Validation of blood pressure monitoring using pulse transit time in heart failure patients with Cheyne–Stokes respiration undergoing adaptive servoventilation therapy

PurposeUsing pulse transit time (PTT) and an ECG appears to be a promising alternative for invasive or noninvasive monitoring of blood pressure (BP). This study assessed the validity of PTT for BP monitoring in clinical practice.MethodsTwenty-nine patients with chronic heart failure (HF; 27 male, 70.5 ± 9.9 years) and nocturnal Cheyne–Stokes respiration were noninvasively ventilated for one hour using adaptive servoventilation (ASV) therapy (PaceWave™, ResMed). BP was measured using two devices (oscillometrically via Task Force Monitor™, CNSystems and PTT via SOMNOscreen™, Somnomedics) at least every 7 min for 30 min before, during, and after ASV.ResultsMean systolic BP was 118.1 ± 14.4 mmHg vs. 115.9 ± 14.1 mmHg for oscillometric method vs PTT, respectively. Corresponding values for diastolic BP were 72.3 ± 10.3 mmHg and 69.4 ± 11.1 mmHg. While clinically comparable, differences between the two methods were statistically significant (p < 0.05). The difference between the two methods showed an increasing trend over time. A total of 18.5 % of PTT-based measurements could not be analyzed. The direction of a change in BP was opposite for PTT vs oscillometry for 17.0 % and 32.8 % of systolic and diastolic BP measurements, respectively.ConclusionsWhen monitoring BP in HF patients, overall BP monitoring using PTT is comparable to oscillometry for a period of 2 h (including a 1-h ASV phase). However, PTT shows a tendency to underestimate BP over time and during ASV.

[1]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[2]  Emma Pickwell-MacPherson,et al.  The acute effects of running on blood pressure estimation using pulse transit time in normotensive subjects , 2009, European Journal of Applied Physiology.

[3]  C. Lim,et al.  Evaluation of blood pressure changes using vascular transit time , 2006, Physiological measurement.

[4]  U. Roggenbuck,et al.  [The Stamp of Quality (Prüfsiegel) of the German Hypertension League for the clinical validation of blood pressure measuring devices. Results from the testing of 51 devices]. , 2006, Deutsche Medizinische Wochenschrift.

[5]  James B Seward,et al.  Central arterial pressure and arterial pressure pulse: new views entering the second century after Korotkov. , 2006, Mayo Clinic proceedings.

[6]  J. Bruner,et al.  Comparison of direct and indirect measuring arterial blood pressure. , 1981, Medical instrumentation.

[7]  M. Safar,et al.  Aortic pulse wave velocity: an independent marker of cardiovascular risk. , 2002, The American journal of geriatric cardiology.

[8]  P. Heidenreich,et al.  Severe obstructive sleep apnea increases mortality in patients with ischemic heart disease and myocardial injury , 2013, Sleep and Breathing.

[9]  M. Anlauf,et al.  Das Prüfsiegelprotokoll der Deutschen Hochdruckliga zur klinischen Validierung von Blutdruckmessgeräten , 2006 .

[10]  J. Bruner,et al.  Comparison of direct and indirect methods of measuring arterial blood pressure, part II. , 1981, Medical instrumentation.

[11]  P. Ducimetiere,et al.  Aortic Stiffness Is an Independent Predictor of All-Cause and Cardiovascular Mortality in Hypertensive Patients , 2001, Hypertension.

[12]  S. Eckert,et al.  [Validation of continuous blood pressure measurements by pulse transit time: a comparison with invasive measurements in a cardiac intensive care unit]. , 2010, Deutsche medizinische Wochenschrift.

[13]  Michael Böhm,et al.  Relationship between pulse transit time and blood pressure is impaired in patients with chronic heart failure , 2010, Clinical Research in Cardiology.

[14]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[15]  Siegfried Eckert,et al.  Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation , 2012, Sleep and Breathing.

[16]  Nima Ghasemzadeh,et al.  A Brief Journey into the History of the Arterial Pulse , 2011, Cardiology research and practice.

[17]  C. Young,et al.  Clinical evaluation of continuous noninvasive blood pressure monitoring: Accuracy and tracking capabilities , 1995, Journal of Clinical Monitoring.

[18]  Kozo Hirata,et al.  Pulse wave analysis and pulse wave velocity: a review of blood pressure interpretation 100 years after Korotkov. , 2006, Circulation journal : official journal of the Japanese Circulation Society.

[19]  J. Mackenzie Principles of Diagnosis and Treatment in Heart Affections , 1916, Bristol Medico-Chirurgical Journal (1883).

[20]  J. Stradling,et al.  Effect of measuring ambulatory blood pressure on sleep and on blood pressure during sleep , 1994, BMJ.

[21]  Nikolaus J Buchner,et al.  Treatment of obstructive sleep apnea reduces arterial stiffness , 2012, Sleep and Breathing.

[22]  C. Caceres,et al.  A computer program for the peripheral pulse wave. , 1967, The American journal of cardiology.

[23]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[24]  S. Eckert,et al.  Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exercise , 2002, Blood pressure monitoring.

[25]  D. Horstkotte,et al.  Hypotensive effects of positive airway pressure ventilation in heart failure patients with sleep-disordered breathing , 2012, Sleep and Breathing.

[26]  T. Togawa,et al.  Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration , 2000, Medical and Biological Engineering and Computing.

[27]  J. Concato,et al.  Obstructive sleep apnea as a risk factor for coronary events or cardiovascular death , 2010, Sleep and Breathing.

[28]  J. Naschitz,et al.  Pulse Transit Time by R-Wave-Gated Infrared Photoplethysmography: Review of the Literature and Personal Experience , 2004, Journal of Clinical Monitoring and Computing.

[29]  H. Hartung,et al.  [Monoclonal antibodies improve therapy of relapsing multiple sclerosis. Molecular basis and clinical results of anti-VLA4 (natalizumab) therapy]. , 2006, Deutsche Medizinische Wochenschrift.

[30]  Li-li Han,et al.  Mal-effects of obstructive sleep apnea on the heart , 2012, Sleep and Breathing.

[31]  A. Campbell,et al.  Pulse wave analysis in a pilot randomised controlled trial of auto-adjusting and continuous positive airway pressure for obstructive sleep apnoea , 2011, Sleep and Breathing.

[32]  H E HOFF,et al.  The Pulse Wave Velocity as an Early Indicator of Atherosclerosis in Diabetic Subjects , 1962, Circulation.

[33]  M. O'Rourke,et al.  The arterial pulse in health and disease. , 1971, American heart journal.

[34]  R. Collins,et al.  Blood pressure, stroke, and coronary heart disease Part 2, short-term reductions in blood pressure: overview of randomised drug trials in their epidemiological context , 1990, The Lancet.

[35]  R. Collins,et al.  Blood pressure, stroke, and coronary heart disease Part 1, prolonged differences in blood pressure: prospective observational studies corrected for the regression dilution bias , 1990, The Lancet.

[36]  Yutaka Imai,et al.  Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults , 2002, Blood pressure monitoring.

[37]  M. Nakayama,et al.  How much influence does inspiration have on pulse transit time in sleep apnea? , 2008, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[38]  S. Quan,et al.  Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. , 2012, Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine.

[39]  P. Wach,et al.  Non-invasive beat-to-beat cardiac output monitoring by an improved method of transthoracic bioimpedance measurement , 2006, Comput. Biol. Medicine.

[40]  E. Nye,et al.  THE EFFECT OF BLOOD PRESSURE ALTERATION ON THE PULSE WAVE VELOCITY , 1964, British heart journal.