System Calibration and Image Reconstruction for a New Small-Animal SPECT System

A novel small-animal SPECT imager, FastSPECT II, was recently developed at the Center for Gamma-Ray Imaging. FastSPECT II consists of two rings of eight modular scintillation cameras and list-mode data-acquisition electronics that enable stationary and dynamic imaging studies. The instrument is equipped with exchangeable aperture assemblies and adjustable camera positions for selections of magnifications, pinhole sizes, and fields of view (FOVs). The purpose of SPECT imaging is to recover the radiotracer distribution in the object from the measured image data. Accurate knowledge of the imaging system matrix (referred to as H) is essential for image reconstruction. To assure that all of the system physics is contained in the matrix, experimental calibration methods for the individual cameras and the whole imaging system were developed and carefully performed. The average spatial resolution over the FOV of FastSPECT II in its lowmagnification (2.4X) configuration is around 2.4 mm, computed from the Fourier crosstalk matrix. The system sensitivity measured with a Tc point source at the center of the FOV is about 267 cps/MBq. The system detectability was evaluated by computing the ideal-observer performance on SKE/BKE (signal-known-exactly/background-knownexactly) detection tasks. To reduce the system-calibration time and achieve finer reconstruction grids, two schemes for interpolating H were implemented and compared: these are centroid interpolation with Gaussian fitting and Fourier interpolation. Reconstructed phantom and mouse-cardiac images demonstrated the effectiveness of the H-matrix interpolation.

[1]  Donald W. Wilson,et al.  SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. , 2006, Medical physics.

[2]  Tomio Inoue,et al.  Myocardial infarction in rats: high-resolution single-photon emission tomographic imaging with a pinhole collimator , 1996, European Journal of Nuclear Medicine.

[3]  Harrison H. Barrett,et al.  Detectors for Small-Animal SPECT II , 2005 .

[4]  J C Chen,et al.  Scatter rejection in modular gamma cameras for use in dynamic 3D spect brain imaging system. , 1997, Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society.

[5]  P. Voorn,et al.  Imaging of dopamine transporters in rats using high-resolution pinhole single-photon emission tomography , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[6]  J. Narula,et al.  Targeting human breast tumour in xeno-grafted SCID mice with 99Tcm-glucarate. , 1997, Nuclear medicine communications.

[7]  D. R. Allen,et al.  Tc-99m methylene diphosphonate versus Tc-99m pyrophosphate: biologic and clinical comparison. , 1977, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  H H Barrett,et al.  A stationary hemispherical SPECT imager for three-dimensional brain imaging. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  D. L. Kirch,et al.  Advantages Of List-Mode Acquisition Of Dynamic Cardiac Data , 1997 .

[10]  Guimin Zhang,et al.  Integrated CT-SPECT system for small-animal imaging , 2000, SPIE Optics + Photonics.

[11]  Jeffrey A. Fessler Penalized weighted least-squares image reconstruction for positron emission tomography , 1994, IEEE Trans. Medical Imaging.

[12]  Donald W. Wilson,et al.  FastSPECT II: a second-generation high-resolution dynamic SPECT imager , 2002, IEEE Transactions on Nuclear Science.

[13]  D S Lalush,et al.  Performance of ordered-subset reconstruction algorithms under conditions of extreme attenuation and truncation in myocardial SPECT. , 2000, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  I Buvat,et al.  Monte Carlo simulations in SPET and PET. , 2002, The quarterly journal of nuclear medicine : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology.

[15]  Alessandra Boschi,et al.  YAP-(S)PET small animal scanner: Quantitative results , 2003 .

[16]  P. Acton,et al.  Occupancy of dopamine D2 receptors in the mouse brain measured using ultra-high-resolution single-photon emission tomography and [123I]IBF , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[17]  A. Lent,et al.  Iterative algorithms for large partitioned linear systems, with applications to image reconstruction , 1981 .

[18]  H. Atkins,et al.  Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[19]  T Yokoi,et al.  [Development of a high-resolution pinhole SPECT system using dual-head gamma camera for small animal studies]. , 1998, Kaku igaku. The Japanese journal of nuclear medicine.

