Noise of surface bio-potential electrodes based on NASICON ceramic and Ag−AgCl

The electrochemical noise from dry NASICON-based surface electrodes and pregelled Ag−AgCl electrodes is evaluated in saline solutions and on the skin. The electrochemical noise from the electrode/electrolyte interface is found to be negligible (less than 1 μV peak to peak). On the skin, the noise level is highly dependent on the patient. At high frequencies, the skin/electrode interface noise is equal to ‘thermal noise’ and can be related to the real part of the skin/electrode impedance. At low frequencies (f<100 Hz), excess noise is observed that varies as f−2. It is tentatively ascribed to a non-stationary process or noise of electrochemical origin due to the ionic nature of the skin. The contribution of residual EMG signal of low amplitude (5 μV peak to peak) is suggested for electrodes with large surface area.

[1]  A. Nightingale,et al.  Background noise in electromyography. , 1959, Physics in medicine and biology.

[2]  Electrical Noise from ‘Polarization Cells’ and from Human Tissues , 1958, Nature.

[3]  R. Pallàs-Areny,et al.  Electrode contact noise in surface biopotential measurements , 1992, 1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[4]  Gideon F. Inbar,et al.  On Surface EMG Spectral Characterization and Its Application to Diagnostic Classification , 1984, IEEE Transactions on Biomedical Engineering.

[5]  P. Fitch Signal/noise ratio in the recording of human nerve-action potentials , 1973, Medical and biological engineering.

[6]  H. Callen,et al.  Irreversibility and Generalized Noise , 1951 .

[7]  F. Huet,et al.  Fluctuations in electrochemical systems. I. General theory on diffusion limited electrochemical reactions , 1993 .

[8]  M. Fleischmann,et al.  Generation-recombination noise in weak electrolytes , 1970 .

[9]  E. Novakov,et al.  Non-polarisable dry electrode based on NASICON ceramic , 1995, Medical and Biological Engineering and Computing.

[10]  R. N. Scott,et al.  Noise characteristics of stainless-steel surface electrodes , 1991, Medical and Biological Engineering and Computing.

[11]  P. Fabry,et al.  NASICON based electrodes for bioelectric signal measurements , 1992, 1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[12]  A. Flasterstein A general analysis of voltage fluctuations of metal-electrolyte interfaces , 1966, Medical and biological engineering.

[13]  H. Nyquist Thermal Agitation of Electric Charge in Conductors , 1928 .

[15]  F. Huet,et al.  Investigation of electrochemical processes by an electrochemical noise analysis. Theoretical and experimental aspects in potentiostatic regime , 1986 .

[16]  H A deVries,et al.  Relationship of resting EMG level to total body metabolism with reference to the origin of "tissue noise". , 1976, American journal of physical medicine.

[17]  C D Ferris,et al.  Electrode-produced signal distortion in electrophysiological recording systems. , 1974, IEEE transactions on bio-medical engineering.

[18]  deVries Ha,et al.  Relationship of resting EMG level to total body metabolism with reference to the origin of "tissue noise". , 1976 .

[19]  Dry electrode based on nasicon ceramic for surface biopotential measurements : skin-electrode impedance and noise , 1995 .

[20]  G. C. Barker Noise connected with electrode processes , 1969 .

[21]  Claude Gabrielli,et al.  Fluctuations in electrochemical systems. II. Application to a diffusion limited redox process , 1993 .