Webcam-based flow cytometer using wide-field imaging for low cell number detection at high throughput.

Here we describe a novel low-cost flow cytometer based on a webcam capable of low cell number detection in a large volume which may overcome the limitations of current flow cytometry. Several key elements have been combined to yield both high throughput and high sensitivity. The first element is a commercially available webcam capable of 187 frames per second video capture at a resolution of 320 × 240 pixels. The second element in this design is a 1 W 450 nm laser module for area-excitation, which combined with the webcam allows for rapid interrogation of a flow field. The final element is a 2D flow-cell which overcomes the flow limitation of hydrodynamic focusing and allows for higher sample throughput in a wider flow field. This cell allows for the linear velocity of target cells to be lower than in a conventional "1D" hydrodynamic focusing flow-cells typically used in cytometry at similar volumetric flow rates. It also allows cells to be imaged at the full frame rate of the webcam. Using this webcam-based flow cytometer with wide-field imaging, it was confirmed that the detection of fluorescently tagged 5 μm polystyrene beads in "1D" hydrodynamic focusing flow-cells was not practical for low cell number detection due to streaking from the motion of the beads, which did not occur with the 2D flow-cell design. The sensitivity and throughput of this webcam-based flow cytometer was then investigated using THP-1 human monocytes stained with SYTO-9 florescent dye in the 2D flow-cell. The flow cytometer was found to be capable of detecting fluorescently tagged cells at concentrations as low as 1 cell per mL at flow rates of 500 μL min(-1) in buffer and in blood. The effectiveness of detection was concentration dependent: at 100 cells per mL 84% of the cells were detected compared to microscopy, 10 cells per mL 79% detected and 1 cell per mL 59% of the cells were detected. With the blood samples spiked to 100 cells per mL, the average concentration for all samples was 91.4 cells per mL, with a 95% confidence interval of 86-97 cells per mL. These low cell concentrations and the large volume capabilities of the system may overcome the limitations of current cytometry, and are applicable to rare cell (such as circulating tumor cell) detection The simplicity and low cost of this device suggests that it may have a potential use in developing point-of-care clinical flow cytometry for resource-poor settings associated with global health.

[1]  Derek Tseng,et al.  Fluorescent imaging of single nanoparticles and viruses on a smart phone. , 2013, ACS nano.

[2]  John L. Crassidis,et al.  Sensors and actuators , 2005, Conference on Electron Devices, 2005 Spanish.

[3]  M. Ratajczak,et al.  Analytical capabilities of the ImageStream cytometer. , 2011, Methods in cell biology.

[4]  Yordan Kostov,et al.  Charged-coupled device (CCD) detectors for Lab-on-a Chip (LOC) optical analysis. , 2013, Methods in molecular biology.

[5]  Jeffrey S Erickson,et al.  Two simple and rugged designs for creating microfluidic sheath flow. , 2008, Lab on a chip.

[6]  W. Marsden I and J , 2012 .

[7]  Joshua Balsam,et al.  Thousand-fold fluorescent signal amplification for mHealth diagnostics. , 2014, Biosensors & bioelectronics.

[8]  Lisa C Shriver-Lake,et al.  Multiplexed measurement of serum antibodies using an array biosensor. , 2006, Biosensors & bioelectronics.

[9]  Yordan Kostov,et al.  A fluorescence detection platform using spatial electroluminescent excitation for measuring botulinum neurotoxin A activity. , 2008, Biosensors & bioelectronics.

[10]  Joshua Balsam,et al.  Low-cost technologies for medical diagnostics in low-resource settings. , 2013, Expert opinion on medical diagnostics.

[11]  Lisa C Shriver-Lake,et al.  Detection of deoxynivalenol in foods and indoor air using an array biosensor. , 2006, Environmental science & technology.

[12]  Hongying Zhu,et al.  Cost-effective and rapid blood analysis on a cell-phone. , 2013, Lab on a chip.

[13]  Yordan Kostov,et al.  Lab-on-a-chip for botulinum neurotoxin a (BoNT-A) activity analysis. , 2009, Lab on a chip.