[20]  E. Cisbani,et al.  Small animal imaging by single photon emission using pinhole and coded aperture collimation , 2005, IEEE Transactions on Nuclear Science.

[21]  Hiroyuki Okada,et al.  A high resolution animal PET scanner using compact PS-PMT detectors , 1996 .

[22]  Thomas D Milster,et al.  Design and Construction of a Modular Gamma Camera , 1987 .

[23]  V. Perez-Mendez,et al.  A method of tomographic imaging using a multiple pinhole-coded aperture. , 1974, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[24]  H. B. Dietrich,et al.  Kinetics of Self-Trapped Holes in Alkali-Halide Crystals: Experiments in NaI(Tl) and KI(Tl) , 1973 .

[25]  Markus Cremer,et al.  Evaluation of the TierPET system , 1998 .

[26]  D Paix,et al.  Pinhole imaging of gamma rays. , 1967, Physics in medicine and biology.

[27]  M. Schmand,et al.  Design of a small animal PET imaging system with 1 microliter volume resolution , 2004, IEEE Transactions on Nuclear Science.

[28]  B. Cassen EVOLUTION OF SCINTILLATION SCANNING. , 1969 .

[29]  R S Balaban,et al.  Challenges in small animal noninvasive imaging. , 2001, ILAR journal.

[30]  B. Hesse,et al.  Radionuclide tracers in the evaluation of resting myocardial ischaemia and viability , 1997, European Journal of Nuclear Medicine.

[31]  R. Malaguti,et al.  Preliminary test results on the new electronic readout of the YAP(S)PET small animal scanner , 2001 .

[32]  Koon-Pong Wong,et al.  An investigation of coded aperture imaging for small animal SPECT , 2000 .

[33]  S Weber,et al.  The design of an animal PET: flexible geometry for achieving optimal spatial resolution or high sensitivity , 1997, IEEE Transactions on Medical Imaging.

[34]  C.A. Bouman,et al.  Quantitative comparison of FBP, EM, and Bayesian reconstruction algorithms for the IndyPET scanner , 2003, IEEE Transactions on Medical Imaging.

[35]  P. Steele,et al.  A new method of multiplanar emission tomography using a seven pinhole collimator and an Anger scintillation camera. , 1978, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[36]  R. Weissleder,et al.  A coded aperture for high-resolution nuclear medicine planar imaging with a conventional Anger camera: experimental results , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[37]  H. B. Dietrich,et al.  Kinetics of the diffusion of self-trapped holes in alkali halide scintillators , 1972 .

[38]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[39]  M. Pillai,et al.  Evolution of Tc-99m in diagnostic radiopharmaceuticals. , 2001, Seminars in nuclear medicine.

[40]  J.A. Correia,et al.  Performance evaluation of MMP-II:A second-generation small animal PET , 2004, IEEE Transactions on Nuclear Science.

[41]  Yuan-Chuan Tai,et al.  Instrumentation aspects of animal PET. , 2005, Annual review of biomedical engineering.

[42]  Bruce D. Smith Image Reconstruction from Cone-Beam Projections: Necessary and Sufficient Conditions and Reconstruction Methods , 1985, IEEE Transactions on Medical Imaging.

[43]  M. Rafecas,et al.  Characterization and readout of MADPET-II detector modules: validation of a unique design concept for high resolution small animal PET , 2005, IEEE Transactions on Nuclear Science.

[44]  R. F. Wagner,et al.  Objective assessment of image quality. II. Fisher information, Fourier crosstalk, and figures of merit for task performance. , 1995, Journal of the Optical Society of America. A, Optics, image science, and vision.

[45]  L. Landweber An iteration formula for Fredholm integral equations of the first kind , 1951 .

[46]  Paul D Acton,et al.  Small animal imaging with high resolution single photon emission tomography. , 2003, Nuclear medicine and biology.

[47]  E E Fenimore,et al.  Tomographical imaging using uniformly redundant arrays. , 1979, Applied optics.

[48]  A Kobayashi,et al.  A study of the multiple pinhole coded aperture and the application of the minicomputer in image decoding. , 1979, Radioisotopes.

