Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage.

PURPOSE To determine the dosimetric effects of rotational errors on target coverage using volumetric modulated arc therapy (VMAT) for multitarget stereotactic radiosurgery (SRS). METHODS AND MATERIALS This retrospective study included 50 SRS cases, each with 2 intracranial planning target volumes (PTVs). Both PTVs were planned for simultaneous treatment to 21 Gy using a single-isocenter, noncoplanar VMAT SRS technique. Rotational errors of 0.5°, 1.0°, and 2.0° were simulated about all axes. The dose to 95% of the PTV (D95) and the volume covered by 95% of the prescribed dose (V95) were evaluated using multivariate analysis to determine how PTV coverage was related to PTV volume, PTV separation, and rotational error. RESULTS At 0.5° rotational error, D95 values and V95 coverage rates were ≥95% in all cases. For rotational errors of 1.0°, 7% of targets had D95 and V95 values <95%. Coverage worsened substantially when the rotational error increased to 2.0°: D95 and V95 values were >95% for only 63% of the targets. Multivariate analysis showed that PTV volume and distance to isocenter were strong predictors of target coverage. CONCLUSIONS The effects of rotational errors on target coverage were studied across a broad range of SRS cases. In general, the risk of compromised coverage increased with decreasing target volume, increasing rotational error and increasing distance between targets. Multivariate regression models from this study may be used to quantify the dosimetric effects of rotational errors on target coverage given patient-specific input parameters of PTV volume and distance to isocenter.

[1]  J. Kuo,et al.  The use of RapidArc volumetric-modulated arc therapy to deliver stereotactic radiosurgery and stereotactic body radiotherapy to intracranial and extracranial targets. , 2012, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[2]  Matthias Guckenberger,et al.  Dosimetric consequences of translational and rotational errors in frame-less image-guided radiosurgery , 2012, Radiation oncology.

[3]  L. Lin,et al.  A concordance correlation coefficient to evaluate reproducibility. , 1989, Biometrics.

[4]  Mauro Iori,et al.  Testing of the analytical anisotropic algorithm for photon dose calculation. , 2006, Medical physics.

[5]  R. Popple,et al.  Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases. , 2010, International Journal of Radiation Oncology, Biology, Physics.

[6]  H. Shu,et al.  Six degrees of freedom CBCT‐based positioning for intracranial targets treated with frameless stereotactic radiosurgery , 2012, Journal of applied clinical medical physics.

[7]  G. Mahon A Proposal for Strength-of-Agreement Criteria for Lin’s Concordance Correlation Coefficient , 2005 .

[8]  B. Winey,et al.  Geometric and dosimetric uncertainties in intracranial stereotatctic treatments for multiple nonisocentric lesions , 2014, Journal of applied clinical medical physics.

[9]  Robert B. McBride,et al.  A proposal for strength-of-agreement criteria for Lins Concordance Correlation Coefficient , 2005 .

[10]  K Y Liang,et al.  Longitudinal data analysis for discrete and continuous outcomes. , 1986, Biometrics.

[11]  M A Schork,et al.  A logistic regression model for the decision to perform access surgery. , 1997, Journal of clinical periodontology.

[12]  Naren Ramakrishna,et al.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[13]  Z. Ying,et al.  Model‐Checking Techniques Based on Cumulative Residuals , 2002, Biometrics.

[14]  J. Wong,et al.  A method for optimizing LINAC treatment geometry for volumetric modulated arc therapy of multiple brain metastases. , 2010, Medical physics.

[15]  Thierry Gevaert,et al.  Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery. , 2012, International journal of radiation oncology, biology, physics.