroid of which is subsequently used as a fiducial for calibration. Calibration accuracy was evaluated for these procedures based on the leave-one-out target registration error. It was shown that larger diameter phantoms with lower echogenicity are more accurately segmented in comparison to highly reflective thin phantoms. This improvement in segmentation accuracy leads to a lower fiducial localization error, which ultimately results in low target registration error. This would have a profound effect on calibration procedures and the feasibility of different calibration procedures in the context of image-guided procedures.

[1]  K. Cleary,et al.  Image-guided interventions : technology and applications , 2008 .

[2]  Mohammed Bennani Dosse,et al.  Anisotropic Orthogonal Procrustes Analysis , 2010, J. Classif..

[3]  Jørgen Arendt Jensen,et al.  Synthetic aperture ultrasound imaging. , 2006, Ultrasonics.

[4]  Terry M. Peters,et al.  Solving for free-hand and real-time 3D ultrasound calibration with anisotropic orthogonal Procrustes analysis , 2014, Medical Imaging.

[5]  R W Prager,et al.  Rapid calibration for 3-D freehand ultrasound. , 1998, Ultrasound in medicine & biology.

[6]  P A Lewin,et al.  Modified synthetic transmit aperture algorithm for ultrasound imaging. , 2012, Ultrasonics.

[7]  Michael C. Kolios,et al.  BIOMEDICAL ULTRASOUND IMAGING FROM 1 TO 1000 MHZ , 2009 .

[8]  K. Cleary,et al.  Image Guided Interventions. , 2020, Biomedizinische Technik. Biomedical engineering.

[9]  Terry M. Peters,et al.  Registration of 3D shapes under anisotropic scaling , 2015, International Journal of Computer Assisted Radiology and Surgery.

[10]  Berthold K. P. Horn,et al.  Closed-form solution of absolute orientation using unit quaternions , 1987 .

[11]  Terry M. Peters,et al.  Synthetic aperture imaging in ultrasound calibration , 2014, Medical Imaging.

[12]  Purang Abolmaesumi,et al.  A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. , 2009, Ultrasound in medicine & biology.

[13]  Richard Brull,et al.  Is Ultrasound Guidance Advantageous for Interventional Pain Management? A Systematic Review of Chronic Pain Outcomes , 2013, Anesthesia and analgesia.

[14]  Guido Gerig,et al.  Medical Image Computing and Computer-Assisted Intervention - MICCAI 2005, 8th International Conference, Palm Springs, CA, USA, October 26-29, 2005, Proceedings, Part I , 2005, MICCAI.

[15]  Frank Sauer,et al.  A Novel Phantom-Less Spatial and Temporal Ultrasound Calibration Method , 2005, MICCAI.

[16]  Andras Lasso,et al.  Design of a tracked ultrasound calibration phantom made of LEGO bricks , 2014, Medical Imaging.

[17]  Andras Lasso,et al.  PLUS: Open-Source Toolkit for Ultrasound-Guided Intervention Systems , 2014, IEEE Transactions on Biomedical Engineering.

[18]  Paul J. Besl,et al.  A Method for Registration of 3-D Shapes , 1992, IEEE Trans. Pattern Anal. Mach. Intell..