[49]  R. G. Simpson,et al.  A Modular Scintillation Camera for Use in Nuclear Medicine , 1984, IEEE Transactions on Nuclear Science.

[50]  F D Thomas,et al.  Technetium-99m-methylene diphosphonate--a superior agent for skeletal imaging: comparison with other technetium complexes. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[51]  R. Jaszczak,et al.  Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT. , 1994, Physics in medicine and biology.

[52]  G. Hutchins,et al.  Design and characterization of IndyPET-II:A high-resolution, high-sensitivity dedicated research scanner , 2003 .

[53]  H H Barrett,et al.  A full-field modular gamma camera. , 1990, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[54]  R. F. Wagner,et al.  Unified SNR analysis of medical imaging systems , 1985, Physics in medicine and biology.

[55]  R. F. Wagner,et al.  Overview of a Unified SNR Analysis of Medical Imaging Systems , 1982, IEEE Transactions on Medical Imaging.

[56]  A.J. Da Silva,et al.  An ultra high resolution ECG-gated myocardial imaging system for small animals , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[57]  A. Kak,et al.  Simultaneous Algebraic Reconstruction Technique (SART): A Superior Implementation of the Art Algorithm , 1984, Ultrasonic imaging.

[58]  H H Barrett,et al.  Cone-beam tomography with discrete data sets. , 1994, Physics in medicine and biology.

[59]  H. Tuy AN INVERSION FORMULA FOR CONE-BEAM RECONSTRUCTION* , 1983 .

[60]  Ronald J. Jaszczak,et al.  Molecular imaging of small animals with a triple-head SPECT system using pinhole collimation , 2005, IEEE Transactions on Medical Imaging.

[61]  Alan J. Fischman,et al.  Development of a small animal PET imaging device with resolution approaching 1 mm , 1999 .

[62]  Scott D. Metzler,et al.  Analytic determination of the resolution-equivalent effective diameter of a pinhole collimator , 2004, IEEE Transactions on Medical Imaging.

[63]  G. Herman,et al.  Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. , 1970, Journal of theoretical biology.

[64]  James F. Young,et al.  MicroPET: a high resolution PET scanner for imaging small animals , 1996, IEEE Nuclear Science Symposium Conference Record.

[65]  J. M. Ollinger,et al.  Positron Emission Tomography , 2018, Handbook of Small Animal Imaging.

[66]  R. E. Burgess,et al.  Homophase and heterophase fluctuations in semiconducting crystals , 1959 .

[67]  Roger Lecomte,et al.  Initial results from the Sherbrooke avalanche photodiode positron tomograph , 1996 .

[68]  Duccio Volterrani,et al.  A pilot clinical trial on scintimammography (SMM) with 99mTc-D-Glucaric Acid in patients with breast cancer , 2002 .

[69]  H H Barrett,et al.  Effect of random background inhomogeneity on observer detection performance. , 1992, Journal of the Optical Society of America. A, Optics and image science.

[70]  Harrison H Barrett,et al.  99mTc glucarate high-resolution imaging of drug sensitive and drug resistant human breast cancer xenografts in SCID mice , 2004, Nuclear medicine communications.

[71]  J. Narula,et al.  Very early noninvasive detection of acute experimental nonreperfused myocardial infarction with 99mTc-labeled glucarate. , 1997, Circulation.

[72]  L.R. Furenlid,et al.  Real-time data acquisition and maximum-likelihood estimation for gamma cameras , 2005, 14th IEEE-NPSS Real Time Conference, 2005..

[73]  D L Kirch,et al.  Seven-pinhole tomography--a technical description. , 1981, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[74]  Jyhcheng. Chen,et al.  Modular gamma cameras: Improvements in scatter rejection, and characterization and initial clinical application. , 1995 .

[75]  S S Gambhir,et al.  Use of positron emission tomography in animal research. , 2001, ILAR journal.

[76]  P Wollmer,et al.  Pinhole emission computed tomography: method and experimental evaluation. , 1990, Physics in medicine and biology.

[77]  E. Hoffman,et al.  Performance evaluation of A-SPECT: a high resolution desktop pinhole SPECT system for imaging small animals , 2002 .

