Physiotherapy as a Rare Cause of Twiddler’s Syndrome in a Patient With an Implanted Cardioverter Defibrillator

A 65-year-old male patient with a history of ischemic cardiomyopathy developed ventricular tachycardia resulting in presyncope. An ICD was indicated for secondary prophylaxis of ventricular tachyarrhythmias. A dual chamber ICD was implanted from the right side because insertion of the device from the left side was unfeasible after surgery of a left subscapularis tendon lesion. ICD implantation and testing of defibrillation threshold were uneventful. During early follow-up a progressive increase of the stimulation threshold was detected. On chest X-ray coiling of both atrial and ventricular leads was noted and caused inadvertently by active shoulder-arm physiotherapy. Complete revision of the ICD system was necessary for restoration of the pacemaker function of the ICD. This unique case highlights important steps for early recognition and prevention of Twiddler’s syndrome that may occur due to physiotherapy treatment even without abnormal manipulations by the patient.

[1]  R. Strasser,et al.  Twiddler’s syndrome in an adolescent patient with ICD during neurological and physical rehabilitation , 2009, Clinical Research in Cardiology.

[2]  S. Spencker,et al.  An unusual cause of oversensing leading to inappropriate ICD discharges. , 2008, International journal of cardiology.

[3]  B. Wilkoff,et al.  Pacemaker and ICD leads: Strategies for long-term management , 2008, Journal of Interventional Cardiac Electrophysiology.

[4]  C. Bode,et al.  Multiple inappropriate defibrillator discharges due to Twiddler's syndrome , 2005, Wiener klinische Wochenschrift.

[5]  C. Israel,et al.  “Twiddler Syndrome” in a Subpectorally Implanted Cardioverter Defibrillator , 2002, Journal of cardiovascular electrophysiology.

[6]  M. Josephson,et al.  Twiddler's Syndrome Variants in ICD Patients , 1998, Pacing and clinical electrophysiology : PACE.

[7]  D. Fitzgerald,et al.  Defibrillator Twiddler's Syndrome Causing Device Failure in a Subpectoral Transvenous System , 1996, Pacing and clinical electrophysiology : PACE.

[8]  J. Windle,et al.  Defibrillator twiddler's syndrome. , 1994, The Annals of thoracic surgery.

[9]  D. Beanlands,et al.  The pacemaker-twiddler's syndrome: a new complication of implantable transvenous pacemakers. , 1968, Canadian Medical Association journal.

[10]  B. Dijkman,et al.  Lead system causing twiddler's syndrome in patients with an implantable cardioverter-defibrillator. , 2005, The Journal of thoracic and cardiovascular surgery.

[11]  V. Dörnberger,et al.  [Twiddler-Syndrome in a subpectoral implanted unipolar cardioverter-defibrillator]. , 1998, Zeitschrift für Kardiologie.

[12]  L. Seipel,et al.  Twiddler-Syndrom bei einem subpektoral implantierten unipolaren Cardioverter-Defibrillator , 1998, Zeitschrift für Kardiologie.