Accuracy of commercial devices and methods for noninvasive estimation of aortic systolic blood pressure a systematic review and meta-analysis of invasive validation studies

Background: Although compelling evidence has established the physiological and clinical relevance of aortic SBP (a-SBP), no consensus exists regarding the validity of the available methods/techniques that noninvasively measure it. Objectives: The systematic review and meta-analysis aimed to determine the accuracy of commercial devices estimating a-SBP noninvasively, which have been validated by invasive measurement of a-SBP. Moreover their optimal mode of application, in terms of calibration, as well as specific technique and arterial site of pulse wave acquisition were further investigated. Methods: The study was performed according to the PRISMA guidelines; 22 eligible studies were included, which validated invasively 11 different commercial devices in 808 study participants. Results: Overall, the error in a-SBP estimation (estimated minus actual value) was −4.49 mmHg [95% confidence interval (CI): −6.06 to −2.92 mmHg]. The estimated (noninvasive) a-SBP differed from the actual (invasive) value depending on calibration method: by −1.08 mmHg (95% CI: −2.81, 0.65 mmHg) and by −5.81 mmHg (95% CI: −7.79, −3.84 mmHg), when invasively and noninvasively measured brachial BP values were used respectively; by −1.83 mmHg, (95% CI: −3.32, −0.34 mmHg), and by 7.78 mmHg (95% CI: −10.28, −5.28 mmHg), when brachial mean arterial pressure/DBP and SBP/DBP were used, respectively. Conclusion: Automated recording of waveforms, calibrated noninvasively by brachial mean arterial pressure/DBP values seems the most promising approach that can provide relatively more accurate, noninvasive estimation of a-SBP. It is still uncertain whether a specific device can be recommended as ‘gold standard’; however, a consensus is currently demanding.

[1]  Kenji Takazawa,et al.  Relationship between Radial and Central Arterial Pulse Wave and Evaluation of Central Aortic Pressure Using the Radial Arterial Pulse Wave , 2007, Hypertension Research.

[2]  D. Moher,et al.  Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. , 2010, International journal of surgery.

[3]  S. Millasseau,et al.  Non-invasive estimation of aortic blood pressures: a close look at current devices and methods. , 2014, Current pharmaceutical design.

[4]  Simon Redwood,et al.  The noninvasive estimation of central aortic blood pressure in patients with aortic stenosis , 2008, Journal of hypertension.

[5]  J. Lekakis,et al.  Transmission of calibration errors (input) by generalized transfer functions to the aortic pressures (output) at different hemodynamic states. , 2006, International journal of cardiology.

[6]  Paolo Salvi,et al.  Validation of a new non-invasive portable tonometer for determining arterial pressure wave and pulse wave velocity: the PulsePen device , 2004, Journal of hypertension.

[7]  Siegfried Wassertheurer,et al.  Validation of a Brachial Cuff-Based Method for Estimating Central Systolic Blood Pressure , 2011, Hypertension.

[8]  Yasmin,et al.  Central Pressure: Variability and Impact of Cardiovascular Risk Factors: The Anglo-Cardiff Collaborative Trial II , 2008, Hypertension.

[9]  J. Lekakis,et al.  The effect of antihypertensive drugs on central blood pressure beyond peripheral blood pressure. Part I: (Patho)-physiology, rationale and perspective on pulse pressure amplification. , 2009, Current pharmaceutical design.

[10]  E. Lakatta,et al.  Hypertension Derivation and Validation of Diagnostic Thresholds for Central Blood Pressure Measurements Based on Long-Term Cardiovascular Risks , 2022 .

[11]  M. Safar,et al.  Cuff and aortic pressure differences during dobutamine infusion: a study of the effects of systolic blood pressure amplification. , 2010, American heart journal.

[12]  Hiroshi Miyashita,et al.  Clinical Assessment of Central Blood Pressure , 2012, Current hypertension reviews.

[13]  T. Gillebert Central blood pressure and its amplification: a final breakthrough or do we need more? , 2014, European heart journal.

[14]  Ruiyan Zhang,et al.  Validation of the central blood pressure estimation by the SphygmoCor system in Chinese , 2009, Blood pressure monitoring.

[15]  Patrick Segers,et al.  Role of pulse pressure amplification in arterial hypertension: experts' opinion and review of the data. , 2009, Hypertension.

[16]  Alice Stanton,et al.  Differential Impact of Blood Pressure–Lowering Drugs on Central Aortic Pressure and Clinical Outcomes: Principal Results of the Conduit Artery Function Evaluation (CAFE) Study , 2006, Circulation.

[17]  J. Sharman,et al.  Brachial-to-radial SBP amplification: implications of age and estimated central blood pressure from radial tonometry , 2015, Journal of hypertension.

[18]  Pascal Verdonck,et al.  Amplification of the Pressure Pulse in the Upper Limb in Healthy, Middle-Aged Men and Women , 2009, Hypertension.