[78]  L. Lucy An iterative technique for the rectification of observed distributions , 1974 .

[79]  H. Barrett,et al.  Objective assessment of image quality. III. ROC metrics, ideal observers, and likelihood-generating functions. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.

[80]  Bruce H. Hasegawa,et al.  ECG-gated pinhole SPECT in mice with millimeter spatial resolution , 2000 .

[81]  Timothy Joseph. Roney Coded-aperture transaxial tomography using modular gamma cameras. , 1989 .

[82]  W.P. Klein,et al.  FASTSPECT: electrical and mechanical design of a high-resolution dynamic SPECT imager , 1995, 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference Record.

[83]  Eric Clarkson,et al.  Experimental determination of object statistics from noisy images. , 2003, Journal of the Optical Society of America. A, Optics, image science, and vision.

[84]  J Fessler,et al.  Spatial resolution and noise tradeoffs in pinhole imaging system design: a density estimation approach. , 1998, Optics express.

[85]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[86]  P. Grangeat Mathematical framework of cone beam 3D reconstruction via the first derivative of the radon transform , 1991 .

[87]  T J Spinks,et al.  Three-dimensional performance of a small-diameter positron emission tomograph. , 1997, Physics in medicine and biology.

[88]  Norio Miura,et al.  Phosphors for X-ray and ionizing radiation , 2006 .

[89]  E. Hoffman,et al.  Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. , 1999, Physics in medicine and biology.

[90]  Harrison H. Barrett,et al.  Modular Scintillation Cameras: A Progress Report , 1988, Medical Imaging.

[91]  John Missimer,et al.  Performance evaluation of the 16-module quad-HIDAC small animal PET camera. , 2004, Physics in medicine and biology.

[92]  Miles N. Wernick,et al.  CHAPTER 21 – Iterative Image Reconstruction , 2004 .

[93]  R. F. Wagner,et al.  Aperture optimization for emission imaging: effect of a spatially varying background. , 1990, Journal of the Optical Society of America. A, Optics and image science.

[94]  J. B. Birks,et al.  The Theory and Practice of Scintillation Counting , 1965 .

[95]  C. Cork,et al.  A germanium-based, coded aperture imager , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[96]  Brandon Dominic Gallas Signal detection in lumpy backgrounds , 2001 .

[97]  H. B. Barber,et al.  Compact CT/SPECT small-animal imaging system , 2002, IEEE Transactions on Nuclear Science.

[98]  C A Jayachandran,et al.  Calculated effective atomic number and kerma values for tissue-equivalent and dosimetry materials. , 1971, Physics in medicine and biology.

[99]  P. Gilbert Iterative methods for the three-dimensional reconstruction of an object from projections. , 1972, Journal of theoretical biology.

[100]  H H Barrett,et al.  List-mode likelihood. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[101]  John Nathan. Aarsvold Multiple-pinhole transaxial tomography: A model and analysis. , 1993 .

[102]  Herbert N. Hersh,et al.  The Scintillation Mechanism in Thallium Doped Alkali Halides , 1970 .

[103]  Eric C. Frey,et al.  Pinhole SPECT of mice using the LumaGEM gamma camera , 2001 .

[104]  Donald W. Wilson,et al.  Calibration of Scintillation Cameras and Pinhole SPECT Imaging Systems , 2005 .

[105]  R. Wojcik,et al.  Performance evaluation of a multipinhole small animal SPECT system , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[106]  H H Barrett,et al.  Fresnel zone plate imaging of gamma rays; theory. , 1973, Applied optics.

[107]  Eiji Sakai,et al.  Recent Measurements on Scintillator-Photodetector Systems , 1987, IEEE Transactions on Nuclear Science.

[108]  Stefan Eberl,et al.  A prototype coded aperture detector for small animal SPECT , 2001 .

[109]  R. Fisher,et al.  On the Mathematical Foundations of Theoretical Statistics , 1922 .

[110]  Magdalena Rafecas,et al.  A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals , 2001, European Journal of Nuclear Medicine.