[14]  A. Rasooly,et al.  A simple portable electroluminescence illumination-based CCD detector. , 2009, Methods in molecular biology.

[15]  Yordan Kostov,et al.  Image stacking approach to increase sensitivity of fluorescence detection using a low cost complementary metal-oxide-semiconductor (CMOS) webcam. , 2012, Sensors and actuators. B, Chemical.

[16]  Aydogan Ozcan,et al.  Smart-phone based computational microscopy using multi-frame contact imaging on a fiber-optic array. , 2013, Lab on a chip.

[17]  Joshua Balsam,et al.  Orthographic projection capillary array fluorescent sensor for mHealth. , 2013, Methods.

[18]  Hongying Zhu,et al.  Wide-field fluorescent microscopy and fluorescent imaging flow cytometry on a cell-phone. , 2013, Journal of visualized experiments : JoVE.

[19]  Yordan Kostov,et al.  A simple 96 well microfluidic chip combined with visual and densitometry detection for resource-poor point of care testing. , 2011, Sensors and actuators. B, Chemical.

[20]  Lisa R. Hilliard,et al.  Multi-wavelength microflow cytometer using groove-generated sheath flow. , 2009, Lab on a chip.

[21]  Lisa C Shriver-Lake,et al.  The Array Biosensor: Portable, Automated Systems , 2007, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[22]  Hongying Zhu,et al.  Wide-field fluorescent microscopy on a cell-phone , 2011, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[23]  F. Ligler,et al.  Evanescent wave fluorescence biosensors. , 2005, Biosensors & bioelectronics.

[24]  A. Rasooly,et al.  Multi-wavelength Spatial LED illumination based detector for in vitro detection of Botulinum Neurotoxin A Activity. , 2010, Sensors and actuators. B, Chemical.

[25]  P. Morrissey,et al.  Quantitative analysis of protein co-localization on B cells opsonized with rituximab and complement using the ImageStream multispectral imaging flow cytometer. , 2006, Journal of immunological methods.

[26]  A. Rasooly,et al.  Miniaturized 96-well ELISA chips for staphylococcal enterotoxin B detection using portable colorimetric detector , 2009, Analytical and bioanalytical chemistry.

[27]  A. Rasooly,et al.  ELISA-LOC: lab-on-a-chip for enzyme-linked immunodetection. , 2010, Lab on a chip.

[28]  Yordan Kostov,et al.  An ELISA Lab-on-a-Chip (ELISA-LOC). , 2013, Methods in molecular biology.

[29]  David A Basiji,et al.  Distinguishing modes of cell death using the ImageStream® multispectral imaging flow cytometer , 2004, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[30]  A. Minguela,et al.  Imaging cytometry for counting circulating tumor cells: comparative analysis of the CellSearch vs ImageStream systems , 2013, APMIS : acta pathologica, microbiologica, et immunologica Scandinavica.

[31]  A. Ozcan,et al.  Quantum dot enabled detection of Escherichia coli using a cell-phone. , 2012, The Analyst.

[32]  Yordan Kostov,et al.  Lensless CCD-based fluorometer using a micromachined optical Söller collimator. , 2011, Lab on a chip.

[33]  Hugh Alan Bruck,et al.  Modeling and design of micromachined optical Söller collimators for lensless CCD-based fluorometry. , 2012, The Analyst.

[34]  Hongying Zhu,et al.  Cost-effective and compact wide-field fluorescent imaging on a cell-phone. , 2011, Lab on a chip.

[35]  Joshua Balsam,et al.  Capillary Array Waveguide Amplified Fluorescence Detector for mHealth. , 2013, Sensors and actuators. B, Chemical.

[36]  Aydogan Ozcan,et al.  Albumin testing in urine using a smart-phone. , 2013, Lab on a chip.

[37]  Hongying Zhu,et al.  Optofluidic fluorescent imaging cytometry on a cell phone. , 2011, Analytical chemistry.