The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging

Today the majority of clinical molecular imaging procedures are carried out with single-photon emitters and gamma cameras, in planar mode and single-photon emission computed tomography (SPECT) mode. Thanks to the development of advanced multi-pinhole collimation technologies, SPECT imaging of small experimental animals is rapidly gaining in popularity. Whereas resolutions in routine clinical SPECT are typically larger than 1 cm (corresponding to >1,000 μl), it has recently proved possible to obtain spatial resolutions of about 0.35 mm (≈0.04 μl) in the mouse. Meanwhile, SPECT systems that promise an even better performance are under construction. The new systems are able to monitor functions in even smaller structures of the mouse than was possible with dedicated small animal positron emission tomography (≈1 mm resolution, corresponding to 1 μl). This paper provides a brief history of image formation with pinholes and explains the principles of pinhole imaging and pinhole tomography and the basics of modern image reconstruction methods required for such systems. Some recently introduced ultra-high-resolution small animal SPECT instruments are discussed and new avenues for improving system performance are explored. This may lead to many completely new biomedical applications. We also demonstrate that clinical SPECT systems with focussing pinhole gamma cameras will be able to produce images with a resolution that may become superior to that of PET for major clinical applications. A design study of a cardiac pinhole SPECT system indicates that the heart can be imaged an order of magnitude faster or with much more detail than is possible with currently used parallel-hole SPECT (e.g. 3–4 mm instead of ≈8 mm system resolution).

[1]  P Wollmer,et al.  Pinhole emission computed tomography: method and experimental evaluation. , 1990, Physics in medicine and biology.

[2]  Eustace L. Dereniak,et al.  Progress in the development of large-area modular 64/spl times/64 CdZnTe imaging arrays for nuclear medicine , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[3]  H.H. Barrett,et al.  Progress in the development of large-area modular 64/spl times/64 CdZnTe imaging arrays for nuclear medicine , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[4]  Bruce H. Hasegawa,et al.  ECG-gated pinhole SPECT in mice with millimeter spatial resolution , 2000 .

[5]  G. Zeng,et al.  An analytical algorithm for skew-slit imaging geometry with nonuniform attenuation correction. , 2006, Medical physics.

[6]  P. V. van Rijk,et al.  U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[7]  Feng Zhang,et al.  3D position-sensitive CdZnTe gamma-ray spectrometers: improved performance with new ASICs , 2004, SPIE Optics + Photonics.

[8]  Donald W. Wilson,et al.  FastSPECT II: a second-generation high-resolution dynamic SPECT imager , 2002, IEEE Transactions on Nuclear Science.

[9]  K. Tamura,et al.  Novel Photon-Counting Detector for 0.1–100 keV X-Ray Imaging Possessing High Spatial Resolution , 2003 .

[10]  Matthew A. Kupinski,et al.  Small-Animal Spect Imaging , 2005 .

[11]  J. Lees,et al.  The high resolution gamma imager (HRGI): a CCD based camera for medical imaging , 2003 .

[12]  Hiroyuki Kudo,et al.  A new reconstruction strategy for image improvement in pinhole SPECT , 2004, European Journal of Nuclear Medicine and Molecular Imaging.

[13]  R. Dinapoli,et al.  Medipix2: A 64-k pixel readout chip with 55-/spl mu/m square elements working in single photon counting mode , 2001 .

[14]  P J Serlemitsos Conical foil x-ray mirrors: performance and projections. , 1988, Applied optics.

[15]  V. Perez-Mendez,et al.  A method of tomographic imaging using a multiple pinhole-coded aperture. , 1974, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[16]  Emi Miyata,et al.  Development of new X-ray imaging device sensitive to 0.1–100 keV , 2004 .

[17]  Antonio Cossu,et al.  The usefulness of neck pinhole SPECT as a complementary tool to planar scintigraphy in primary and secondary hyperparathyroidism. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  N. Clinthorne,et al.  Design and feasibility study of a single photon emission microscope system for small animal I-125 imaging , 2006, IEEE Transactions on Nuclear Science.