[111]  K. Erlandsson,et al.  Small animal imaging with pinhole single‐photon emission computed tomography , 1994, Cancer.

[112]  C. Metz Basic principles of ROC analysis. , 1978, Seminars in nuclear medicine.

[113]  Y. Yonekura,et al.  Ultra-high resolution SPECT system using four pinhole collimators for small animal studies. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[114]  Hyunki Kim SemiSPECT: A small-animal SPECT imager based on eight cadmium zinc tellurium detector arrays , 2004 .

[115]  W C Eckelman,et al.  High resolution PET, SPECT and projection imaging in small animals. , 2001, Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society.

[116]  A. Del Guerra,et al.  An integrated PET-SPECT small animal imager: preliminary results , 1999 .

[117]  Craig K. Abbey,et al.  Practical issues and methodology in assessment of image quality using model observers , 1997, Medical Imaging.

[118]  R Accorsi,et al.  Near-field artifact reduction in planar coded aperture imaging. , 2001, Applied optics.

[119]  G. Knoll Radiation detection and measurement , 1979 .

[120]  H. Barrett,et al.  List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET , 1998, IEEE Transactions on Medical Imaging.

[121]  Hiroaki Takajo,et al.  Noniterative method for obtaining the exact solution for the normal equation in least-squares phase estimation from the phase difference , 1988 .

[122]  Rolf Clackdoyle,et al.  CHAPTER 20 – Analytic Image Reconstruction Methods , 2004 .

[123]  Bobby R. Hunt,et al.  Matrix formulation of the reconstruction of phase values from phase differences , 1979 .

[124]  Andre Lehovich,et al.  List-mode SPECT reconstruction using empirical likelihood , 2005 .

[125]  Matthew A. Kupinski,et al.  Objective Assessment of Image Quality , 2005 .

[126]  Harrison H Barrett,et al.  Validating the use of channels to estimate the ideal linear observer. , 2003, Journal of the Optical Society of America. A, Optics, image science, and vision.

[127]  G. Herman,et al.  Algebraic reconstruction techniques using smooth basis functions for helical cone-beam tomography , 2001 .

[128]  K. Murthy,et al.  "ANIPET" a versatile PET scanner for imaging small animals , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[129]  N. Schramm,et al.  High-resolution SPECT using multi-pinhole collimation , 2002, IEEE Nuclear Science Symposium Conference Record.

[130]  M. Gmar,et al.  Development of coded-aperture imaging with a compact gamma camera , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[131]  P. V. van Rijk,et al.  U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[132]  Michael W Dae,et al.  Performance evaluation of a pinhole SPECT system for myocardial perfusion imaging of mice. , 2002, Medical physics.

[133]  A. Jeavons,et al.  A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[134]  S. Kecani,et al.  Operational and performance report on ANIPET, a versatile PET scanner for laboratory animals , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[135]  H. Barrett Fresnel zone plate imaging in nuclear medicine. , 1972, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[136]  Olga I Vassilieva,et al.  Method for reducing background artifacts from images in single-photon emission computed tomography with a uniformly redundant array coded aperture. , 2002, Applied optics.

[137]  Gary D. Hutchins,et al.  IndyPET-a high resolution, high sensitivity dedicated research scanner , 1999, 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019).

[138]  M. SIAMJ.,et al.  BLOCK-ITERATIVE ALGORITHMS WITH DIAGONALLY SCALED OBLIQUE PROJECTIONS FOR THE LINEAR FEASIBILITY PROBLEM , 2002 .

[139]  William H. Richardson,et al.  Bayesian-Based Iterative Method of Image Restoration , 1972 .

[140]  G Mariani,et al.  Detection of acute myocardial infarction by 99mTc-labeled D-glucaric acid imaging in patients with acute chest pain. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[141]  Kevin Anthony Gross Assessing and Optimizing Pinhole SPECT Imaging Systems for Detection Tasks , 2006 .

[142]  R. N. Goble,et al.  Performance evaluation of the microPET P4: a PET system dedicated to animal imaging. , 2001, Physics in medicine and biology.

[143]  S. Siegel,et al.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[144]  H. Anger SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS. , 1964, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[145]  T J Spinks,et al.  The design and physical characteristics of a small animal positron emission tomograph. , 1995, Physics in medicine and biology.

