An improved Monte-Carlo model of the Varian EPID separating support arm and rear-housing backscatter

Previous investigators of EPID dosimetric properties have ascribed the backscatter, that contaminates dosimetric EPID images, to its supporting arm. Accordingly, Monte-Carlo (MC) EPID models have approximated the backscatter signal from the layers under the detector and the robotic support arm using either uniform or non-uniform solid water slabs, or through convolutions with back-scatter kernels. The aim of this work is to improve the existent MC models by measuring and modelling the separate backscatter contributions of the robotic arm and the rear plastic housing of the EPID. The EPID plastic housing is non-uniform with a 11.9 cm wide indented section that runs across the cross-plane direction in the superior half of the EPID which is 1.75 cm closer to the EPID sensitive layer than the rest of the housing. The thickness of the plastic housing is 0.5 cm. Experiments were performed with and without the housing present by removing all components of the EPID from the housing. The robotic support arm was not present for these measurements. A MC model of the linear accelerator and the EPID was modified to include the rear-housing indentation and results compared to the measurement. The rear housing was found to contribute a maximum of 3% additional signal. The rear housing contribution to the image is non-uniform in the in-plane direction with 2% asymmetry across the central 20 cm of an image irradiating the entire detector. The MC model was able to reproduce this non-uniform contribution. The EPID rear housing contributes a non-uniform backscatter component to the EPID image, which has not been previously characterized. This has been incorporated into an improved MC model of the EPID.

[1]  P. Greer,et al.  Development and testing of an improved dosimetry system using a backscatter shielded electronic portal imaging device. , 2012, Medical physics.

[2]  Pejman Rowshanfarzad,et al.  Measurement and modeling of the effect of support arm backscatter on dosimetry with a varian EPID. , 2010, Medical physics.

[3]  R. Mohan,et al.  Monte Carlo computation of dosimetric amorphous silicon electronic portal images. , 2004, Medical physics.

[4]  Peter B Greer,et al.  An energy fluence-convolution model for amorphous silicon EPID dose prediction. , 2009, Medical physics.

[5]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  C. Ma,et al.  BEAM: a Monte Carlo code to simulate radiotherapy treatment units. , 1995, Medical physics.

[7]  J. Siebers,et al.  Verification of the optimal backscatter for an aSi electronic portal imaging device , 2004, Physics in medicine and biology.

[8]  J. Siebers,et al.  Investigation of the optimal backscatter for an aSi electronic portal imaging device , 2004, Physics in Medicine and Biology.

[9]  M B Aufderheide,et al.  Blurring artifacts in megavoltage radiography with a flat-panel imaging system: comparison of Monte Carlo simulations with measurements. , 2002, Medical physics.

[10]  E Spezi,et al.  An investigation of the accuracy of Monte Carlo portal dosimetry for verification of IMRT with extended fields , 2010, Physics in medicine and biology.

[11]  J. Siebers,et al.  Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers. , 2009, Medical physics.