[19]  S. Walrand,et al.  Evaluation of novel whole-body high-resolution rodent SPECT (Linoview) based on direct acquisition of linogram projections. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[20]  Donald W. Wilson,et al.  SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. , 2006, Medical physics.

[21]  J. H. Hammond The Camera Obscura, A Chronicle , 1981 .

[22]  R. Jaszczak,et al.  Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT. , 1994, Physics in medicine and biology.

[23]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[24]  R. Wojcik,et al.  A restraint-free small animal SPECT imaging system with motion tracking , 2005, IEEE Transactions on Nuclear Science.

[25]  M. Wernick,et al.  Emission Tomography: The Fundamentals of PET and SPECT , 2004 .

[26]  Heike Soltau,et al.  A monolithic array of silicon drift detectors coupled to a single scintillator for γ-ray imaging with sub-millimeter position resolution , 2003 .

[27]  Donald W. Wilson,et al.  Quantitative analysis of acute myocardial infarct in rat hearts with ischemia-reperfusion using a high-resolution stationary SPECT system. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[28]  Daniel Vavrik,et al.  First tests of a Medipix-1 pixel detector for X-ray dynamic defectoscopy , 2002 .

[29]  Ronald J. Jaszczak,et al.  Determination of mechanical and electronic shifts for pinhole SPECT using a single point source , 2005, IEEE Transactions on Medical Imaging.

[30]  Zhong He,et al.  Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers , 2000 .

[31]  K. Lange,et al.  EM reconstruction algorithms for emission and transmission tomography. , 1984, Journal of computer assisted tomography.

[32]  F. Beekman,et al.  Submillimeter total-body murine imaging with U-SPECT-I. , 2007, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[33]  P Hendrik Pretorius,et al.  Introduction to the physics of molecular imaging with radioactive tracers in small animals , 2002, Journal of cellular biochemistry. Supplement.

[34]  Koon-Pong Wong,et al.  An investigation of coded aperture imaging for small animal SPECT , 2000 .

[35]  J. L. MARIGNIER Historical light on photography , 1990, Nature.

[36]  B. Lengeler,et al.  A compound refractive lens for focusing high-energy X-rays , 1996, Nature.

[37]  G. Entine,et al.  A CCD-based detector for SPECT , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[38]  Yong-hwi Pak Combined scintigraphic and radiographic diagnosis of bone and joint diseases , 1994 .

[39]  Y. Yonekura,et al.  Ultra-high resolution SPECT system using four pinhole collimators for small animal studies. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[40]  J. Mallard,et al.  The performance of a gamma camera for the visualization of radioactive isotope in vivo. , 1963, Physics in medicine and biology.

[41]  S. Rose,et al.  Complementary molecular imaging technologies: High resolution SPECT, PET and MRI. , 2006, Drug discovery today. Technologies.

[42]  F. van der Have,et al.  Continuous model of multi-pinhole SPECT devices , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[43]  H. Bradaczek,et al.  APPROACHING REAL X-RAY OPTICS , 2000 .

[44]  Michael F Land,et al.  The spatial resolution of the pinhole eyes of giant clams (Tridacna maxima , 2003, Proceedings of the Royal Society of London. Series B: Biological Sciences.

[45]  J. Booij,et al.  Evaluation of high-resolution pinhole SPECT using a small rotating animal. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[46]  Michael F Land,et al.  The optical structures of animal eyes , 2005, Current Biology.

[47]  F. Beekman,et al.  Photon-counting gamma camera based on an electron-multiplying CCD , 2005, IEEE Transactions on Nuclear Science.

[48]  P. Steele,et al.  A new method of multiplanar emission tomography using a seven pinhole collimator and an Anger scintillation camera. , 1978, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[49]  H. B. Barber APPLICATIONS OF SEMICONDUCTOR DETECTORS TO NUCLEAR MEDICINE , 1999 .

[50]  Alain Seret,et al.  180° Pinhole SPET with a tilted detector and OS-EM reconstruction: phantom studies and potential clinical applications , 2001, European Journal of Nuclear Medicine.

[51]  Hyunki Kim,et al.  SPECT Imager Design and Data-Acquisition Systems , 2005 .