[146]  Harrison H Barrett,et al.  SemiSPECT: a small-animal imaging system based on eight CdZnTe pixel detectors , 2002, 2002 IEEE Nuclear Science Symposium Conference Record.

[147]  Michael D. Tipton The ODCAT: One Dimensional Coded Aperture Tomography , 1978, Optics & Photonics.

[148]  Kyle J Myers,et al.  Objective assessment of image quality. IV. Application to adaptive optics. , 2006, Journal of the Optical Society of America. A, Optics, image science, and vision.

[149]  R. B. Murray Energy Transfer in Alkali Halide Scintillators by Electron-Hole Diffusion and Capture , 1975, IEEE Transactions on Nuclear Science.

[150]  S. Cherry,et al.  MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. , 2003, Physics in medicine and biology.

[151]  Georgios Kastis Multi-modality imaging of small animals , 2002 .

[152]  R. Wojcik,et al.  SPECT-CT system for small animal imaging , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[153]  H. Barrett,et al.  Ideal-observer computation in medical imaging with use of Markov-chain Monte Carlo techniques. , 2003, Journal of the Optical Society of America. A, Optics, image science, and vision.

[154]  D. Weber,et al.  Ultra-high-resolution imaging of small animals: Implications for preclinical and research studies , 1999, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[155]  H. B. Barber,et al.  Tomographic small-animal imaging using a high-resolution semiconductor camera , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[156]  A. Del Guerra,et al.  YAP-PET: first results of a small animal positron emission tomograph based on YAP:Ce finger crystals , 1998 .

[157]  T. M. Cannon,et al.  Coded aperture imaging with uniformly redundant arrays. , 1978, Applied optics.

[158]  S. Cherry,et al.  Performance evaluation of microPET: a high-resolution lutetium oxyorthosilicate PET scanner for animal imaging. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[159]  John David Sain,et al.  Optical modeling, design optimization, and performance analysis of a gamma camera for detection of breast cancer , 2001 .

[160]  A. Del Guerra,et al.  High spatial resolution small animal YAP-PET , 1998 .

[161]  Jagat Narula,et al.  Avidity of technetium 99m glucarate for the necrotic myocardium: In vivo and in vitro assessment , 1997, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[162]  Jorge Llacer Theory of Imaging with a Very Limited Number of Projections , 1979, IEEE Transactions on Nuclear Science.

[163]  Koichi Ogawa,et al.  Ultra high resolution pinhole SPECT for small animal study , 1998 .

[164]  Ronald J. Jaszczak,et al.  Analytic determination of the pinhole collimator's point-spread function and RMS resolution with penetration , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[165]  Louis A. Romero,et al.  Robust two-dimensional weighted and unweighted phase unwrapping that uses fast transforms and iterative methods , 1994 .

[166]  J. Booij,et al.  Evaluation of high-resolution pinhole SPECT using a small rotating animal. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[167]  G. Mageras,et al.  A measurement of the light yield of common inorganic scintillators , 1988 .

[168]  L. Landweber An iteration formula for Fredholm integral equations of the first kind with application to the axially symmetric potential flow about elongated bodies of revolution , 1951 .

[169]  H Zaidi,et al.  Relevance of accurate Monte Carlo modeling in nuclear medical imaging. , 1999, Medical physics.

[170]  Harrison H. Barrett,et al.  Likelihood window, energy window, and Bayesian window for scatter rejection in gamma cameras , 1993, 1993 IEEE Conference Record Nuclear Science Symposium and Medical Imaging Conference.

[171]  C. Le Goaller,et al.  Development of coded-aperture imaging with a compact gamma camera , 2003 .

[172]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[173]  P. Acton,et al.  Quantification of dopamine transporters in the mouse brain using ultra-high resolution single-photon emission tomography , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[174]  R. K. Rowe,et al.  A system for three-dimensional SPECT without motion. , 1991 .

[175]  D. Rubin,et al.  Maximum likelihood from incomplete data via the EM - algorithm plus discussions on the paper , 1977 .