Voxel-based dose reconstruction for total body irradiation with helical tomotherapy.

PURPOSE We have developed a megavoltage CT (MVCT)-based dose reconstruction strategy for total body irradiation (TBI) with helical TomoTherapy (HT) using a deformable registration model to account for the patient's interfraction changes. The proposed technique serves as an efficient tool for delivered dose verification and, potentially, plan adaptation. METHODS AND MATERIALS Four patients with acute myelogenous leukemia treated with TBI using HT were selected for this study. The prescription was 12 Gy, 2 Gy/fraction, twice per day, given at least 6 h apart. The original plan achieved coverage of 80% of the clinical target volume (CTV) by the 12 Gy isodose surface. MVCTs were acquired prior to each treatment. Regions of interest were contoured on each MVCT. The dose for each fraction was calculated based on the MVCT using the HT planned adaptive station. B-spline deformable registration was conducted to establish voxel-to-voxel correspondence between the MVCT and the planning CT. The resultant deformation vector was employed to map the reconstructed dose from each fraction to the same point as the plan dose, and a voxel-to-voxel summed dose from all six fractions was obtained. The reconstructed dose distribution and its dosimetric parameters were compared with those of the original treatment plan. RESULTS While changes in CTV contours occurred in all patients, the reconstructed dose distribution showed that the dose-volume histogram for CTV coverage was close (<1.5%) to that of the original plan. For sensitive structures, the differences between the reconstructed and the planned doses were less than 3.0%. CONCLUSION Voxel-based dose reconstruction strategy that takes into account interfraction anatomical changes using MVCTs is a powerful tool for treatment verification of the delivered doses. This proposed technique can also be applied to adaptive TBI therapy using HT.

[1]  G. Marchal,et al.  Multi-modal volume registration by maximization of mutual information , 1997 .

[2]  Brian B. Avants,et al.  Directly Manipulated Free-Form Deformation Image Registration , 2009, IEEE Transactions on Image Processing.

[3]  Jie Yang,et al.  Monte Carlo based IMRT dose verification using MLC log files and R/V outputs. , 2006, Medical physics.

[4]  K. Brock Results of a multi-institution deformable registration accuracy study (MIDRAS). , 2010, International journal of radiation oncology, biology, physics.

[5]  J. Mazziotta,et al.  Rapid Automated Algorithm for Aligning and Reslicing PET Images , 1992, Journal of computer assisted tomography.

[6]  Lei Dong,et al.  Patient-specific point dose measurement for IMRT monitor unit verification. , 2003, International journal of radiation oncology, biology, physics.

[7]  A Collignon,et al.  Automated multimodality image registration using information theory , 1995 .

[8]  Pierre Hellier,et al.  Level Set Methods in an EM Framework for Shape Classification and Estimation , 2004, International Conference on Medical Image Computing and Computer-Assisted Intervention.

[9]  An Liu,et al.  Image-guided total-marrow irradiation using helical tomotherapy in patients with multiple myeloma and acute leukemia undergoing hematopoietic cell transplantation. , 2009, International journal of radiation oncology, biology, physics.

[10]  Lei Xing,et al.  Automated contour mapping with a regional deformable model. , 2008, International journal of radiation oncology, biology, physics.

[11]  Sven Kabus,et al.  Estimation of Organ Motion from 4D CT for 4D Radiation Therapy Planning of Lung Cancer , 2004, MICCAI.

[12]  E. Podgoršak,et al.  Total body irradiation with a reconditioned cobalt teletherapy unit , 2006, Journal of applied clinical medical physics.

[13]  L. Xing,et al.  Retrospective IMRT dose reconstruction based on cone-beam CT and MLC log-file. , 2008, International journal of radiation oncology, biology, physics.

[14]  Lei Dong,et al.  Dosimetric accuracy of Kodak EDR2 film for IMRT verifications. , 2005, Medical physics.

[15]  Luc Beaulieu,et al.  Commissioning and evaluation of an extended SSD photon model for PINNACLE3 : An application to total body irradiation. , 2009, Medical physics.

[16]  Charles S Mayo,et al.  A systematic benchmark method for analysis and comparison of IMRT treatment planning algorithms. , 2003, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[17]  Jorge Nocedal,et al.  On the limited memory BFGS method for large scale optimization , 1989, Math. Program..

[18]  Yaoqin Xie,et al.  Auto-propagation of contours for adaptive prostate radiation therapy , 2008, Physics in medicine and biology.

[19]  J. Mazziotta,et al.  MRI‐PET Registration with Automated Algorithm , 1993, Journal of computer assisted tomography.

[20]  Indrin J Chetty,et al.  Analysis of deformable image registration accuracy using computational modeling. , 2010, Medical physics.

[21]  Timothy E Schultheiss,et al.  Image-guided total marrow and total lymphatic irradiation using helical tomotherapy. , 2007, International journal of radiation oncology, biology, physics.

[22]  Zhe Chen,et al.  A Constrained Non-rigid Registration Algorithm for Use in Prostate Image-Guided Radiotherapy , 2008, MICCAI.

[23]  Kai Ding,et al.  A cubic B-spline-based hybrid registration of lung CT images for a dynamic airway geometric model with large deformation , 2011, Physics in medicine and biology.

[24]  Quan Chen,et al.  Objective assessment of deformable image registration in radiotherapy: a multi-institution study. , 2007, Medical physics.

[25]  Joseph O Deasy,et al.  CERR: a computational environment for radiotherapy research. , 2003, Medical physics.

[26]  Guy Marchal,et al.  Automated multi-modality image registration based on information theory , 1995 .

[27]  E. Moros,et al.  Pediatric craniospinal axis irradiation with helical tomotherapy: patient outcome and lack of acute pulmonary toxicity. , 2009, International journal of radiation oncology, biology, physics.

[28]  Guy Marchal,et al.  Automated multi-moda lity image registration based on information theory , 1995 .

[29]  Ole Vilhelm Larsen,et al.  Medical Imaging 2001 : Image Processing, 19-22 February 2001, San Diego, CA, USA , 2001 .

[30]  L. Xing,et al.  Image interpolation in 4D CT using a BSpline deformable registration model. , 2006, International journal of radiation oncology, biology, physics.