[19]  T. Tamaki,et al.  Augmentation index does not reflect risk of coronary artery disease in elderly patients. , 2014, Circulation journal : official journal of the Japanese Circulation Society.

[20]  Yan Li,et al.  Comparison of the SphygmoCor and Omron devices in the estimation of pressure amplification against the invasive catheter measurement , 2013, Journal of hypertension.

[21]  A. Lowe,et al.  Evaluation of a novel sphygmomanometer, which estimates central aortic blood pressure from analysis of brachial artery suprasystolic pressure waves , 2012, Journal of hypertension.

[22]  Simon R Redwood,et al.  Pressure Wave Reflection Assessed From the Peripheral Pulse: Is a Transfer Function Necessary? , 2003, Hypertension.

[23]  Alberto Avolio,et al.  Estimation of central aortic systolic pressure using late systolic inflection of radial artery pulse and its application to vasodilator therapy , 2012, Journal of hypertension.

[24]  Hao-Min Cheng,et al.  Quantification of the calibration error in the transfer function-derived central aortic blood pressures. , 2011, American journal of hypertension.

[25]  Hao-Min Cheng,et al.  Measurement accuracy of non-invasively obtained central blood pressure by applanation tonometry: a systematic review and meta-analysis. , 2013, International journal of cardiology.

[26]  Jaspal Kooner,et al.  Estimation of central aortic pressure by SphygmoCor requires intra-arterial peripheral pressures. , 2003, Clinical science.

[27]  M. Band,et al.  Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure , 2003, Journal of hypertension.

[28]  I. Roberts,et al.  Blood pressure lowering and cardiovascular risk , 2014, The Lancet.

[29]  H. Bøtker,et al.  Assessment of central blood pressure in patients with type 2 diabetes: a comparison between SphygmoCor and invasively measured values. , 2014, American journal of hypertension.

[30]  B. Gaszner,et al.  Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity , 2010, Journal of hypertension.

[31]  Christodoulos Stefanadis,et al.  Prediction of cardiovascular events and all-cause mortality with central haemodynamics: a systematic review and meta-analysis. , 2010, European heart journal.

[32]  H. Bøtker,et al.  Invasive validation of arteriograph estimates of central blood pressure in patients with type 2 diabetes. , 2014, American journal of hypertension.

[33]  Christian Ott,et al.  Comparison of Two Noninvasive Devices for Measurement of Central Systolic Blood Pressure With Invasive Measurement During Cardiac Catheterization , 2012, Journal of clinical hypertension.

[34]  Telmo Pereira,et al.  Invasive validation of the Complior Analyse in the assessment of central artery pressure curves: a methodological study , 2014, Blood pressure monitoring.

[35]  I. Wilkinson,et al.  Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure , 2013, Journal of hypertension.

[36]  M. O'Rourke,et al.  Pulse wave analysis. , 1996, Journal of hypertension. Supplement : official journal of the International Society of Hypertension.

[37]  Michele Tarsilla Cochrane Handbook for Systematic Reviews of Interventions , 2010, Journal of MultiDisciplinary Evaluation.

[38]  J. Cockcroft,et al.  Central blood pressure: current evidence and clinical importance. , 2014, European heart journal.

[39]  J. Sharman,et al.  Evaluation of a brachial cuff and suprasystolic waveform algorithm method to noninvasively derive central blood pressure. , 2015, American journal of hypertension.

[40]  Martin Szarski,et al.  Estimation of central aortic blood pressure: a systematic meta-analysis of available techniques , 2014, Journal of hypertension.

[41]  Michael F O'Rourke,et al.  Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform. , 2004, American journal of hypertension.

[42]  Bryan Williams,et al.  Development and validation of a novel method to derive central aortic systolic pressure from the radial pressure waveform using an n-point moving average method. , 2011, Journal of the American College of Cardiology.

[43]  M. Safar,et al.  Clinical Utility of Aortic Pulses and Pressures Calculated From Applanated Radial-Artery Pulses , 2003, Hypertension.

[44]  Siegfried Wassertheurer,et al.  Assessment of systolic aortic pressure and its association to all cause mortality critically depends on waveform calibration , 2015, Journal of hypertension.

[45]  Joseph Cheriyan,et al.  The accuracy of central SBP determined from the second systolic peak of the peripheral pressure waveform , 2009, Journal of hypertension.

[46]  Ira Tabas,et al.  Letter by Williams and Tabas regarding article "atherosclerosis 2005: recent discoveries and novel hypotheses". , 2006, Circulation.

[47]  Brian Clapp,et al.  Estimating central systolic blood pressure during oscillometric determination of blood pressure: proof of concept and validation by comparison with intra-aortic pressure recording and arterial tonometry , 2012, Blood pressure monitoring.