Detection of hyperoxaemia in neonates: data from three new pulse oximeters

Aim: To determine the sensitivity and specificity of three newly developed pulse oximeters in the detection of hyperoxaemia, defined as an arterial partial pressure of oxygen (Pao2) of > 80 mm Hg. Methods: Spo2 readings from three oximeters (Agilent Viridia (AgV), Masimo SET (MaS), Nellcor Oxismart (NeO)) were documented in 56 infants (median gestational age at birth 35.5 weeks, range 24–41) whenever an arterial blood gas was taken for clinical purposes. Blood samples were analysed within one minute in a Radiometer ABL 505 blood gas analyser and OSM3 co-oximeter. Results: Between 280 and 291 blood gases were analysed for each instrument; 105–112 showed a Pao2 > 80 mm Hg. At an upper alarm limit of 95%, the three instruments detected hyperoxaemia with 93–95% sensitivity. Specificity at this alarm level ranged from 26 to 45%. The mean (SD) difference between arterial oxygen saturation and Spo2 (bias) was −0.25 (2.5)% for AgV, −0.06 (2.5)% for MaS, and −0.91 (2.6)% for NeO (p < 0.01, NeO v AgV and MaS). Conclusion: These instruments detected hyperoxaemia with sufficient sensitivity at an upper alarm limit of 95%, but showed differences in their specificity, which was probably related to differences in measurement bias.

[1]  G. Koch,et al.  Adjustment of arterial blood gases and acid base balance in the normal newborn infant during the first week of life. , 1968, Biologia neonatorum. Neo-natal studies.

[2]  J. Flynn,et al.  A cohort study of transcutaneous oxygen tension and the incidence and severity of retinopathy of prematurity. , 1992, The New England journal of medicine.

[3]  C. Poets,et al.  Oxygen saturation during the first 24 hours of life , 2000, Archives of disease in childhood. Fetal and neonatal edition.

[4]  F Paky,et al.  Pulse oximetry in ventilated preterm newborns: reliability of detection of hyperoxaemia and hypoxaemia, and feasibility of alarm settings , 1995, Acta paediatrica.

[5]  C. Poets,et al.  Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern. , 1994, Pediatrics.

[6]  H. Bucher,et al.  Hyperoxemia in newborn infants: detection by pulse oximetry. , 1989, Pediatrics.

[7]  M. Jopling,et al.  Issues in the laboratory evaluation of pulse oximeter performance. , 2002, Anesthesia and analgesia.

[8]  M. Healy Statistics from the inside. 14. Diagnostic and screening tests and reference values. , 1995, Archives of disease in childhood.

[9]  C. Poets,et al.  [Accuracy of measurements of Kontrol 7840, Nellcor N200 and Radiometer OX13 pulse oximeters in infants and young children]. , 1993, Klinische Padiatrie.

[10]  W. Hay Physiology of oxygenation and its relation to pulse oximetry in neonates. , 1987, Journal of perinatology : official journal of the California Perinatal Association.

[11]  S. Lawless Crying wolf: False alarms in a pediatric intensive care unit , 1994, Critical care medicine.

[12]  C. Poets,et al.  Reliability of a pulse oximeter in the detection of hyperoxemia. , 1993, Jornal de Pediatria.