Delayed prosthesis malposition after transcatheter aortic valve implantation causing coronaries obstruction.

A case of delayed malposition of a CoreValve device causing obstruction of coronary ostia is described. Nine months after the original implant, the patient developed an acute coronary syndrome and was readmitted to hospital. Angiogram demonstrated an ostial stenosis of both the left main stem and the right coronary ostia, which were filled by a paravalvular leakage of the bioprosthesis. Gated computed tomography scan with 3D reconstruction showed valve malposition with cusps situated 14 mm above the ostium of the right coronary and the presence of fibrous and calcific agglomerations associated to one of the cusp causing a tight stenosis of the left ostium. Computed tomography scan is a crucial imaging technique in the transcatheter aortic valve replacement field and in this case enabled us to identify an interesting phenomenon of fibrosis/calcification originating at the level of the misplaced valve, which was actually the triggering cause of the coronary obstruction. Considering the reported need for more accurate investigations regarding the predictors of negative outcomes and the selection of transcatheter aortic valve replacement candidates, the use of cardiac-gated computed tomography should be stimulated and promoted as a valuable aid for the diagnosis and further clinical decision making in those patients.

[1]  L. Nombela‐Franco,et al.  Coronary obstruction following transcatheter aortic valve implantation: a systematic review. , 2013, JACC. Cardiovascular interventions.

[2]  A. Diegeler,et al.  Transcatheter aortic valve implantation using the ACURATE TA™ system: 1-year outcomes and comparison of 500 patients from the SAVI registries , 2017, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[3]  R. Hetzer,et al.  Delayed subtotal coronary obstruction after transapical aortic valve implantation. , 2011, Interactive cardiovascular and thoracic surgery.

[4]  J. Ye Coronary artery occlusion as a complication of transcatheter aortic valve implantation. , 2012, The Canadian journal of cardiology.

[5]  P. Serruys,et al.  Incidence, timing, and predictors of valve dislodgment during TAVI with the medtronic corevalve system , 2012, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.