Accuracy of intraocular pressure measurements in dogs using two different tonometers and plano therapeutic soft contact lenses.

OBJECTIVE To compare and evaluate the accuracy of intraocular pressure (IOP) measured through a therapeutic contact lens, using applanation (TonoPen XL(®)) and rebound (TonoVet(®)) tonometers in enucleated dog eyes. ANIMALS STUDIED A total of 30 enucleated eyes from 15 beagle dogs. PROCEDURES To measure accurate IOP, the anterior chamber of each enucleated eye was cannulated with two 26-gauge needles and two polyethylene tubes were connected vertically to an adjustable reservoir bag of normal saline and a pressure transducer. IOP was measured by the TonoPen XL(®) followed by the TonoVet(®) without a contact lens. After a contact lens was applied to the cornea, IOP was re-measured in the same order. Three consecutive IOP measurements were performed using both tonometers. RESULTS Without the contact lens, the IOP values obtained by both tonometers correlated well according to the regression analysis (TonoVet(®): γ(2) = 0.98, TonoPen XL(®): γ(2) = 0.97, P < 0.001). The TonoPen XL(®) consistently underestimated values as transducer IOP increased; however, IOP values measured with the TonoPen XL(®) were in close agreement and were less variable than those determined with the TonoVet(®) when a contact lens was applied to the cornea. Bland-Altman analysis was used to determine the lower and upper limits of agreement (TonoVet(®): -29.7 and +21.1 mmHg, TonoPen XL(®): -3.9 and +3.6 mmHg) between the two devices. CONCLUSIONS This study suggests that the TonoPen XL(®) is a useful tonometer for dogs wearing therapeutic contact lenses, and importantly, contact lenses would not need to be removed prior to IOP measurement.

[1]  K. Seo,et al.  Effect of central corneal thickness on intraocular pressure with the rebound tonometer and the applanation tonometer in normal dogs. , 2011, Veterinary ophthalmology.

[2]  A. Loewenstein,et al.  Intraocular pressure measurements in relation to head position and through soft contact lenses: comparison of three portable instruments. , 2011, Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye.

[3]  M. Hässig,et al.  Evaluation of a rebound tonometer (Tonovet) in clinically normal cat eyes. , 2010, Veterinary ophthalmology.

[4]  E. Chihara Assessment of true intraocular pressure: the gap between theory and practical data. , 2008, Survey of ophthalmology.

[5]  G. Saleh,et al.  Applanation tonometry in silicone hydrogel contact lens wearers. , 2004, Contact lens & anterior eye : the journal of the British Contact Lens Association.

[6]  H. Uusitalo,et al.  Comparison of simultaneous readings of intraocular pressure in rabbits using Perkins handheld, Tono-Pen XL, and TonoVet tonometers , 2007, Graefe's Archive for Clinical and Experimental Ophthalmology.

[7]  M. Boevé,et al.  Comparison of the use of new handheld tonometers and established applanation tonometers in dogs. , 2006, American journal of veterinary research.

[8]  P. Khaw,et al.  Accuracy of intraocular pressure measurements in new zealand white rabbits. , 2005, Investigative ophthalmology & visual science.

[9]  P. Barrett,et al.  Abstract Title: An Evaluation of the TonoVet® Rebound Tonometer for Measuring Intraocular Pressure in Dogs and Horses , 2005 .

[10]  D. Luce Determining in vivo biomechanical properties of the cornea with an ocular response analyzer , 2005, Journal of cataract and refractive surgery.

[11]  W. Illahi,et al.  Non-contact tonometry over soft contact lenses: effect of contact lens power on the measurement of intra-ocular pressure. , 2004, Contact lens & anterior eye : the journal of the British Contact Lens Association.

[12]  D. Tan,et al.  Accurate intraocular pressure measurement in contact lens wearers with normal pressures. , 1997, The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc.

[13]  P. Donshik,et al.  The effects of therapeutic contact lenses on intraocular pressure measurement. , 1996, The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc.

[14]  P. Kaufman,et al.  Intraocular pressure measurement in cynomolgus monkeys. Tono-Pen versus manometry. , 1996, Investigative ophthalmology & visual science.

[15]  C. Murphy,et al.  Intraocular pressure measurement through two types of plano therapeutic soft contact lenses in dogs. , 1995, American journal of veterinary research.

[16]  J. Morrison,et al.  Noninvasive measurement of rat intraocular pressure with the Tono-Pen. , 1993, Investigative ophthalmology & visual science.

[17]  P. Asbell,et al.  Accuracy of Intraocular Pressure Measurements with Two Different Tonometers Through Bandage Contact Lenses , 1992, Cornea.

[18]  L. Janoff Intraocular pressure measurements with the tonopen II over a disposable soft lens , 1991 .

[19]  I. Kurzman,et al.  Evaluation of two applanation tonometers in cats. , 1991, American journal of veterinary research.

[20]  J. Pickett,et al.  Evaluation of two applanation tonometers in horses. , 1990, American journal of veterinary research.

[21]  W. A. Boothe,et al.  Intraocular pressure measurement with the Tono-Pen through soft contact lenses. , 1990, American journal of ophthalmology.

[22]  J. A. Khan,et al.  Tono-Pen estimation of intraocular pressure through bandage contact lenses. , 1989, American journal of ophthalmology.

[23]  C. McMonnies Noncontact Tonometry through Soft Contact Lenses , 1986, American journal of optometry and physiological optics.

[24]  T. Deutsch,et al.  Pneumatonometry through bandage contact lenses. , 1985, Archives of ophthalmology.

[25]  Club Jules Gonin,et al.  Graefe's archive for clinical and experimental ophthalmology , 1982 .

[26]  R. Meyer,et al.  Mackay-Marg tonometry over therapeutic soft contact lenses. , 1978, American journal of ophthalmology.

[27]  Schmidt Tf On applanation tonometry. , 1961 .