Comparison of computational analysis with clinical measurement of stresses on below-knee residual limb in a prosthetic socket.

Interface pressures and shear stresses between a below-knee residual limb and prosthetic socket predicted using finite element analyses were compared with experimental measurements. A three-dimensional nonlinear finite element model, based on actual residual geometry and incorporating PTB socket rectification and interfacial friction/slip conditions, was developed to predict the stress distribution. A system for measuring pressures and bi-axial shear stresses was used to measure the stresses in the PTB socket of a trans-tibial amputee. The FE-predicted results indicated that the peak pressure of 226 kPa occurred at the patellar tendon area and the peak shear stress of 50 kPa at the anterolateral tibia area. Quantitatively, FE-predicted pressures were 11%, on average, lower than those measured by triaxial transducers placed at all the measurement sites. Because friction/slip conditions between the residual limb and socket liner were taken into consideration by using interface elements in the FE model, the directions and magnitudes of shear stresses match well between the FE prediction and clinical measurements. The results suggest that the nonlinear mechanical properties of soft tissues and dynamic effects during gait should be addressed in future work.

[1]  D. Childress,et al.  Normal and shear stresses on a residual limb in a prosthetic socket during ambulation: comparison of finite element results with experimental measurements. , 1993, Journal of rehabilitation research and development.

[2]  M Zhang,et al.  A finite element analysis of the load transfer between an above-knee residual limb and its prosthetic socket--roles of interface friction and distal-end boundary conditions. , 1996, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[3]  V. C. Roberts,et al.  Frictional action at lower limb/prosthetic socket interface. , 1996, Medical engineering & physics.

[4]  Krouskop Ta,et al.  Computer-aided design of a prosthetic socket for an above-knee amputee. , 1987 .

[5]  V. C. Roberts,et al.  Development of a non-linear finite element modelling of the below-knee prosthetic socket interface. , 1995, Medical engineering & physics.

[6]  D S Childress,et al.  Parametric analysis using the finite element method to investigate prosthetic interface stresses for persons with trans-tibial amputation. , 1996, Journal of rehabilitation research and development.

[7]  M Zhang,et al.  Finite element modelling of a residual lower-limb in a prosthetic socket: a survey of the development in the first decade. , 1998, Medical engineering & physics.

[8]  J. R. Engsberg,et al.  Quantifying interface pressures in below-knee amputee sockets , 1993 .

[9]  Ming Zhang Biomechanics of the residual limb and prosthetic socket interface in below-knee amputees. , 1995 .

[10]  D. Smith,et al.  Foot loading in amputee stance , 1984, Prosthetics and orthotics international.

[11]  Koepke Gh,et al.  Effect of liner materials on interface pressures in below-knee prostheses. , 1970 .

[12]  H B Skinner,et al.  Analysis of a below-knee patellar tendon-bearing prosthesis: a finite element study. , 1991, Journal of rehabilitation research and development.

[13]  A. Mak,et al.  In vivo friction properties of human skin , 1999, Prosthetics and orthotics international.

[14]  V. C. Roberts,et al.  Triaxial force transducer for investigating stresses at the stump/socket interface , 2006, Medical and Biological Engineering and Computing.

[15]  E M Burgess,et al.  Clinical measurement of normal and shear stresses on a trans-tibial stump: Characteristics of wave-form shapes during walking , 1993, Prosthetics and orthotics international.

[16]  V. C. Roberts,et al.  Clinical investigation of the pressure and shear stress on the trans-tibial stump with a prosthesis. , 1998, Medical engineering & physics.