Intraoperative localization of brachytherapy implants using intensity-based registration

In prostate brachytherapy, a transrectal ultrasound (TRUS) will show the prostate boundary but not all the implanted seeds, while fluoroscopy will show all the seeds clearly but not the boundary. We propose an intensity-based registration between TRUS images and the implant reconstructed from fluoroscopy as a means of achieving accurate intra-operative dosimetry. The TRUS images are first filtered and compounded, and then registered to the fluoroscopy model via mutual information. A training phantom was implanted with 48 seeds and imaged. Various ultrasound filtering techniques were analyzed, and the best results were achieved with the Bayesian combination of adaptive thresholding, phase congruency, and compensation for the non-uniform ultrasound beam profile in the elevation and lateral directions. The average registration error between corresponding seeds relative to the ground truth was 0.78 mm. The effect of false positives and false negatives in ultrasound were investigated by masking true seeds in the fluoroscopy volume or adding false seeds. The registration error remained below 1.01 mm when the false positive rate was 31%, and 0.96 mm when the false negative rate was 31%. This fully automated method delivers excellent registration accuracy and robustness in phantom studies, and promises to demonstrate clinically adequate performance on human data as well.

[1]  R. Abugharbieh,et al.  2A-4 Enhancement of Bone Surface Visualization from 3D Ultrasound Based on Local Phase Information , 2006, 2006 IEEE Ultrasonics Symposium.

[2]  Gabor Fichtinger,et al.  Prostate Implant Reconstruction with Discrete Tomography , 2007, MICCAI.

[3]  Russell H. Taylor,et al.  Fast algorithm for probabilistic bone edge detection (FAPBED) , 2005, SPIE Medical Imaging.

[4]  R. Robb,et al.  Seed localization and TRUS-fluoroscopy fusion for intraoperative prostate brachytherapy dosimetry , 2007, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[5]  R. Abugharbieh,et al.  ENHANCEMENT OF BONE SURFACE VISUALIZATION FROM 3 D ULTRASOUND BASED ON LOCAL PHASE INFORMATION , 2006 .

[6]  I. Kaplan,et al.  Real-time computed tomography dosimetry during ultrasound-guided brachytherapy for prostate cancer. , 2006, Brachytherapy.

[7]  S. Pathak,et al.  Ultrasonography and fluoroscopic fusion for prostate brachytherapy dosimetry. , 2002, International journal of radiation oncology, biology, physics.

[8]  M. Oldham,et al.  Cone-beam-CT guided radiation therapy: technical implementation. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  Purang Abolmaesumi,et al.  A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness , 2008, SPIE Medical Imaging.

[10]  Marco Zaider,et al.  On the question of 3D seed reconstruction in prostate brachytherapy: the determination of x-ray source and film locations. , 2004, Physics in medicine and biology.

[11]  Gabor Fichtinger,et al.  Intra-operative 3D Guidance in Prostate Brachytherapy Using a Non-isocentric C-arm , 2007, MICCAI.

[12]  Yongmin Kim,et al.  Seed-based ultrasound and fluoroscopy registration using iterative optimal assignment for intraoperative prostate brachytherapy dosimetry , 2007, SPIE Medical Imaging.

[13]  N. Dubrawsky Cancer statistics , 1989, CA: a cancer journal for clinicians.

[14]  Gabor Fichtinger,et al.  Intensity-based registration of prostate brachytherapy implants and ultrasound , 2008, 2008 5th IEEE International Symposium on Biomedical Imaging: From Nano to Macro.

[15]  David R. Haynor,et al.  PET-CT image registration in the chest using free-form deformations , 2003, IEEE Transactions on Medical Imaging.

[16]  Peter Kovesi,et al.  Image Features from Phase Congruency , 1995 .