[52]  Yong Whee Bahk Combined Scintigraphic and Radiographic Diagnosis of Bone and Joint Diseases , 1994 .

[53]  H. B. Barber,et al.  Ultra-high-resolution brain SPECT imaging: simulation results , 1992, IEEE Conference on Nuclear Science Symposium and Medical Imaging.

[54]  D E COPELAND,et al.  Pinhole camera for gamma-ray sources. , 1949, Nucleonics.

[55]  H H Barrett,et al.  A stationary hemispherical SPECT imager for three-dimensional brain imaging. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[56]  B. Hasegawa,et al.  A multipinhole small animal SPECT system with submillimeter spatial resolution. , 2006, Medical physics.

[57]  Alain Seret,et al.  Body contour 180° pinhole SPET with or without tilted detector: a phantom study , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[58]  K. Itoh,et al.  Intraperitoneal leakage of technetium-99m-DTPA following renal transplantation: a sign of chylous ascites. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[59]  Simon R. Cherry,et al.  Towards in vivo nuclear microscopy: iodine-125 imaging in mice using micro-pinholes , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[60]  F. Beekman,et al.  Design and simulation of a high-resolution stationary SPECT system for small animals. , 2004, Physics in medicine and biology.

[61]  E. Hoffman,et al.  Performance evaluation of A-SPECT: a high resolution desktop pinhole SPECT system for imaging small animals , 2002 .

[62]  Tobias Funk,et al.  Small-animal radionuclide imaging with focusing gamma-ray optics , 2003, SPIE Optics + Photonics.

[63]  C. Tavernier,et al.  Bone scintigraphy equipped with a pinhole collimator for diagnosis of avascular necrosis of the femoral head , 1997, Clinical Rheumatology.

[64]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[65]  P. Acton,et al.  Quantification of dopamine transporters in the mouse brain using ultra-high resolution single-photon emission tomography , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[66]  Roberto Accorsi,et al.  Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging—a simulation study , 2005, Physics in medicine and biology.

[67]  N. Schramm,et al.  High-resolution SPECT using multi-pinhole collimation , 2002, IEEE Nuclear Science Symposium Conference Record.

[68]  Scott D. Metzler,et al.  Analytic determination of the resolution-equivalent effective diameter of a pinhole collimator , 2004, IEEE Transactions on Medical Imaging.

[69]  B.H. Hasegawa,et al.  ECG-gated pinhole SPECT in mice , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[70]  F. Beekman,et al.  Photon-counting versus an integrating CCD-based gamma camera: important consequences for spatial resolution , 2005, Physics in medicine and biology.

[71]  Miles N. Wernick,et al.  CHAPTER 21 – Iterative Image Reconstruction , 2004 .

[72]  Brian F. Hutton,et al.  A clinical perspective of accelerated statistical reconstruction , 1997, European Journal of Nuclear Medicine.

[73]  Copeland De,et al.  Pinhole camera for gamma-ray sources. , 1949 .

[74]  Patrick Dupont,et al.  Characterization of pinhole SPECT acquisition geometry , 2003, IEEE Transactions on Medical Imaging.

[75]  Yang Soong,et al.  Foil X-ray mirrors , 1996 .

[76]  H. B. Barber,et al.  Ultra-high-resolution brain SPECT imaging: simulation results , 1992, IEEE Conference on Nuclear Science Symposium and Medical Imaging.

[77]  Ronald J. Jaszczak,et al.  Analytic determination of the pinhole collimator's point-spread function and RMS resolution with penetration , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[78]  B. Vastenhouw,et al.  System calibration and statistical image reconstruction for sub-mm stationary pinhole SPECT , 2005, IEEE Nuclear Science Symposium Conference Record, 2005.

[79]  A. Wouters,et al.  Direct method of decoding multiple images. , 1973, Applied optics.

[80]  H. B. Barber,et al.  Tomographic small-animal imaging using a high-resolution semiconductor camera , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[81]  P M Wanet,et al.  Physical and clinical evaluation of high-resolution thyroid pinhole tomography. , 1996, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[82]  S. Meikle,et al.  Small animal SPECT and its place in the matrix of molecular imaging technologies , 2005, Physics in medicine and biology.